Chihyun Cho, Joo-Gwang Lee, P. Hale, J. Jargon, P. Jeavons, J. Schlager, A. Dienstfrey
{"title":"Calibration of channel mismatch in time-interleaved real-time digital oscilloscopes","authors":"Chihyun Cho, Joo-Gwang Lee, P. Hale, J. Jargon, P. Jeavons, J. Schlager, A. Dienstfrey","doi":"10.1109/ARFTG.2015.7162896","DOIUrl":null,"url":null,"abstract":"A novel method is proposed for the calibration of channel mismatch in a time-interleaved real-time digital oscilloscope (RTDO). A simple simultaneous equation is derived from the Fourier transform of the time-interleaved signals. Thus, it only requires a transfer function of time-interleaved ADCs (TIADCs), while most of previous works have employed additional filters. A measurement method for the transfer function of a commercial RTDO is also proposed. The accuracy of the calibration method is determined by the noise produced after the interleaving process. To validate the proposed method, we measure two-tone signals using a commercial RTDO. The calibrated results clearly show signals at spurious frequencies are substantially reduced.","PeriodicalId":228314,"journal":{"name":"2015 85th Microwave Measurement Conference (ARFTG)","volume":"1 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2015-05-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"12","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2015 85th Microwave Measurement Conference (ARFTG)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ARFTG.2015.7162896","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 12
Abstract
A novel method is proposed for the calibration of channel mismatch in a time-interleaved real-time digital oscilloscope (RTDO). A simple simultaneous equation is derived from the Fourier transform of the time-interleaved signals. Thus, it only requires a transfer function of time-interleaved ADCs (TIADCs), while most of previous works have employed additional filters. A measurement method for the transfer function of a commercial RTDO is also proposed. The accuracy of the calibration method is determined by the noise produced after the interleaving process. To validate the proposed method, we measure two-tone signals using a commercial RTDO. The calibrated results clearly show signals at spurious frequencies are substantially reduced.