[Transcutaneous measurement of PaO2 by the "TCM 1" electrode (author's transl)].

P Carthaus, I Podlesch
{"title":"[Transcutaneous measurement of PaO2 by the \"TCM 1\" electrode (author's transl)].","authors":"P Carthaus,&nbsp;I Podlesch","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Arterial oxygen pressures were compared with oxygen pressures measured transcutaneously with the aid of \"TCM 1\", an apparatus manufactured by Radiometer. At normal oxygen partial pressures the transcutaneously determined values for pO2 ran parallel with those measured intraarterially, but were on average 10.7% lower than the latter. In cases where oxygen partial pressures were above or below normal the figures did not agree. In these conditions, therefore, determinations in the arterial blood cannot be dispensed with. Halothane in concentrations of 0.3-0.5% and nitrous oxide (66%) in the inspired air did not affect the transcutaneous determination of pO2, but the method failed during surgical electrocoagulation.</p>","PeriodicalId":76342,"journal":{"name":"Praktische Anasthesie, Wiederbelebung und Intensivtherapie","volume":"14 4","pages":"345-7"},"PeriodicalIF":0.0000,"publicationDate":"1979-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Praktische Anasthesie, Wiederbelebung und Intensivtherapie","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Arterial oxygen pressures were compared with oxygen pressures measured transcutaneously with the aid of "TCM 1", an apparatus manufactured by Radiometer. At normal oxygen partial pressures the transcutaneously determined values for pO2 ran parallel with those measured intraarterially, but were on average 10.7% lower than the latter. In cases where oxygen partial pressures were above or below normal the figures did not agree. In these conditions, therefore, determinations in the arterial blood cannot be dispensed with. Halothane in concentrations of 0.3-0.5% and nitrous oxide (66%) in the inspired air did not affect the transcutaneous determination of pO2, but the method failed during surgical electrocoagulation.

[“TCM 1”电极经皮测定PaO2(作者译)]。
用Radiometer公司生产的“TCM 1”仪器与经皮测得的动脉血氧压进行比较。在正常氧分压下,经皮测定的pO2值与动脉内测量的值平行,但平均比后者低10.7%。在氧分压高于或低于正常的情况下,数据不一致。因此,在这种情况下,动脉血的测定是不能省略的。吸入空气中浓度为0.3-0.5%的氟烷和一氧化二氮(66%)对经皮测定pO2无影响,但手术电凝时该方法失败。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信