ANALISIS PAPARAN RADIASI SKYSHINE DINDING RUANGAN ANGIOGRAFI/CATHLAB RUMAH SAKIT MITRA KELUARGA KEMAYORAN BERSUDUT GANTRY 0 DERAJAT

Adi Andhika, Ghifari Muhammad Fajri, Hendra Suroso
{"title":"ANALISIS PAPARAN RADIASI SKYSHINE DINDING RUANGAN ANGIOGRAFI/CATHLAB RUMAH SAKIT MITRA KELUARGA KEMAYORAN BERSUDUT GANTRY 0 DERAJAT","authors":"Adi Andhika, Ghifari Muhammad Fajri, Hendra Suroso","doi":"10.53862/ssi.v3.092023.020","DOIUrl":null,"url":null,"abstract":"Based on NCRP (2005), skyshine radiation generally occurs in radiotherapy facilities, the cause of which is that some radiation facilities are designed with minimal shielding on the linear accelerator (Linac) ceiling. As time goes by, Skyshine radiation monitoring is not only focused on radiotherapy facilities, but radiodiagnostic facilities also need to be monitored, especially interventional radiology, which uses real-time X-ray radiation at certain times. Regarding this matter, it is necessary to check whether the walls of the existing angiography room or cath lab are protected from skyshine radiation. This study aims to determine whether the Angiography/Cathlab room at Mitra Keluarga Kemayoran Hospital does not produce skyshine radiation. The method for measuring skyshine radiation exposure uses a survey meter at a distance of 30 cm and 2 meters 3 times from the outer wall of the Angiography/Cathlab room on the north, south, and west sides of the gantry; then the gantry is positioned at 0 degree. The research results show that radiation exposure is 0.14 µSv/hour at a distance of 30 cm. Skyshine radiation was detected at a distance of 2 meters with an exposure value of 0.32 µSv/hour on the west side of the gantry and 0.39 µSv/hour on the north side of the gantry, this occurred due to the radiation barrier wall (Pb ) not complete until the ceiling is only 2 meters high from the floor. The conclusion is that the walls of the Angiography/Cathlab room at Mitra Keluarga Kemayoran Hospital have exposure to skyshine radiation at a distance of 2 meters measured from the outer walls of the Angiography/Cathlab room on the west and north sides of the gantry of 0.32 µSv/hour and 0.39 µSv/hour. Keywords: angiography/cath lab, radiation, X-ray, skyshine, interventional radiology.","PeriodicalId":476035,"journal":{"name":"Prosiding Seminar Si-INTAN","volume":"50 3-4","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2023-10-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Prosiding Seminar Si-INTAN","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.53862/ssi.v3.092023.020","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Based on NCRP (2005), skyshine radiation generally occurs in radiotherapy facilities, the cause of which is that some radiation facilities are designed with minimal shielding on the linear accelerator (Linac) ceiling. As time goes by, Skyshine radiation monitoring is not only focused on radiotherapy facilities, but radiodiagnostic facilities also need to be monitored, especially interventional radiology, which uses real-time X-ray radiation at certain times. Regarding this matter, it is necessary to check whether the walls of the existing angiography room or cath lab are protected from skyshine radiation. This study aims to determine whether the Angiography/Cathlab room at Mitra Keluarga Kemayoran Hospital does not produce skyshine radiation. The method for measuring skyshine radiation exposure uses a survey meter at a distance of 30 cm and 2 meters 3 times from the outer wall of the Angiography/Cathlab room on the north, south, and west sides of the gantry; then the gantry is positioned at 0 degree. The research results show that radiation exposure is 0.14 µSv/hour at a distance of 30 cm. Skyshine radiation was detected at a distance of 2 meters with an exposure value of 0.32 µSv/hour on the west side of the gantry and 0.39 µSv/hour on the north side of the gantry, this occurred due to the radiation barrier wall (Pb ) not complete until the ceiling is only 2 meters high from the floor. The conclusion is that the walls of the Angiography/Cathlab room at Mitra Keluarga Kemayoran Hospital have exposure to skyshine radiation at a distance of 2 meters measured from the outer walls of the Angiography/Cathlab room on the west and north sides of the gantry of 0.32 µSv/hour and 0.39 µSv/hour. Keywords: angiography/cath lab, radiation, X-ray, skyshine, interventional radiology.
这是对中心室血管造影暴露的分析
根据NCRP(2005),在放射治疗设施中普遍存在天光辐射,其原因是一些放射设施在设计时对直线加速器(Linac)天花板的屏蔽很小。随着时间的推移,Skyshine辐射监测不仅集中在放射治疗设施,放射诊断设施也需要监测,特别是介入放射,它在某些时间使用实时x射线辐射。对于这个问题,有必要检查现有的血管造影室或导管室的墙壁是否受到日光辐射的保护。本研究旨在确定Mitra Keluarga Kemayoran医院的血管造影/Cathlab室是否不产生天光辐射。测量天光辐射暴露的方法是在龙门北侧、南侧、西侧血管造影/Cathlab室外墙上30厘米、2米处放置测量仪3次;然后将龙门架定位在0度。研究结果表明,在30 cm距离处的辐射暴露量为0.14µSv/h。在2米距离处检测到天光辐射,龙门西侧的照射值为0.32µSv/小时,龙门北侧的照射值为0.39µSv/小时,这是由于辐射屏障墙(Pb)在天花板距地面仅2米高时才完成。结论是,Mitra Keluarga Kemayoran医院血管造影/Cathlab室的墙壁在距离龙门西侧和北侧血管造影/Cathlab室外墙2米的距离处受到的天光辐射分别为0.32µSv/小时和0.39µSv/小时。关键词:血管造影/导管实验室,放射线,x线,天光,介入放射学。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书