Correlation study between blood pressure and pulse transit time

T. Pereira, R. Sanches, P. Reis, J. Pêgo, Ricardo Simões
{"title":"Correlation study between blood pressure and pulse transit time","authors":"T. Pereira, R. Sanches, P. Reis, J. Pêgo, Ricardo Simões","doi":"10.1109/ENBENG.2015.7088864","DOIUrl":null,"url":null,"abstract":"Blood pressure (BP) determination is a fundamental parameter in cardiovascular assessment. The gold standard method to measure BP is based on the inflatable arm cuff, however has several disadvantages for continuous monitoring. New techniques were developed to overcome these limitations using correlations between the pulse transit time (PTT) and BP. This work draws attention to the PTT rationale using several methods. In order to determine the PTT, an electrocardiogram (ECG) was used combined with multiple photoplethysmography (PPG) sensors applied to different arm locations, these signals were acquired with a bioPLUX device. The Ultrassound system (SonoSite Edge) was used to measure the artery diameter. As reference, BP was measured using a cuff-based sphygmomanometric device. Measurements were performed in a study population of 36 volunteers. The correlation coefficient for DBP determined and DBP measured was r = 0,689. The results suggest PTT deduced from different locations can be used to measure BP.","PeriodicalId":285567,"journal":{"name":"2015 IEEE 4th Portuguese Meeting on Bioengineering (ENBENG)","volume":"23 4 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2015-04-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"4","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2015 IEEE 4th Portuguese Meeting on Bioengineering (ENBENG)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ENBENG.2015.7088864","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 4

Abstract

Blood pressure (BP) determination is a fundamental parameter in cardiovascular assessment. The gold standard method to measure BP is based on the inflatable arm cuff, however has several disadvantages for continuous monitoring. New techniques were developed to overcome these limitations using correlations between the pulse transit time (PTT) and BP. This work draws attention to the PTT rationale using several methods. In order to determine the PTT, an electrocardiogram (ECG) was used combined with multiple photoplethysmography (PPG) sensors applied to different arm locations, these signals were acquired with a bioPLUX device. The Ultrassound system (SonoSite Edge) was used to measure the artery diameter. As reference, BP was measured using a cuff-based sphygmomanometric device. Measurements were performed in a study population of 36 volunteers. The correlation coefficient for DBP determined and DBP measured was r = 0,689. The results suggest PTT deduced from different locations can be used to measure BP.
血压与脉搏传递时间的相关性研究
血压(BP)的测定是心血管疾病评估的基本参数。测量血压的金标准方法是基于充气臂袖带,然而对于连续监测有几个缺点。利用脉冲传递时间(PTT)和BP之间的相关性,开发了新的技术来克服这些限制。这项工作通过几种方法引起了人们对PTT理论基础的关注。为了确定PTT,使用心电图(ECG)和多个光电体积脉搏波(PPG)传感器应用于不同的手臂位置,这些信号由bioPLUX设备获取。超声系统(SonoSite Edge)测量动脉直径。作为参考,使用袖带式血压计测量血压。研究人员对36名志愿者进行了测量。测定DBP与测定DBP的相关系数r = 0.689。结果表明,从不同位置推导出的PTT可以用于测量BP。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信