Comparison of an open view autorefractor with an open view aberrometer in determining peripheral refraction in children

IF 2.2 Q2 OPHTHALMOLOGY
Pelsin Demir , Antonio Filipe Macedo , Ranjay Chakraborty , Karthikeyan Baskaran
{"title":"Comparison of an open view autorefractor with an open view aberrometer in determining peripheral refraction in children","authors":"Pelsin Demir ,&nbsp;Antonio Filipe Macedo ,&nbsp;Ranjay Chakraborty ,&nbsp;Karthikeyan Baskaran","doi":"10.1016/j.optom.2021.12.002","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><p>The aim of this study was to compare central and peripheral refraction using an open view Shin-Nippon NVision-K 5001 autorefractor and an open view COAS-HD VR aberrometer in young children.</p></div><div><h3>Methods</h3><p>Cycloplegic central and peripheral autorefraction was measured in the right eye of 123 children aged 8 to 16 years. Three measurements each were obtained with both Shin-Nippon NVision-K 5001 autorefractor and COAS-HD VR aberrometer along the horizontal visual field up to 30° (nasal and temporal) in 10° steps. The refraction from the autorefractor was compared with aberrometer refraction for pupil analysis diameters of 2.5-mm and 5.0-mm.</p></div><div><h3>Results</h3><p>The Shin-Nippon was 0.30 D more hyperopic than COAS-HD VR at 2.5-mm pupil and 0.50 D more hyperopic than COAS-HD VR at 5-mm pupil for central refraction. For both pupil sizes, the 95% limits of agreement were approximately 0.50 D for central refraction, and limits were wider in the nasal visual field compared to the temporal visual field. The mean difference for both J<sub>0</sub> and J<sub>45</sub> were within 0.15 D and the 95% limits of agreement within 0.90 D across the horizontal visual field.</p></div><div><h3>Conclusion</h3><p>Defocus components were similar between the Shin-Nippon autorefractor and the COAS-HD VR aberrometer with a 2.5-mm pupil for most visual field angles. However, there was a significant difference in defocus component between the Shin-Nippon autorefractor and the COAS-HD VR aberrometer with a 5.0-mm pupil, wherein the autorefractor measured more hyperopia. The astigmatic components J<sub>0</sub> and J<sub>45</sub> were similar between instruments for both central and peripheral refraction.</p></div>","PeriodicalId":46407,"journal":{"name":"Journal of Optometry","volume":"16 1","pages":"Pages 20-29"},"PeriodicalIF":2.2000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9811364/pdf/","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Optometry","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S188842962100090X","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 2

Abstract

Purpose

The aim of this study was to compare central and peripheral refraction using an open view Shin-Nippon NVision-K 5001 autorefractor and an open view COAS-HD VR aberrometer in young children.

Methods

Cycloplegic central and peripheral autorefraction was measured in the right eye of 123 children aged 8 to 16 years. Three measurements each were obtained with both Shin-Nippon NVision-K 5001 autorefractor and COAS-HD VR aberrometer along the horizontal visual field up to 30° (nasal and temporal) in 10° steps. The refraction from the autorefractor was compared with aberrometer refraction for pupil analysis diameters of 2.5-mm and 5.0-mm.

Results

The Shin-Nippon was 0.30 D more hyperopic than COAS-HD VR at 2.5-mm pupil and 0.50 D more hyperopic than COAS-HD VR at 5-mm pupil for central refraction. For both pupil sizes, the 95% limits of agreement were approximately 0.50 D for central refraction, and limits were wider in the nasal visual field compared to the temporal visual field. The mean difference for both J0 and J45 were within 0.15 D and the 95% limits of agreement within 0.90 D across the horizontal visual field.

Conclusion

Defocus components were similar between the Shin-Nippon autorefractor and the COAS-HD VR aberrometer with a 2.5-mm pupil for most visual field angles. However, there was a significant difference in defocus component between the Shin-Nippon autorefractor and the COAS-HD VR aberrometer with a 5.0-mm pupil, wherein the autorefractor measured more hyperopia. The astigmatic components J0 and J45 were similar between instruments for both central and peripheral refraction.

Abstract Image

Abstract Image

Abstract Image

开放视野自动折射仪与开放视野像差仪测定儿童外周屈光度的比较
目的本研究的目的是使用新日本NVision-K 5001自动验光仪和COAS-HD VR像差仪对幼儿的中心和外周屈光进行比较。方法对123例8~16岁儿童的右眼进行了中央和外周自行屈光测量。使用新日本NVision-K 5001自动折射仪和COAS-HD VR像差仪,沿水平视野以10°的步长分别获得三个测量值,最高可达30°(鼻腔和颞部)。在瞳孔分析直径为2.5毫米和5.0毫米的情况下,将自动折射器的折射与像差仪的折射进行比较。结果在中心折射方面,新日本在2.5毫米瞳孔处的远视度比COAS-HD VR高0.30 D,在5毫米瞳孔处比COAS-HD VR高0.50 D。对于两种瞳孔大小,中心屈光的95%一致性极限约为0.50 D,与颞视野相比,鼻视野的极限更宽。J0和J45的平均差异在0.15D以内,95%的一致性范围在0.90D以内。结论在大多数视场角下,新日本自动折射仪和瞳孔为2.5 mm的COAS-HD VR像差仪的散焦成分相似。然而,新日本自动折射器和具有5.0mm瞳孔的COAS-HD VR像差仪之间的散焦成分存在显著差异,其中自动折射器测量到更多的远视。散光分量J0和J45在用于中心折射和外围折射的仪器之间是相似的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Optometry
Journal of Optometry OPHTHALMOLOGY-
CiteScore
5.20
自引率
0.00%
发文量
60
审稿时长
66 days
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信