{"title":"A wireless passive sensor for pH monitoring employing temperature compensation","authors":"S. Bhadra, Greg E. Bridges, Douglas J Thomson, Michael S Freund","doi":"10.1109/ICSENS.2011.6127181","DOIUrl":null,"url":null,"abstract":"We present an integrated coupled-coil passive sensor for temperature compensated remote pH monitoring. The sensor is a RLC resonant circuit consisting of an inductive coil connected in parallel to a thermistor, a varactor and a pH combination electrode. pH variation of the solution changes the voltage-dependent capacitance and shifts the resonant frequency of the sensor, while temperature change of the solution affects the resistance and changes the quality factor. The resonant frequency and quality factor of the sensor are tracked remotely by an interrogator coil coupled to the sensor coil. A temperature range of 25°C–60°C and pH over a 2–12 dynamic range can be measured simultaneously. With temperature compensation, the accuracy of the pH measurement is demonstrated to be 0.1 pH. The sensor overcomes the pH measurement error due to the temperature dependence of electrode based passive pH sensors.","PeriodicalId":201386,"journal":{"name":"2011 IEEE SENSORS Proceedings","volume":"28 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2011-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"7","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2011 IEEE SENSORS Proceedings","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ICSENS.2011.6127181","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 7

Abstract

We present an integrated coupled-coil passive sensor for temperature compensated remote pH monitoring. The sensor is a RLC resonant circuit consisting of an inductive coil connected in parallel to a thermistor, a varactor and a pH combination electrode. pH variation of the solution changes the voltage-dependent capacitance and shifts the resonant frequency of the sensor, while temperature change of the solution affects the resistance and changes the quality factor. The resonant frequency and quality factor of the sensor are tracked remotely by an interrogator coil coupled to the sensor coil. A temperature range of 25°C–60°C and pH over a 2–12 dynamic range can be measured simultaneously. With temperature compensation, the accuracy of the pH measurement is demonstrated to be 0.1 pH. The sensor overcomes the pH measurement error due to the temperature dependence of electrode based passive pH sensors.
一种采用温度补偿的用于pH监测的无线无源传感器
我们提出了一种集成的耦合线圈无源传感器,用于温度补偿的远程pH监测。该传感器是一个RLC谐振电路,由一个电感线圈与一个热敏电阻、一个变容管和一个pH组合电极并联组成。溶液的pH变化会改变电压依赖性电容,使传感器的谐振频率发生移位,而溶液的温度变化会影响电阻,改变品质因子。传感器的谐振频率和质量因数由耦合在传感器线圈上的询问线圈远程跟踪。温度范围为25°C - 60°C, pH值在2-12动态范围内可以同时测量。通过温度补偿,pH测量精度可达0.1 pH。该传感器克服了电极无源pH传感器由于温度依赖而导致的pH测量误差。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信