The Direction of Flow and Phase-encoding Schemes Effects on Signal Intensity in T1-weighted Inversion Recovery TurboFLASH Images.

Q3 Medicine
Mahmood Nazarpoor
{"title":"The Direction of Flow and Phase-encoding Schemes Effects on Signal Intensity in T1-weighted Inversion Recovery TurboFLASH Images.","authors":"Mahmood Nazarpoor","doi":"10.31661/jbpe.v0i0.2203-1468","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>It is needed to minimize the effect of flow direction on the desired area, such as arterial input function (AIF) in magnetic resonance imaging (MRI).</p><p><strong>Objective: </strong>The current study aimed to investigate the effect of flow direction on different velocities (0-80.39 cm/s) for the strength of the signal intensity (SI) at the linear phase-encoding (LPE) and the center out phase-encoding (COPE) schemes and to recommend the best flow direction in a selected slice and scheme for absolute perfusion measurement by inversion recovery T1-weighted turbo fast low-angle shot (TurboFLASH) MR images.</p><p><strong>Material and methods: </strong>In this experimental study, the flow rates were measured using a flow phantom, and the signal intensity (SI) was measured at the two opposite flow directions in the Z-axis perpendicular to the coronal image at a concentration of 0.8 mmol/L of gadolinium-diethylenetriaminepantaacetic acid (Gd-DTPA) by using the LPE and COPE schemes.</p><p><strong>Results: </strong>The increase in velocity along with the growth in SI and inflow affected the use of LPE and COPE acquisitions in both directions. The velocity of the arterial input function is needed to calculate the inflow correction factor by using two schemes in two opposite flow directions to investigate perfusion.</p><p><strong>Conclusion: </strong>The COPE scheme was better than the LPE scheme in measuring perfusion since the velocity and direction of blood flow affect SI less.</p>","PeriodicalId":38035,"journal":{"name":"Journal of Biomedical Physics and Engineering","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2023-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/f6/c1/JBPE-13-45.PMC9923244.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Biomedical Physics and Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.31661/jbpe.v0i0.2203-1468","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Background: It is needed to minimize the effect of flow direction on the desired area, such as arterial input function (AIF) in magnetic resonance imaging (MRI).

Objective: The current study aimed to investigate the effect of flow direction on different velocities (0-80.39 cm/s) for the strength of the signal intensity (SI) at the linear phase-encoding (LPE) and the center out phase-encoding (COPE) schemes and to recommend the best flow direction in a selected slice and scheme for absolute perfusion measurement by inversion recovery T1-weighted turbo fast low-angle shot (TurboFLASH) MR images.

Material and methods: In this experimental study, the flow rates were measured using a flow phantom, and the signal intensity (SI) was measured at the two opposite flow directions in the Z-axis perpendicular to the coronal image at a concentration of 0.8 mmol/L of gadolinium-diethylenetriaminepantaacetic acid (Gd-DTPA) by using the LPE and COPE schemes.

Results: The increase in velocity along with the growth in SI and inflow affected the use of LPE and COPE acquisitions in both directions. The velocity of the arterial input function is needed to calculate the inflow correction factor by using two schemes in two opposite flow directions to investigate perfusion.

Conclusion: The COPE scheme was better than the LPE scheme in measuring perfusion since the velocity and direction of blood flow affect SI less.

Abstract Image

Abstract Image

Abstract Image

流方向和相位编码对t1加权反演恢复图像信号强度的影响。
背景:需要将血流方向对所需区域的影响最小化,例如磁共振成像(MRI)中的动脉输入功能(AIF)。目的:本研究旨在探讨不同流速(0-80.39 cm/s)下线性相位编码(LPE)和中心向外相位编码(COPE)方案下信号强度(SI)的方向影响,并为t1加权turbo快速低角度拍摄(TurboFLASH) MR图像反演恢复推荐最佳的流方向和绝对灌注测量方案。材料与方法:在本实验研究中,采用流动模体测量流速,并采用LPE和COPE方案,在浓度为0.8 mmol/L的钆-二乙烯三胺乙酸(Gd-DTPA)浓度下,在垂直于冠状像的z轴上测量两个相反流动方向上的信号强度(SI)。结果:随着SI和流入的增加,流速的增加在两个方向上影响了LPE和COPE的使用。采用两种方案在两个相反的流动方向上研究灌注情况,需要动脉输入函数的速度来计算流入校正因子。结论:COPE方案比LPE方案更适合于测量血流灌注,血流速度和方向对SI的影响较小。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Biomedical Physics and Engineering
Journal of Biomedical Physics and Engineering Medicine-Radiology, Nuclear Medicine and Imaging
CiteScore
2.90
自引率
0.00%
发文量
64
审稿时长
10 weeks
期刊介绍: The Journal of Biomedical Physics and Engineering (JBPE) is a bimonthly peer-reviewed English-language journal that publishes high-quality basic sciences and clinical research (experimental or theoretical) broadly concerned with the relationship of physics to medicine and engineering.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信