Measurement and prediction of drug solvent losses by evaporation from a jet nebuliser.

S W Smye, M I Jollie, H Cunliffe, J M Littlewood
{"title":"Measurement and prediction of drug solvent losses by evaporation from a jet nebuliser.","authors":"S W Smye,&nbsp;M I Jollie,&nbsp;H Cunliffe,&nbsp;J M Littlewood","doi":"10.1088/0143-0815/13/2/004","DOIUrl":null,"url":null,"abstract":"<p><p>The evaporative losses of solvent (water) from a commercial jet nebuliser (Unineb, Unimed Ltd) were estimated and found to be approximately 14% of total nebuliser output. A simple analysis of the nebulisation process then enabled the amount of the drug colomycin released as potentially respirable aerosol to be predicted from measurements of the total nebulisation time. Good agreement was found between the predicted (P) and measured (M) values such that P = 1.07 M (r2 = 0.98). The analysis also indicates that the proportion of the drug released as aerosol will increase as nebulisation proceeds provided nebuliser output is continuous and that for a given initial mass of drug a larger amount of drug is released as an aerosol if the volume of administration used in the nebuliser is as large as possible.</p>","PeriodicalId":77070,"journal":{"name":"Clinical physics and physiological measurement : an official journal of the Hospital Physicists' Association, Deutsche Gesellschaft fur Medizinische Physik and the European Federation of Organisations for Medical Physics","volume":"13 2","pages":"129-34"},"PeriodicalIF":0.0000,"publicationDate":"1992-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1088/0143-0815/13/2/004","citationCount":"11","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical physics and physiological measurement : an official journal of the Hospital Physicists' Association, Deutsche Gesellschaft fur Medizinische Physik and the European Federation of Organisations for Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1088/0143-0815/13/2/004","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 11

Abstract

The evaporative losses of solvent (water) from a commercial jet nebuliser (Unineb, Unimed Ltd) were estimated and found to be approximately 14% of total nebuliser output. A simple analysis of the nebulisation process then enabled the amount of the drug colomycin released as potentially respirable aerosol to be predicted from measurements of the total nebulisation time. Good agreement was found between the predicted (P) and measured (M) values such that P = 1.07 M (r2 = 0.98). The analysis also indicates that the proportion of the drug released as aerosol will increase as nebulisation proceeds provided nebuliser output is continuous and that for a given initial mass of drug a larger amount of drug is released as an aerosol if the volume of administration used in the nebuliser is as large as possible.

用喷射喷雾器蒸发药物溶剂损失的测量和预测。
据估计,商用喷射喷雾器(Unineb, Unimed Ltd)的溶剂(水)蒸发损失约占喷雾器总产量的14%。然后,对雾化过程进行简单的分析,就可以通过对雾化总时间的测量来预测作为潜在可吸入气溶胶释放的药物量。预测值(P)与实测值(M)吻合良好,P = 1.07 M (r2 = 0.98)。分析还表明,如果喷雾器的输出是连续的,那么作为气溶胶释放的药物比例将随着雾化的进行而增加,并且对于给定的初始质量的药物,如果喷雾器中使用的给药量尽可能大,则作为气溶胶释放的药物量就会更多。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信