{"title":"In vitro characterization of the RP-6 dialyzer: co-current and counter-current clearance as a function of ultrafiltration.","authors":"A Zelman, D Gisser","doi":"10.3109/08860227909063943","DOIUrl":null,"url":null,"abstract":"<p><p>Analytical and graphical techniques are described for presenting clearance data as a function of ultrafiltration; thus for the first time comparison of clearance values between dialyzers can be made unambiguously. This simple method uses a hand calculator to determine three coefficients. Clearance has been shown to be accurately described by the equation C = alpha 1 + alpha 2 Qv + alpha 3 (1/QBi). This formula permits construction of easily read families of curves which can predict clearance values under most clinical situations. The RP-6 dialyzer is characterized by these methods for clearance of urea, glucose and vitamin B 12 for counter-current and co-current modes.</p>","PeriodicalId":75998,"journal":{"name":"Journal of dialysis","volume":"3 2-3","pages":"237-50"},"PeriodicalIF":0.0000,"publicationDate":"1979-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.3109/08860227909063943","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of dialysis","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3109/08860227909063943","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 2
Abstract
Analytical and graphical techniques are described for presenting clearance data as a function of ultrafiltration; thus for the first time comparison of clearance values between dialyzers can be made unambiguously. This simple method uses a hand calculator to determine three coefficients. Clearance has been shown to be accurately described by the equation C = alpha 1 + alpha 2 Qv + alpha 3 (1/QBi). This formula permits construction of easily read families of curves which can predict clearance values under most clinical situations. The RP-6 dialyzer is characterized by these methods for clearance of urea, glucose and vitamin B 12 for counter-current and co-current modes.