Pre-analytical conditions influencing analysis of folate in dried plasma microsamples

IF 3.1 4区 医学 Q2 MEDICAL LABORATORY TECHNOLOGY
Christopher M. Shuford, Evan W. McConnell, Stacy Dee, Russell P. Grant
{"title":"Pre-analytical conditions influencing analysis of folate in dried plasma microsamples","authors":"Christopher M. Shuford,&nbsp;Evan W. McConnell,&nbsp;Stacy Dee,&nbsp;Russell P. Grant","doi":"10.1016/j.jmsacl.2023.01.003","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><p>Determination of folate insufficiency is of considerable interest given its importance in fetal development and red blood cell formation; however, access to blood tests may be limited due to the requirement for phlebotomy as well as controlled temperature shipping of blood specimens to laboratories for testing due to the inherent instability of folate and its vitamers.</p></div><div><h3>Methods</h3><p>An LC-MS/MS test was developed and validated for the measurement of 5-methyltetrahydrofolate (5MTHF) in dried plasma specimens collected from fingerstick blood using a laminar flow blood separation device, as well as liquid venous plasma for comparison. Two pre-analytical factors investigated influencing the measurement of 5MTHF in dried plasma were hemolysis of the fingerstick blood during collection and storage/shipment of the dried plasma.</p></div><div><h3>Results</h3><p>Although observed infrequently, hemolysis &gt;10 % resulted in elevated 5MTHF measurements, but hemolysis &gt;1 % resulted in elevated chloride measurements, which were necessary to normalize 5MTHF measurements for variation in volume of dried plasma specimens. Stability of 5MTHF was improved in dried plasma relative to liquid plasma at ambient temperatures, but not sufficiently to allow for uncontrolled temperature shipping despite controlling for humidity and light exposure. Shipping studies emulating ISTA procedure 7D were conducted with a reusable cold packaging solution. The packaging failed to stabilize 5MTHF in dried plasma specimens during a 2-day summer shipping evaluation, but did provide sufficient temperature control to stabilize 5MTHF during the overnight shipping evaluation.</p></div><div><h3>Conclusion</h3><p>Our studies provide boundary conditions with respect to hemolysis, storage, and shipping for successful analysis of 5MTHF from dried plasma specimens.</p></div>","PeriodicalId":52406,"journal":{"name":"Journal of Mass Spectrometry and Advances in the Clinical Lab","volume":"28 ","pages":"Pages 1-8"},"PeriodicalIF":3.1000,"publicationDate":"2023-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/a6/97/main.PMC9894916.pdf","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Mass Spectrometry and Advances in the Clinical Lab","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2667145X23000032","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 1

Abstract

Introduction

Determination of folate insufficiency is of considerable interest given its importance in fetal development and red blood cell formation; however, access to blood tests may be limited due to the requirement for phlebotomy as well as controlled temperature shipping of blood specimens to laboratories for testing due to the inherent instability of folate and its vitamers.

Methods

An LC-MS/MS test was developed and validated for the measurement of 5-methyltetrahydrofolate (5MTHF) in dried plasma specimens collected from fingerstick blood using a laminar flow blood separation device, as well as liquid venous plasma for comparison. Two pre-analytical factors investigated influencing the measurement of 5MTHF in dried plasma were hemolysis of the fingerstick blood during collection and storage/shipment of the dried plasma.

Results

Although observed infrequently, hemolysis >10 % resulted in elevated 5MTHF measurements, but hemolysis >1 % resulted in elevated chloride measurements, which were necessary to normalize 5MTHF measurements for variation in volume of dried plasma specimens. Stability of 5MTHF was improved in dried plasma relative to liquid plasma at ambient temperatures, but not sufficiently to allow for uncontrolled temperature shipping despite controlling for humidity and light exposure. Shipping studies emulating ISTA procedure 7D were conducted with a reusable cold packaging solution. The packaging failed to stabilize 5MTHF in dried plasma specimens during a 2-day summer shipping evaluation, but did provide sufficient temperature control to stabilize 5MTHF during the overnight shipping evaluation.

Conclusion

Our studies provide boundary conditions with respect to hemolysis, storage, and shipping for successful analysis of 5MTHF from dried plasma specimens.

Abstract Image

Abstract Image

Abstract Image

影响干燥血浆微量样品中叶酸分析的分析前条件
叶酸缺乏的测定在胎儿发育和红细胞形成中具有重要意义,因此引起了人们的极大兴趣;然而,由于叶酸及其维生素的固有不稳定性,需要静脉切开术以及将血液样本运送到实验室进行检测的温度控制,因此血液检测的机会可能会受到限制。方法采用液相色谱-质谱联用技术(LC-MS/MS)测定指尖血液干燥血浆样品中5-甲基四氢叶酸(5MTHF)的含量,并与液体静脉血浆进行比较。研究的影响干血浆中5MTHF测量的两个预分析因素是干血浆收集和储存/运输过程中手指棒血的溶血。结果尽管观察不多,但溶血>;10%导致5MTHF测量值升高,但溶血>;1%导致氯化物测量值升高,这对于使5MTHF测量值标准化以测量干燥血浆样品的体积变化是必要的。相对于环境温度下的液体等离子体,5MTHF在干燥等离子体中的稳定性得到了改善,但不足以允许不受控制的温度运输,尽管控制了湿度和光暴露。使用可重复使用的冷包装解决方案进行了模拟ISTA程序7D的装运研究。在为期两天的夏季装运评估中,包装未能稳定干燥血浆样品中的5MTHF,但在隔夜装运评估中提供了足够的温度控制以稳定5MTHF。结论我们的研究为成功分析干燥血浆样品中的5MTHF提供了溶血、储存和运输的边界条件。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Mass Spectrometry and Advances in the Clinical Lab
Journal of Mass Spectrometry and Advances in the Clinical Lab Health Professions-Medical Laboratory Technology
CiteScore
4.30
自引率
18.20%
发文量
41
审稿时长
81 days
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信