{"title":"A network for verifying AC resistance measuring instruments","authors":"K. Jones","doi":"10.1109/CPEM.1998.699999","DOIUrl":null,"url":null,"abstract":"A resistance network for verifying the accuracy of resistance measuring instruments up described. The network has been instrument accuracies from 50 Hz to 1 MHz. Relative standard uncertainties of the method range from better than 1/spl times/10/sup -6/ at low frequency to 100/spl times/10/sup -6/ at 1 MHz.","PeriodicalId":239228,"journal":{"name":"1998 Conference on Precision Electromagnetic Measurements Digest (Cat. No.98CH36254)","volume":"14 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1998-07-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"1998 Conference on Precision Electromagnetic Measurements Digest (Cat. No.98CH36254)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/CPEM.1998.699999","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
A resistance network for verifying the accuracy of resistance measuring instruments up described. The network has been instrument accuracies from 50 Hz to 1 MHz. Relative standard uncertainties of the method range from better than 1/spl times/10/sup -6/ at low frequency to 100/spl times/10/sup -6/ at 1 MHz.