P Liu, H Griffiths, C M Wiles, K M Nathadwarawala, W Stewart
{"title":"Measurement of pharyngeal transit time by electrical impedance tomography.","authors":"P Liu, H Griffiths, C M Wiles, K M Nathadwarawala, W Stewart","doi":"10.1088/0143-0815/13/a/038","DOIUrl":null,"url":null,"abstract":"<p><p>The feasibility of using electrical impedance tomography (EIT) for measuring pharyngeal transit time has been investigated. Imaging sequences were obtained on the neck in a group of eight healthy subjects when swallowing solutions of different electrical conductivities. The solutions used were aqueous Dioralyte (1.5 S m-1) and tap water (0.04 S m-1). The effects of movement upon the images were large compared with the changes due to the solutions, but these were cancelled by a subtraction technique. The mean value of the EIT pharyngeal transit time parameter was 370 +/- 120 ms. EIT has the advantage that repeated measurements could be performed without fear of hazard in contrast to the established technique of video fluoroscopy which involves ionising</p>","PeriodicalId":77070,"journal":{"name":"Clinical physics and physiological measurement : an official journal of the Hospital Physicists' Association, Deutsche Gesellschaft fur Medizinische Physik and the European Federation of Organisations for Medical Physics","volume":"13 Suppl A ","pages":"197-200"},"PeriodicalIF":0.0000,"publicationDate":"1992-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1088/0143-0815/13/a/038","citationCount":"16","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical physics and physiological measurement : an official journal of the Hospital Physicists' Association, Deutsche Gesellschaft fur Medizinische Physik and the European Federation of Organisations for Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1088/0143-0815/13/a/038","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 16
Abstract
The feasibility of using electrical impedance tomography (EIT) for measuring pharyngeal transit time has been investigated. Imaging sequences were obtained on the neck in a group of eight healthy subjects when swallowing solutions of different electrical conductivities. The solutions used were aqueous Dioralyte (1.5 S m-1) and tap water (0.04 S m-1). The effects of movement upon the images were large compared with the changes due to the solutions, but these were cancelled by a subtraction technique. The mean value of the EIT pharyngeal transit time parameter was 370 +/- 120 ms. EIT has the advantage that repeated measurements could be performed without fear of hazard in contrast to the established technique of video fluoroscopy which involves ionising
研究了电阻抗断层扫描(EIT)测量咽传输时间的可行性。在一组8名健康受试者中,当吞咽不同电导率的溶液时,获得了颈部的成像序列。使用的溶液为水透析液(1.5 S m-1)和自来水(0.04 S m-1)。运动对图像的影响比由解引起的变化更大,但这些影响通过减法技术被抵消了。EIT咽部传输时间参数平均值为370±120 ms。与已建立的涉及电离的视频透视技术相比,EIT的优点是可以进行重复测量而不必担心危险