OPTIMISASI DOSIS DAN KUALITAS CITRA PADA PEMERIKSAAN PULMONARY ARTERY CATHETER (PAC) DAN PERCUTANEUS CORONARY INTERVENTION (PCI) SEBAGAI UPAYA PROTEKSI RADIASI PASIEN DI CATHLAB RSUP ADAM MALIK MEDAN

Martua Damanik, Josepa Simanjuntak, Gembira Karo-Karo, Artiyanti Artiyanti
{"title":"OPTIMISASI DOSIS DAN KUALITAS CITRA PADA PEMERIKSAAN PULMONARY ARTERY CATHETER (PAC) DAN PERCUTANEUS CORONARY INTERVENTION (PCI) SEBAGAI UPAYA PROTEKSI RADIASI PASIEN DI CATHLAB RSUP ADAM MALIK MEDAN","authors":"Martua Damanik, Josepa Simanjuntak, Gembira Karo-Karo, Artiyanti Artiyanti","doi":"10.53862/ssi.v3.092023.005","DOIUrl":null,"url":null,"abstract":"The radiation exposure a patient receives during diagnosis and treatment cannot be fully avoided. However, it is crucial to optimize the radiation dose to ensure it is only as much as needed. The purpose of this study is to minimize the patient's radiation exposure during interventions and produce optimal image quality. To accomplish this, dose optimization and image quality were evaluated during PAC and PCI procedures in the Cathlab room at Adam Malik Hospital using Philips Allura Xper FD20. The study was retrospective, using patient dose audits from January 2020 to February 2023, with 250 PACs and 139 PCI mature patients. The median dose value (Q2) was compared to the National diagnostic reference level (NDRL) third quartile (Q3) value to determine optimal patient dose levels. We used a 28 cm PMMA phantom and a 2 cm imaging quality in-house phantom to represent the patient and a solid-state detector to measure surface dose. Medical physicists, radiographers, cardiologists, and nurses worked together in optimization cycles to ensure dose levels were kept to a minimum. Image quality was assessed by calculating the signal difference-to-noise ratio (SDNR), and images were recorded in DICOM format and analyzed using Image J software. Entrance Surface Air Kerma (ESAK) values were recorded to calculate Figure of Merit (FOM) values for each change in physical parameters. The study showed that the distance between the image detector and the object significantly affects the patient's dose. A distance of 14 cm resulted in a higher dose than a distance of 2 cm. The SDNR value varied with the object image distance (OID) factor, and increasing the object's distance to the image detector did not increase the SDNR value. Setting the OID spacing to a minimum is recommended for PAC and PCI procedures to minimize the patient's radiation exposure. The FOM values were affected by both the ESAK and SDNR values. Keywords: Dose Optimization, Image Quality, objek to image distance (OID), Pulmonary Artery Catheter, Percutaneous Coronary Intervention, Cathlab.","PeriodicalId":476035,"journal":{"name":"Prosiding Seminar Si-INTAN","volume":"66 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2023-10-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Prosiding Seminar Si-INTAN","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.53862/ssi.v3.092023.005","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The radiation exposure a patient receives during diagnosis and treatment cannot be fully avoided. However, it is crucial to optimize the radiation dose to ensure it is only as much as needed. The purpose of this study is to minimize the patient's radiation exposure during interventions and produce optimal image quality. To accomplish this, dose optimization and image quality were evaluated during PAC and PCI procedures in the Cathlab room at Adam Malik Hospital using Philips Allura Xper FD20. The study was retrospective, using patient dose audits from January 2020 to February 2023, with 250 PACs and 139 PCI mature patients. The median dose value (Q2) was compared to the National diagnostic reference level (NDRL) third quartile (Q3) value to determine optimal patient dose levels. We used a 28 cm PMMA phantom and a 2 cm imaging quality in-house phantom to represent the patient and a solid-state detector to measure surface dose. Medical physicists, radiographers, cardiologists, and nurses worked together in optimization cycles to ensure dose levels were kept to a minimum. Image quality was assessed by calculating the signal difference-to-noise ratio (SDNR), and images were recorded in DICOM format and analyzed using Image J software. Entrance Surface Air Kerma (ESAK) values were recorded to calculate Figure of Merit (FOM) values for each change in physical parameters. The study showed that the distance between the image detector and the object significantly affects the patient's dose. A distance of 14 cm resulted in a higher dose than a distance of 2 cm. The SDNR value varied with the object image distance (OID) factor, and increasing the object's distance to the image detector did not increase the SDNR value. Setting the OID spacing to a minimum is recommended for PAC and PCI procedures to minimize the patient's radiation exposure. The FOM values were affected by both the ESAK and SDNR values. Keywords: Dose Optimization, Image Quality, objek to image distance (OID), Pulmonary Artery Catheter, Percutaneous Coronary Intervention, Cathlab.
肺动脉(PAC)和PERCUTANEUS CORONARY INTERVENTION (PCI)上的优化剂量和图像质量
患者在诊断和治疗过程中受到的辐射暴露是无法完全避免的。然而,至关重要的是要优化辐射剂量,以确保它只需要多少。本研究的目的是尽量减少患者在干预期间的辐射暴露,并产生最佳的图像质量。为了实现这一目标,在Adam Malik医院的Cathlab室使用Philips Allura Xper FD20评估PAC和PCI过程中的剂量优化和图像质量。该研究是回顾性的,使用了2020年1月至2023年2月的患者剂量审计,包括250名pac和139名PCI成熟患者。将中位剂量值(Q2)与国家诊断参考水平(NDRL)第三个四分位数(Q3)值进行比较,以确定最佳患者剂量水平。我们使用28厘米的PMMA假体和2厘米成像质量的内部假体来代表患者,并使用固态探测器来测量表面剂量。医学物理学家、放射技师、心脏病专家和护士在优化周期中共同努力,以确保剂量水平保持在最低限度。通过计算信号差噪比(SDNR)评估图像质量,并以DICOM格式记录图像,使用Image J软件进行分析。记录入口表面空气Kerma (ESAK)值,以计算每次物理参数变化的优点图(FOM)值。研究表明,图像检测器与物体之间的距离对患者的剂量有显著影响。距离14厘米产生的剂量高于距离2厘米产生的剂量。SDNR值随目标图像距离(OID)因子的变化而变化,增加目标到图像检测器的距离并没有增加SDNR值。对于PAC和PCI手术,建议将OID间距设置为最小,以减少患者的辐射暴露。FOM值同时受到ESAK和SDNR值的影响。关键词:剂量优化,图像质量,物像距离(OID),肺动脉导管,经皮冠状动脉介入治疗,Cathlab。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书