Use of full automated microfluidic technique for determination of Zinc (II) in pharmaceutical preparations with 8-Hydroxyquinoline

Ihsan A. Mkashaf, Kamel H. Al-Sowdani, Abdulla J. Y. Aldarwish
{"title":"Use of full automated microfluidic technique for determination of Zinc (II) in pharmaceutical preparations with 8-Hydroxyquinoline","authors":"Ihsan A. Mkashaf, Kamel H. Al-Sowdani, Abdulla J. Y. Aldarwish","doi":"10.56714/bjrs.48.2.2","DOIUrl":null,"url":null,"abstract":"The full automated microfluidic system has been designed for the determination of zinc (II) ion in pharmaceutical samples home-made. Two-channel microchips (30 μl×4 cm) was designed in this study. The proposed system was controlled by using Arduino UNO and Mega microcontrollers. The first type was utilized to control the homemade micro-peristaltic pump to withdraw samples and chemical reagents in the microchip and then to the spectrophotometer equipment with 7-microliter flow cell. The other one type is Mega, was used as a data-logger to manipulate and recording the results as peak height corresponding the concentration by using Microsoft Excel 2016 program. The linearity was ranged 1-7 µg/ml, the correlation coefficient (R2) was 0.9998. The relative standard deviation for ten measurements of Zn(II) ion 4 µg/ml was (0.982%), as well as the detection limit was 0.125 µg/ml. The dilution factor of this system was 1.07.","PeriodicalId":377834,"journal":{"name":"Basrah Researches Sciences","volume":"1 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2022-12-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Basrah Researches Sciences","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.56714/bjrs.48.2.2","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The full automated microfluidic system has been designed for the determination of zinc (II) ion in pharmaceutical samples home-made. Two-channel microchips (30 μl×4 cm) was designed in this study. The proposed system was controlled by using Arduino UNO and Mega microcontrollers. The first type was utilized to control the homemade micro-peristaltic pump to withdraw samples and chemical reagents in the microchip and then to the spectrophotometer equipment with 7-microliter flow cell. The other one type is Mega, was used as a data-logger to manipulate and recording the results as peak height corresponding the concentration by using Microsoft Excel 2016 program. The linearity was ranged 1-7 µg/ml, the correlation coefficient (R2) was 0.9998. The relative standard deviation for ten measurements of Zn(II) ion 4 µg/ml was (0.982%), as well as the detection limit was 0.125 µg/ml. The dilution factor of this system was 1.07.
全自动微流控技术测定8-羟基喹啉制剂中锌的含量
设计了一种全自动微流控系统,用于国产药品样品中锌离子的测定。本研究设计了双通道微芯片(30 μl×4 cm)。该系统采用Arduino UNO和Mega微控制器进行控制。第一种是控制自制的微蠕动泵,将微芯片中的样品和化学试剂抽离,然后到7微升流式池的分光光度计设备。另一种是Mega,作为数据记录仪,使用Microsoft Excel 2016程序对结果进行处理并记录为浓度对应的峰高。线性范围为1 ~ 7µg/ml,相关系数(R2)为0.9998。Zn(II)离子4µg/ml 10次测定的相对标准偏差为0.982%,检出限为0.125µg/ml。该体系的稀释系数为1.07。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信