REVISITING COMPARABILITY OF THE SIZES OF CHANGES CORRESPONDING TO BREAST CANCER IN FILM AND DIGITAL MAMMOGRAPHY

D. Pasynkov, Sergei N. Merinov, E. A. Romanycheva, L. K. Mukhamatgaleeva, O. Pavlikova, M. N. Nasrullayev, A. L. Fedorov, Olga V. Busygina
{"title":"REVISITING COMPARABILITY OF THE SIZES OF CHANGES CORRESPONDING TO BREAST CANCER IN FILM AND DIGITAL MAMMOGRAPHY","authors":"D. Pasynkov, Sergei N. Merinov, E. A. Romanycheva, L. K. Mukhamatgaleeva, O. Pavlikova, M. N. Nasrullayev, A. L. Fedorov, Olga V. Busygina","doi":"10.47026/2413-4864-2023-2-23-31","DOIUrl":null,"url":null,"abstract":"In recent years, the traditional film technology of mammography has been replaced by digital technology, but comparative studies aimed at comparability of the size of changes when using these technologies are needed. The aim of the study was to assess the comparability of the size of changes corresponding to breast cancer in digital and film mammography, as well as the clinical significance of the dependencies identified. Material and methods. The study material was 27 sets of digital and film mammograms of patients with breast cancer, performed at intervals of no more than 2 months (digital mammograms – after film ones). The images were used to evaluate the same standardized neoplasm size. Results. The average size of changes corresponding to breast cancer on film mammograms was 16.96±8.44 mm, on digital ones – 15.63± 8.16 mm and 15.67±7.96 mm (p > 0.05). The range of variation in the difference in the neoplasm size according to film and digital mammography was 0-5 mm (0-30%). There were no cases of an increase in the neoplasm size on digital mammograms compared to film ones. In 37.04% (n = 10) of cases, the size of the changes in both types of mammography corresponded, in 29.63% (n = 8) understatement in digital mammography did not exceed 10%; in 25.93% (n = 7) of cases – 20%; in 7.40% (n = 2) – 30%. Underestimation of suspicious changes' sizes during digital mammography became clinically significant when their absolute size was less than 1 cm, when they became poorly visible against the background of fatty and fibrous-fatty parenchyma; as well as in situations when suspicious changes were indistinguishable from benign high-density zones. The use of a computer analysis system of mammograms in all cases ensured labeling changes corresponding to breast cancer. Conclusions. The sizes of changes corresponding to breast cancer (different from microcalcinates clusters, the size differences of which were not demonstrated) in film and digital mammography may differ by 0-30%, which does not allow using these techniques as interchangeable ones. Taking into account the possibility of underestimating the size of small changes, which makes them difficult to identify in the visual analysis of digital mammograms, it is advisable to use routinely a computer analysis system with subsequent targeted analysis of the labeled zones.","PeriodicalId":249759,"journal":{"name":"Acta medica Eurasica","volume":"27 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2023-06-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta medica Eurasica","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.47026/2413-4864-2023-2-23-31","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

In recent years, the traditional film technology of mammography has been replaced by digital technology, but comparative studies aimed at comparability of the size of changes when using these technologies are needed. The aim of the study was to assess the comparability of the size of changes corresponding to breast cancer in digital and film mammography, as well as the clinical significance of the dependencies identified. Material and methods. The study material was 27 sets of digital and film mammograms of patients with breast cancer, performed at intervals of no more than 2 months (digital mammograms – after film ones). The images were used to evaluate the same standardized neoplasm size. Results. The average size of changes corresponding to breast cancer on film mammograms was 16.96±8.44 mm, on digital ones – 15.63± 8.16 mm and 15.67±7.96 mm (p > 0.05). The range of variation in the difference in the neoplasm size according to film and digital mammography was 0-5 mm (0-30%). There were no cases of an increase in the neoplasm size on digital mammograms compared to film ones. In 37.04% (n = 10) of cases, the size of the changes in both types of mammography corresponded, in 29.63% (n = 8) understatement in digital mammography did not exceed 10%; in 25.93% (n = 7) of cases – 20%; in 7.40% (n = 2) – 30%. Underestimation of suspicious changes' sizes during digital mammography became clinically significant when their absolute size was less than 1 cm, when they became poorly visible against the background of fatty and fibrous-fatty parenchyma; as well as in situations when suspicious changes were indistinguishable from benign high-density zones. The use of a computer analysis system of mammograms in all cases ensured labeling changes corresponding to breast cancer. Conclusions. The sizes of changes corresponding to breast cancer (different from microcalcinates clusters, the size differences of which were not demonstrated) in film and digital mammography may differ by 0-30%, which does not allow using these techniques as interchangeable ones. Taking into account the possibility of underestimating the size of small changes, which makes them difficult to identify in the visual analysis of digital mammograms, it is advisable to use routinely a computer analysis system with subsequent targeted analysis of the labeled zones.
重新探讨乳腺x线摄影与数字x线摄影中相应乳腺癌变化大小的可比性
近年来,乳房x线摄影的传统胶片技术已被数字技术所取代,但需要针对这些技术使用时变化大小的可比性进行比较研究。该研究的目的是评估数字乳房x线照相术和胶片乳房x线照相术中乳腺癌相应变化大小的可比性,以及所确定的依赖性的临床意义。材料和方法。研究材料是27组乳腺癌患者的数码乳房x光片和胶片x光片,间隔不超过2个月(数码乳房x光片-在胶片x光片之后)。这些图像用于评估相同的标准化肿瘤大小。结果。乳房x线平片与乳腺癌对应的病灶大小分别为16.96±8.44 mm、15.63±8.16 mm和15.67±7.96 mm (p > 0.05)。根据胶片和数字乳房x线摄影,肿瘤大小差异的变化范围为0-5 mm(0-30%)。与胶片摄影相比,没有数字乳房x光片上肿瘤大小增加的病例。37.04% (n = 10)的病例中,两种类型乳房x线摄影的变化大小相对应,29.63% (n = 8)的病例中,数字乳房x线摄影的低报不超过10%;在25.93% (n = 7)的病例中- 20%;在7.40% (n = 2) - 30%。当其绝对尺寸小于1cm时,当它们在脂肪和纤维性脂肪实质背景下变得不明显时,数字乳房x线摄影中对可疑变化大小的低估就具有临床意义;以及在可疑变化与良性高密度区难以区分的情况下。在所有病例中,乳房x线照片的计算机分析系统的使用确保了与乳腺癌相对应的标签变化。结论。在胶片和数字乳房x线摄影中,与乳腺癌相对应的变化大小(不同于微煅烧团簇,其大小差异未被证明)可能相差0-30%,这使得这些技术不能互换使用。考虑到可能低估小变化的大小,这使得它们难以在数字乳房x线照片的视觉分析中识别,建议常规使用计算机分析系统,随后对标记区域进行针对性分析。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信