Stress analysis of a total hip acetabular component: an FEM study.

D S Burton, H B Skinner, M Eng
{"title":"Stress analysis of a total hip acetabular component: an FEM study.","authors":"D S Burton,&nbsp;H B Skinner,&nbsp;M Eng","doi":"10.3109/10731198909118852","DOIUrl":null,"url":null,"abstract":"<p><p>A stress analysis of a total hip acetabular component was performed using three-dimensional finite element modeling. The model consisted of 548 four-noded quadrilateral shell elements with 582 nodes. A worst-case support condition was assumed in which bony contact with small areas of the ilium, ischium, and pubis was represented by three-point support. Loads corresponding to the peak pressures developed in the hip when rising from a seated position--the activity contributing most to the fatigue of a prosthesis--were applied to the model. Peak stresses of 490 MPa are predicted around a screw hole in the region of greatest loading; according to the fatigue curve for porous-coated Ti-6Al-4V alloy, this translates into a component life of three years. When the area of bony support for the prosthesis is increased by a factor of 1.5, the fatigue life is more than doubled to seven years. The addition of restraints around the loaded hole to represent the placement of a screw extends the lifetime dramatically to 65 years. Substituting the material properties of cobalt-chromium alloy increases the predicted lifetime of the component to twelve years. We conclude that poor bone support can compromise the service life of titanium alloy acetabular components. Furthermore, we suggest that screw fixation be used to secure the prosthesis in the anterior-superior region if the prosthesis is not well supported by bone in the acetabulum.</p>","PeriodicalId":77498,"journal":{"name":"Biomaterials, artificial cells, and artificial organs","volume":"17 4","pages":"371-83"},"PeriodicalIF":0.0000,"publicationDate":"1989-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.3109/10731198909118852","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biomaterials, artificial cells, and artificial organs","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3109/10731198909118852","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 2

Abstract

A stress analysis of a total hip acetabular component was performed using three-dimensional finite element modeling. The model consisted of 548 four-noded quadrilateral shell elements with 582 nodes. A worst-case support condition was assumed in which bony contact with small areas of the ilium, ischium, and pubis was represented by three-point support. Loads corresponding to the peak pressures developed in the hip when rising from a seated position--the activity contributing most to the fatigue of a prosthesis--were applied to the model. Peak stresses of 490 MPa are predicted around a screw hole in the region of greatest loading; according to the fatigue curve for porous-coated Ti-6Al-4V alloy, this translates into a component life of three years. When the area of bony support for the prosthesis is increased by a factor of 1.5, the fatigue life is more than doubled to seven years. The addition of restraints around the loaded hole to represent the placement of a screw extends the lifetime dramatically to 65 years. Substituting the material properties of cobalt-chromium alloy increases the predicted lifetime of the component to twelve years. We conclude that poor bone support can compromise the service life of titanium alloy acetabular components. Furthermore, we suggest that screw fixation be used to secure the prosthesis in the anterior-superior region if the prosthesis is not well supported by bone in the acetabulum.

全髋臼构件的应力分析:有限元研究。
采用三维有限元模型对全髋臼构件进行应力分析。该模型由548个四节点四边形壳单元组成,节点582个。假设最坏情况下的支持条件,即与髂骨、坐骨和耻骨的小区域的骨接触由三点支持表示。当从坐姿上升时,髋部产生的峰值压力对应的负载(这是导致假体疲劳的最主要活动)被应用于模型。在最大载荷区域,螺孔周围的峰值应力可达490 MPa;根据多孔涂层Ti-6Al-4V合金的疲劳曲线,这意味着部件寿命为三年。当假体的骨支撑面积增加1.5倍时,疲劳寿命增加一倍以上,达到7年。在加载孔周围增加约束,代表螺钉的位置,将使用寿命显著延长至65年。取代钴铬合金的材料特性,将部件的预测寿命提高到12年。我们认为,不良的骨支撑会影响钛合金髋臼假体的使用寿命。此外,如果假体不能很好地由髋臼骨支撑,我们建议采用螺钉固定来固定假体的前上区域。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书