A Method to Evaluate Power Sensor Calibration Systems

J. G. Burns
{"title":"A Method to Evaluate Power Sensor Calibration Systems","authors":"J. G. Burns","doi":"10.1109/ARFTG.2000.327448","DOIUrl":null,"url":null,"abstract":"This Power Sensor Calibration System used an indirect comparison method to calibrate a Device Under Test (DUT). A thermistor mount, calibrated by an external Calibration Laboratory, was the Standard Mount. The Effective Efficiencies and the associated accuracies of the Standard Mount were entered into a table the system computer. This system used a two step approach to perform the full measurement cycle. The first step was to measure the complex reflection coefficients of the DUT. The second step is shown in Figure 1. A power divider input was connected to a signal generator and amplifier. The Standard Mount was connected to one output of the divider. Another thermistor mount was connected to the other divider output.","PeriodicalId":166771,"journal":{"name":"56th ARFTG Conference Digest","volume":"746 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2000-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"56th ARFTG Conference Digest","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ARFTG.2000.327448","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

This Power Sensor Calibration System used an indirect comparison method to calibrate a Device Under Test (DUT). A thermistor mount, calibrated by an external Calibration Laboratory, was the Standard Mount. The Effective Efficiencies and the associated accuracies of the Standard Mount were entered into a table the system computer. This system used a two step approach to perform the full measurement cycle. The first step was to measure the complex reflection coefficients of the DUT. The second step is shown in Figure 1. A power divider input was connected to a signal generator and amplifier. The Standard Mount was connected to one output of the divider. Another thermistor mount was connected to the other divider output.
一种评估功率传感器校准系统的方法
该功率传感器校准系统采用间接比较法对被测设备(DUT)进行校准。由外部校准实验室校准的热敏电阻安装是标准安装。标准座的有效效率和相关精度被输入到系统计算机的表格中。该系统采用两步方法来完成整个测量周期。第一步是测量DUT的复反射系数。第二步如图1所示。功率分配器输入连接到信号发生器和放大器。标准安装连接到分压器的一个输出。另一个热敏电阻连接到另一个分压器输出。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信