{"title":"The Tandem Control-IQ advanced hybrid system improves glycemic control in children under 18 years of age with type 1 diabetes and night rest in caregivers","authors":"Andrés Mingorance Delgado ,&nbsp;Fernando Lucas","doi":"10.1016/j.endien.2023.08.005","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><p><span>To determine the impact of switching from the predictive low glucose suspend (PLGS) system to the advanced hybrid Tandem Control-IQ system on glucometrics and glycosylated haemoglobin (HbA1c) at one year. To assess the impact on the </span>quality of life perceived by parents.</p></div><div><h3>Method</h3><p><span>Prospective study in 71 patients aged 6–18 years with type 1 diabetes<span> (DM1), in treatment with PLGS, who switched to an advanced hybrid system. Glucometric data were collected before the change, at 4 and 8 weeks, and at one year of use; </span></span>HbA1c before the change and after one year. The Diabetes Impact and Devices Satisfaction (DIDS) questionnaire was used at weeks 4 and 8.</p></div><div><h3>Results</h3><p>An increase in time in range (TIR) was observed with a median of 76% (<em>P</em> <!-->&lt;<!--> <span>.001) at 4 weeks, which was maintained after one year (+8% in the total group). Overall, 73.24% of patients achieved a TIR above 70%. The subgroup with an initial TIR of less than 56% increased it by 14.4%. After one year there was a 0.3% reduction in HbA1c. Level 1 hypoglycaemia, level 1 and level 2 hyperglycaemia, mean glucose (GM) and coefficient of variation (CV) decreased.</span></p><p>Auto mode stayed on 97% of the time and no dropouts occurred.</p><p>Caregivers had a perception of better glycaemic control and less need to monitor blood glucose variations during the night. None of them would switch back to the previous system and they feel safe with the new system.</p></div><div><h3>Conclusions</h3><p>The Tandem Control-IQ advanced hybrid system was shown to be effective one year after its implementation with improvement in all glucometric parameters and HbA1c, as well as night-time rest in caregivers.</p></div>","PeriodicalId":48650,"journal":{"name":"Endocrinologia Diabetes Y Nutricion","volume":"70 ","pages":"Pages 27-35"},"PeriodicalIF":1.8000,"publicationDate":"2023-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Endocrinologia Diabetes Y Nutricion","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2530018023001245","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENDOCRINOLOGY & METABOLISM","Score":null,"Total":0}
引用次数: 0

Abstract

Objective

To determine the impact of switching from the predictive low glucose suspend (PLGS) system to the advanced hybrid Tandem Control-IQ system on glucometrics and glycosylated haemoglobin (HbA1c) at one year. To assess the impact on the quality of life perceived by parents.

Method

Prospective study in 71 patients aged 6–18 years with type 1 diabetes (DM1), in treatment with PLGS, who switched to an advanced hybrid system. Glucometric data were collected before the change, at 4 and 8 weeks, and at one year of use; HbA1c before the change and after one year. The Diabetes Impact and Devices Satisfaction (DIDS) questionnaire was used at weeks 4 and 8.

Results

An increase in time in range (TIR) was observed with a median of 76% (P < .001) at 4 weeks, which was maintained after one year (+8% in the total group). Overall, 73.24% of patients achieved a TIR above 70%. The subgroup with an initial TIR of less than 56% increased it by 14.4%. After one year there was a 0.3% reduction in HbA1c. Level 1 hypoglycaemia, level 1 and level 2 hyperglycaemia, mean glucose (GM) and coefficient of variation (CV) decreased.

Auto mode stayed on 97% of the time and no dropouts occurred.

Caregivers had a perception of better glycaemic control and less need to monitor blood glucose variations during the night. None of them would switch back to the previous system and they feel safe with the new system.

Conclusions

The Tandem Control-IQ advanced hybrid system was shown to be effective one year after its implementation with improvement in all glucometric parameters and HbA1c, as well as night-time rest in caregivers.

Tandem control - iq先进混合系统改善了18岁以下1型糖尿病儿童的血糖控制和护理人员的夜间休息
目的确定从预测性低血糖悬浮液(PLGS)系统切换到高级混合串联控制IQ系统一年后对血糖指标和糖化血红蛋白(HbA1c)的影响。评估父母对生活质量的影响。方法对71名6-18岁的1型糖尿病(DM1)患者进行前瞻性研究,这些患者接受PLGS治疗,并改用先进的混合系统。在改变前、使用4周和8周以及使用一年时收集血糖测量数据;改变前和一年后的HbA1c。在第4周和第8周使用糖尿病影响和设备满意度(DIDS)问卷。结果观察到范围内时间(TIR)增加,4周时中位数为76%(P<;.001),一年后保持不变(总组为+8%)。总体而言,73.24%的患者的TIR达到70%以上。初始TIR小于56%的亚组增加了14.4%。一年后,HbA1c减少了0.3%。1级低血糖、1级和2级高血糖、平均血糖(GM)和变异系数(CV)下降。“自动”模式在97%的时间内保持打开状态,并且没有出现中途退出的情况。护理人员认为血糖控制更好,夜间不需要监测血糖变化。他们中没有一个人会切换回以前的系统,他们对新系统感到安全。结论串联控制IQ高级混合系统在实施一年后被证明是有效的,其所有血糖测量参数和HbA1c都得到了改善,护理人员夜间休息也得到了改善。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
2.10
自引率
10.50%
发文量
99
期刊介绍: Endocrinología, Diabetes y Nutrición is the official journal of the Spanish Society of Endocrinology and Nutrition (Sociedad Española de Endocrinología y Nutrición, SEEN) and the Spanish Society of Diabetes (Sociedad Española de Diabetes, SED), and was founded in 1954. The aim of the journal is to improve knowledge and be a useful tool in practice for clinical and laboratory specialists, trainee physicians, researchers, and nurses interested in endocrinology, diabetes, nutrition and related disciplines. It is an international journal published in Spanish (print and online) and English (online), covering different fields of endocrinology and metabolism, including diabetes, obesity, and nutrition disorders, as well as the most relevant research produced mainly in Spanish language territories. The quality of the contents is ensured by a prestigious national and international board, and by a selected panel of specialists involved in a rigorous peer review. The result is that only manuscripts containing high quality research and with utmost interest for clinicians and professionals related in the field are published. The Journal publishes Original clinical and research articles, Reviews, Special articles, Clinical Guidelines, Position Statements from both societies and Letters to the editor. Endocrinología, Diabetes y Nutrición can be found at Science Citation Index Expanded, Medline/PubMed and SCOPUS.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信