Achintya D Singh, Deepak Madhu, Mythili Menon Pathiyil, Daryl Ramai, Babu P Mohan, Bhavesh Shah, Douglas G Adler
{"title":"Device malfunctions with use of EUS-guided fine-needle biopsy devices: Analysis of the MAUDE database.","authors":"Achintya D Singh, Deepak Madhu, Mythili Menon Pathiyil, Daryl Ramai, Babu P Mohan, Bhavesh Shah, Douglas G Adler","doi":"10.1097/eus.0000000000000016","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The safety of endoscopic ultrasound-guided tissue acquisition through fine-needle biopsy devices is well-established in clinical trials. The real-world experience of using these devices is not known. The authors analyzed the postmarketing surveillance data from the Food and Drug Administration Manufacturer and User Facility Device Experience (MAUDE) database to answer this question.</p><p><strong>Methods: </strong>The Food and Drug Administration MAUDE database from January 2012 to June 2022 was accessed to evaluate for device malfunctions and patient-related adverse consequences of these malfunctions.</p><p><strong>Results: </strong>There were 344 device-related issues. Most issues were due to detachment or breakage of the device (<i>n</i> = 185 [53.7%]). Seventy-six of the breakages (40.8%) occurred during the procedure, whereas 89 cases (47.8%) occurred while removing the needle from the endoscope. The most common site of tissue biopsy at the time of needle breakage was the pancreas (44 [23.8%]).The common patient-related adverse events were retained foreign body (<i>n</i> = 50 [14.5%]) followed by bleeding (16, 4.6%). Six patients (3.4%) required a second intervention for removal of the retained foreign bodies including surgery in 2 cases. The device breakage damaged the endoscope in 3 cases (1.7%), and there was 1 case of needlestick injury to the nurse.</p><p><strong>Conclusion: </strong>The fine-needle biopsy devices can be associated with needle breakage and bending; these adverse events were not previously reported. Needle breakages can result in a retained foreign body that may require additional procedures including surgery. These real-world findings from the MAUDE database may inform clinical decisions and help improve clinical outcomes.</p>","PeriodicalId":11577,"journal":{"name":"Endoscopic Ultrasound","volume":"12 5","pages":"424-427"},"PeriodicalIF":4.4000,"publicationDate":"2023-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10631607/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Endoscopic Ultrasound","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/eus.0000000000000016","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/8/12 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"GASTROENTEROLOGY & HEPATOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background: The safety of endoscopic ultrasound-guided tissue acquisition through fine-needle biopsy devices is well-established in clinical trials. The real-world experience of using these devices is not known. The authors analyzed the postmarketing surveillance data from the Food and Drug Administration Manufacturer and User Facility Device Experience (MAUDE) database to answer this question.
Methods: The Food and Drug Administration MAUDE database from January 2012 to June 2022 was accessed to evaluate for device malfunctions and patient-related adverse consequences of these malfunctions.
Results: There were 344 device-related issues. Most issues were due to detachment or breakage of the device (n = 185 [53.7%]). Seventy-six of the breakages (40.8%) occurred during the procedure, whereas 89 cases (47.8%) occurred while removing the needle from the endoscope. The most common site of tissue biopsy at the time of needle breakage was the pancreas (44 [23.8%]).The common patient-related adverse events were retained foreign body (n = 50 [14.5%]) followed by bleeding (16, 4.6%). Six patients (3.4%) required a second intervention for removal of the retained foreign bodies including surgery in 2 cases. The device breakage damaged the endoscope in 3 cases (1.7%), and there was 1 case of needlestick injury to the nurse.
Conclusion: The fine-needle biopsy devices can be associated with needle breakage and bending; these adverse events were not previously reported. Needle breakages can result in a retained foreign body that may require additional procedures including surgery. These real-world findings from the MAUDE database may inform clinical decisions and help improve clinical outcomes.
期刊介绍:
Endoscopic Ultrasound, a publication of Euro-EUS Scientific Committee, Asia-Pacific EUS Task Force and Latin American Chapter of EUS, is a peer-reviewed online journal with Quarterly print on demand compilation of issues published. The journal’s full text is available online at http://www.eusjournal.com. The journal allows free access (Open Access) to its contents and permits authors to self-archive final accepted version of the articles on any OAI-compliant institutional / subject-based repository. The journal does not charge for submission, processing or publication of manuscripts and even for color reproduction of photographs.