VALIDATION OF LEAD (Pb) ANALYSIS METHOD IN FISH MEAL MATRIX BY ELECTROTHERMAL ATOMIC ABSORPTION SPECTROPHOTOMETRY

Suwari, S. M. Ledoh, P. Rozari
{"title":"VALIDATION OF LEAD (Pb) ANALYSIS METHOD IN FISH MEAL MATRIX BY ELECTROTHERMAL ATOMIC ABSORPTION SPECTROPHOTOMETRY","authors":"Suwari, S. M. Ledoh, P. Rozari","doi":"10.29121/ijoest.v5.i5.2021.215","DOIUrl":null,"url":null,"abstract":"ABSTRACT \nValidation of analytical methods is the process of proving that the methods used are scientifically good and suitable for certain purposes. The aim of this study was to validate the analytical method of lead in fish meal matrix by electrothermal atomic absorption spectroscopy (ETAAS) using a phosphate modifier. The research begins with optimization of the analytical conditions that include pyrolysis (ashing) and atomizing temperature, and continues with the validation process by evaluating the analytical performance parameters using a certified reference material (CRM). Method validation was carried out by determining the level of accuracy and precision as well as lead content in a CRM Cod Muscle-422 from BCR. Each experiment was conducted at a wavelength of 283.3 nm, a lamp current of 9.0 amperes, a slit width of 1.3 nm and an argon carrier gas. The optimal pyrolysis and atomizing temperature of lead obtained by adding phosphate modifier was 650 oC and 2000 oC, respectively. The results obtained in this study indicate that the ETAAS lead analysis method using a phosphate modifier is feasible as a test method with an RSD of 3.79%, precision of 0.0043, relative error of 4.0%, accuracy of 104% and recovery of 98% ± 3%. The results of the validation method obtained that the 95% confidence interval for Pb content was 0.088 ± 0.023 µg/g, still within the range of Pb levels in the certificate, namely 0.085 ± 0.015 µg/g of Pb. \nKeywords: Validation, CRM, Cod Muscle, ETAAS, Lead","PeriodicalId":331301,"journal":{"name":"International Journal of Engineering Science Technologies","volume":"17 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2021-09-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Engineering Science Technologies","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.29121/ijoest.v5.i5.2021.215","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

ABSTRACT Validation of analytical methods is the process of proving that the methods used are scientifically good and suitable for certain purposes. The aim of this study was to validate the analytical method of lead in fish meal matrix by electrothermal atomic absorption spectroscopy (ETAAS) using a phosphate modifier. The research begins with optimization of the analytical conditions that include pyrolysis (ashing) and atomizing temperature, and continues with the validation process by evaluating the analytical performance parameters using a certified reference material (CRM). Method validation was carried out by determining the level of accuracy and precision as well as lead content in a CRM Cod Muscle-422 from BCR. Each experiment was conducted at a wavelength of 283.3 nm, a lamp current of 9.0 amperes, a slit width of 1.3 nm and an argon carrier gas. The optimal pyrolysis and atomizing temperature of lead obtained by adding phosphate modifier was 650 oC and 2000 oC, respectively. The results obtained in this study indicate that the ETAAS lead analysis method using a phosphate modifier is feasible as a test method with an RSD of 3.79%, precision of 0.0043, relative error of 4.0%, accuracy of 104% and recovery of 98% ± 3%. The results of the validation method obtained that the 95% confidence interval for Pb content was 0.088 ± 0.023 µg/g, still within the range of Pb levels in the certificate, namely 0.085 ± 0.015 µg/g of Pb. Keywords: Validation, CRM, Cod Muscle, ETAAS, Lead
电热原子吸收分光光度法测定鱼粉中铅的方法验证
分析方法的验证是证明所使用的方法在科学上是正确的,并且适合于某些目的的过程。研究了采用磷酸盐改性剂的电热原子吸收光谱法(ETAAS)分析鱼粉基质中铅的方法。研究从优化分析条件开始,包括热解(灰化)和雾化温度,并通过使用认证标准物质(CRM)评估分析性能参数,继续进行验证过程。通过测定BCR的CRM Cod Muscle-422的准确度和精密度以及铅含量,对方法进行了验证。每个实验在波长283.3 nm,灯电流9.0安培,狭缝宽度1.3 nm,氩气载气下进行。添加磷酸盐改性剂对铅的最佳热解温度为650℃,最佳雾化温度为2000℃。结果表明,采用磷酸修饰剂的ETAAS铅分析方法是可行的,RSD为3.79%,精密度为0.0043,相对误差为4.0%,准确度为104%,回收率为98%±3%。验证方法的结果表明,铅含量的95%置信区间为0.088±0.023µg/g,仍在证书规定的铅含量范围内,即0.085±0.015µg/g。关键词:验证,CRM, Cod Muscle, ETAAS, Lead
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信