Jakob Lüker, Marc Strik, Jason G Andrade, Alexandre Raymond-Paquin, Mohamed Hassan Elrefai, Paul R Roberts, Óscar Cano Pérez, Jordana Kron, Jayanthi Koneru, Hilton Franqui-Rivera, Arian Sultan, Angela Ernst, Jörn Schmitt, Alexander Pott, Christian Veltmann, Neil T Srinivasan, Jason Collinson, Antonius M W van Stipdonk, Dominik Linz, Nina Fluschnik, Tobias Tönnis, Andreas Haeberlin, Sylvain Ploux, Daniel Steven
{"title":"Incidence of premature battery depletion in subcutaneous cardioverter-defibrillator patients: insights from a multicenter registry.","authors":"Jakob Lüker, Marc Strik, Jason G Andrade, Alexandre Raymond-Paquin, Mohamed Hassan Elrefai, Paul R Roberts, Óscar Cano Pérez, Jordana Kron, Jayanthi Koneru, Hilton Franqui-Rivera, Arian Sultan, Angela Ernst, Jörn Schmitt, Alexander Pott, Christian Veltmann, Neil T Srinivasan, Jason Collinson, Antonius M W van Stipdonk, Dominik Linz, Nina Fluschnik, Tobias Tönnis, Andreas Haeberlin, Sylvain Ploux, Daniel Steven","doi":"10.1007/s10840-023-01468-1","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The subcutaneous ICD established its role in the prevention of sudden cardiac death in recent years. The occurrence of premature battery depletion in a large subset of potentially affected devices has been a cause of concern. The incidence of premature battery depletion has not been studied systematically beyond manufacturer-reported data.</p><p><strong>Methods: </strong>Retrospective data and the most recent follow-up data on S-ICD devices from fourteen centers in Europe, the US, and Canada was studied. The incidence of generator removal or failure was reported to investigate the incidence of premature S-ICD battery depletion, defined as battery failure within 60 months or less.</p><p><strong>Results: </strong>Data from 1054 devices was analyzed. Premature battery depletion occurred in 3.5% of potentially affected devices over an observation period of 49 months.</p><p><strong>Conclusions: </strong>The incidence of premature battery depletion of S-ICD potentially affected by a battery advisory was around 3.5% after 4 years in this study. Premature depletion occurred exclusively in devices under advisory. This is in line with the most recently published reports from the manufacturer.</p><p><strong>Trial registration: </strong>ClinicalTrials.gov Identifier: NCT04767516 .</p>","PeriodicalId":16202,"journal":{"name":"Journal of Interventional Cardiac Electrophysiology","volume":" ","pages":"1409-1415"},"PeriodicalIF":2.6000,"publicationDate":"2025-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12436463/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Interventional Cardiac Electrophysiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s10840-023-01468-1","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/1/18 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Background: The subcutaneous ICD established its role in the prevention of sudden cardiac death in recent years. The occurrence of premature battery depletion in a large subset of potentially affected devices has been a cause of concern. The incidence of premature battery depletion has not been studied systematically beyond manufacturer-reported data.
Methods: Retrospective data and the most recent follow-up data on S-ICD devices from fourteen centers in Europe, the US, and Canada was studied. The incidence of generator removal or failure was reported to investigate the incidence of premature S-ICD battery depletion, defined as battery failure within 60 months or less.
Results: Data from 1054 devices was analyzed. Premature battery depletion occurred in 3.5% of potentially affected devices over an observation period of 49 months.
Conclusions: The incidence of premature battery depletion of S-ICD potentially affected by a battery advisory was around 3.5% after 4 years in this study. Premature depletion occurred exclusively in devices under advisory. This is in line with the most recently published reports from the manufacturer.
期刊介绍:
The Journal of Interventional Cardiac Electrophysiology is an international publication devoted to fostering research in and development of interventional techniques and therapies for the management of cardiac arrhythmias. It is designed primarily to present original research studies and scholarly scientific reviews of basic and applied science and clinical research in this field. The Journal will adopt a multidisciplinary approach to link physical, experimental, and clinical sciences as applied to the development of and practice in interventional electrophysiology. The Journal will examine techniques ranging from molecular, chemical and pharmacologic therapies to device and ablation technology. Accordingly, original research in clinical, epidemiologic and basic science arenas will be considered for publication. Applied engineering or physical science studies pertaining to interventional electrophysiology will be encouraged. The Journal is committed to providing comprehensive and detailed treatment of major interventional therapies and innovative techniques in a structured and clinically relevant manner. It is directed at clinical practitioners and investigators in the rapidly growing field of interventional electrophysiology. The editorial staff and board reflect this bias and include noted international experts in this area with a wealth of expertise in basic and clinical investigation. Peer review of all submissions, conflict of interest guidelines and periodic editorial board review of all Journal policies have been established.