Evaluation of three dry-sterilized hollow fiber artifical kidneys.

A Streifel, J Ebben, C Sahr, R Meyer, J Shideman, R Lynch, C Kjellstrand
{"title":"Evaluation of three dry-sterilized hollow fiber artifical kidneys.","authors":"A Streifel,&nbsp;J Ebben,&nbsp;C Sahr,&nbsp;R Meyer,&nbsp;J Shideman,&nbsp;R Lynch,&nbsp;C Kjellstrand","doi":"10.3109/08860227809079321","DOIUrl":null,"url":null,"abstract":"<p><p>We investigated three new dry sterilized hollow fiber artificial kidneys (HFAK) (Cordis Dow CDAK 1.3, Travenol CF 1200, Extracorporeal Tri-Ex 1). Dry sterilization makes these dialyzers more economical by shortening set-up time. Dry sterilization also eliminates iatrogenic administration of residual sterilant. Water of imbibition can significantly increase the blood compartment volume of the dialyzer during dialysis. Consequently, a corrected blood volume for each dialyzer was established; these corrected volumes varied from 13--36% greater than the volume determined before dialysate flow. With low dose heparinization of these dialyzers there was between an 18 and 45% decrease in the post dialysis volume, presumably due to fiber clotting during dialysis. This volume added to the residual blood loss measured by a colorimetric technique accounted for a possible blood loss ranging between 26.9 and 53.9 ml per dialysis for these three dialyzers. Our results suggest that a relationship between dialyzer clotting and decreased dialyzer efficiency may exist. These three capillary flow dialyzers had a much lower platelet drop (0--9% pre to post) when compared to 30--40% drop of flat plate dialyzers. These dialyzers were found to be safe and easy to use but the high fiber clotting warrants further investigation in chronic dialysis patients on high dose heparin.</p>","PeriodicalId":75998,"journal":{"name":"Journal of dialysis","volume":"2 4","pages":"347-59"},"PeriodicalIF":0.0000,"publicationDate":"1978-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.3109/08860227809079321","citationCount":"3","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of dialysis","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3109/08860227809079321","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 3

Abstract

We investigated three new dry sterilized hollow fiber artificial kidneys (HFAK) (Cordis Dow CDAK 1.3, Travenol CF 1200, Extracorporeal Tri-Ex 1). Dry sterilization makes these dialyzers more economical by shortening set-up time. Dry sterilization also eliminates iatrogenic administration of residual sterilant. Water of imbibition can significantly increase the blood compartment volume of the dialyzer during dialysis. Consequently, a corrected blood volume for each dialyzer was established; these corrected volumes varied from 13--36% greater than the volume determined before dialysate flow. With low dose heparinization of these dialyzers there was between an 18 and 45% decrease in the post dialysis volume, presumably due to fiber clotting during dialysis. This volume added to the residual blood loss measured by a colorimetric technique accounted for a possible blood loss ranging between 26.9 and 53.9 ml per dialysis for these three dialyzers. Our results suggest that a relationship between dialyzer clotting and decreased dialyzer efficiency may exist. These three capillary flow dialyzers had a much lower platelet drop (0--9% pre to post) when compared to 30--40% drop of flat plate dialyzers. These dialyzers were found to be safe and easy to use but the high fiber clotting warrants further investigation in chronic dialysis patients on high dose heparin.

三种干式灭菌中空纤维人工肾的评价。
我们研究了三种新型干式灭菌中空纤维人工肾(HFAK) (Cordis Dow CDAK 1.3, Travenol CF 1200, Extracorporeal Tri-Ex 1)。干式灭菌通过缩短设置时间使这些透析器更加经济。干式灭菌还消除了残余灭菌剂的医源性管理。渗吸水能显著增加透析器透析时血室容积。因此,建立了每个透析器的校正血容量;校正后的体积比透析液流动前测定的体积大13% ~ 36%。这些透析器的低剂量肝素化后,透析后体积减少18%至45%,可能是由于透析期间的纤维凝血。这一体积加上用比色技术测量的残留失血量,说明这三种透析器每次透析的可能失血量在26.9至53.9毫升之间。我们的结果表明,在透析器凝血和降低透析器效率之间可能存在关系。与平板透析器的血小板下降30- 40%相比,这三种毛细管流透析器的血小板下降(前后为0- 9%)要低得多。这些透析器安全且易于使用,但高纤维凝血在高剂量肝素慢性透析患者中的应用有待进一步研究。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信