{"title":"The effect of forearm rotation on the bone mineral density measurements of the distal radius.","authors":"Keisuke Uemura, Satoshi Miyamura, Yoshito Otake, Hirokazu Mae, Kazuma Takashima, Hidetoshi Hamada, Kosuke Ebina, Tsuyoshi Murase, Yoshinobu Sato, Seiji Okada","doi":"10.1007/s00774-023-01473-4","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Forearm dual-energy X-ray absorptiometry (DXA) is often performed in clinics where central DXA is unavailable. Accurate bone mineral density (BMD) measurement is crucial for clinical assessment. Forearm rotation can affect BMD measurements, but this effect remains uncertain. Thus, we aimed to conduct a simulation study using CT images to clarify the effect of forearm rotation on BMD measurements.</p><p><strong>Materials and methods: </strong>Forearm CT images of 60 women were analyzed. BMD was measured at the total, ultra-distal (UD), mid-distal (MD), and distal 33% radius regions with the radius located at the neutral position using digitally reconstructed radiographs generated from CT images. Then, the rotation was altered from - 30° to 30° (supination set as positive) with a one-degree increment, and the percent BMD changes from the neutral position were quantified for all regions at each angle for each patient.</p><p><strong>Results: </strong>The maximum mean BMD changes were 5.8%, 7.0%, 6.2%, and 7.2% for the total, UD, MD, and distal 33% radius regions, respectively. The analysis of the absolute values of the percent BMD changes from the neutral position showed that BMD changes of all patients remained within 2% when the rotation was between - 5° and 7° for the total region, between - 3° and 2° for the UD region, between - 4° and 3° for the MD region, and between - 3° and 1° for the distal 33% radius region.</p><p><strong>Conclusion: </strong>Subtle rotational changes affected the BMD measurement of each region. The results showed the importance of forearm positioning when measuring the distal radius BMD.</p>","PeriodicalId":15116,"journal":{"name":"Journal of Bone and Mineral Metabolism","volume":" ","pages":"37-46"},"PeriodicalIF":2.4000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Bone and Mineral Metabolism","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00774-023-01473-4","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/12/6 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"ENDOCRINOLOGY & METABOLISM","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction: Forearm dual-energy X-ray absorptiometry (DXA) is often performed in clinics where central DXA is unavailable. Accurate bone mineral density (BMD) measurement is crucial for clinical assessment. Forearm rotation can affect BMD measurements, but this effect remains uncertain. Thus, we aimed to conduct a simulation study using CT images to clarify the effect of forearm rotation on BMD measurements.
Materials and methods: Forearm CT images of 60 women were analyzed. BMD was measured at the total, ultra-distal (UD), mid-distal (MD), and distal 33% radius regions with the radius located at the neutral position using digitally reconstructed radiographs generated from CT images. Then, the rotation was altered from - 30° to 30° (supination set as positive) with a one-degree increment, and the percent BMD changes from the neutral position were quantified for all regions at each angle for each patient.
Results: The maximum mean BMD changes were 5.8%, 7.0%, 6.2%, and 7.2% for the total, UD, MD, and distal 33% radius regions, respectively. The analysis of the absolute values of the percent BMD changes from the neutral position showed that BMD changes of all patients remained within 2% when the rotation was between - 5° and 7° for the total region, between - 3° and 2° for the UD region, between - 4° and 3° for the MD region, and between - 3° and 1° for the distal 33% radius region.
Conclusion: Subtle rotational changes affected the BMD measurement of each region. The results showed the importance of forearm positioning when measuring the distal radius BMD.
期刊介绍:
The Journal of Bone and Mineral Metabolism (JBMM) provides an international forum for researchers and clinicians to present and discuss topics relevant to bone, teeth, and mineral metabolism, as well as joint and musculoskeletal disorders. The journal welcomes the submission of manuscripts from any country. Membership in the society is not a prerequisite for submission. Acceptance is based on the originality, significance, and validity of the material presented. The journal is aimed at researchers and clinicians dedicated to improvements in research, development, and patient-care in the fields of bone and mineral metabolism.