前路种植手术中术中测量指南与常规手术指南的准确性:回顾性病例对照研究

IF 1.7 4区 医学 Q3 DENTISTRY, ORAL SURGERY & MEDICINE
Yingkai Wang, Chenyang Xie, Tinglu Fang, Zhaozhao Chen, Haiyang Yu
{"title":"前路种植手术中术中测量指南与常规手术指南的准确性:回顾性病例对照研究","authors":"Yingkai Wang,&nbsp;Chenyang Xie,&nbsp;Tinglu Fang,&nbsp;Zhaozhao Chen,&nbsp;Haiyang Yu","doi":"10.11607/jomi.9898","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To verify a novel method that improves the accuracy of static computer-aided implant surgery (sCAIS) through intraoperative measurement.</p><p><strong>Materials and methods: </strong>Forty-seven patients were selected for this study, each with a missing tooth or a tooth that required extraction from the anterior area. The patients were divided into the intraoperative measuring guide (MG) and conventional guide (CG) groups. Following the preoperative implant planning, the surgical guides were designed and fabricated. In the MG group, the drill was guided by double-armed zirconia sleeves, and the axial direction of the drill was assessed using the indicator components. The implant was guided using a resin guide tube. In the CG group, the drills were guided using a metal sleeve and handles, and the implants were placed with the guidance of the metal sleeve only. The angular and linear deviations at the entry and apex between the planned and actual implant positions were measured after matching the preoperative and postoperative CBCT data. The independent-samples t test was used to compare the deviation between the MG and CG groups.</p><p><strong>Results: </strong>The 3D deviations for the MG group at the entry and apex were 0.67 ± 0.44 mm and 0.93 ± 0.40 mm, respectively. The angular deviation was 2.27 ± 0.96 degrees. Statistical differences were found in the 3D deviation at the entry point and apical position between the MG and CG groups, yielding relatively smaller deviations in the MG group.</p><p><strong>Conclusion: </strong>The use of an intraoperative measuring guide could improve the accuracy of implant placement in sCAIS.</p>","PeriodicalId":50298,"journal":{"name":"International Journal of Oral & Maxillofacial Implants","volume":"38 2","pages":"277-286"},"PeriodicalIF":1.7000,"publicationDate":"2023-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Accuracy of Intraoperative Measuring Guide and Conventional Surgical Guide in Anterior Implant Surgery: A Retrospective Case-Control Study\",\"authors\":\"Yingkai Wang,&nbsp;Chenyang Xie,&nbsp;Tinglu Fang,&nbsp;Zhaozhao Chen,&nbsp;Haiyang Yu\",\"doi\":\"10.11607/jomi.9898\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>To verify a novel method that improves the accuracy of static computer-aided implant surgery (sCAIS) through intraoperative measurement.</p><p><strong>Materials and methods: </strong>Forty-seven patients were selected for this study, each with a missing tooth or a tooth that required extraction from the anterior area. The patients were divided into the intraoperative measuring guide (MG) and conventional guide (CG) groups. Following the preoperative implant planning, the surgical guides were designed and fabricated. In the MG group, the drill was guided by double-armed zirconia sleeves, and the axial direction of the drill was assessed using the indicator components. The implant was guided using a resin guide tube. In the CG group, the drills were guided using a metal sleeve and handles, and the implants were placed with the guidance of the metal sleeve only. The angular and linear deviations at the entry and apex between the planned and actual implant positions were measured after matching the preoperative and postoperative CBCT data. The independent-samples t test was used to compare the deviation between the MG and CG groups.</p><p><strong>Results: </strong>The 3D deviations for the MG group at the entry and apex were 0.67 ± 0.44 mm and 0.93 ± 0.40 mm, respectively. The angular deviation was 2.27 ± 0.96 degrees. Statistical differences were found in the 3D deviation at the entry point and apical position between the MG and CG groups, yielding relatively smaller deviations in the MG group.</p><p><strong>Conclusion: </strong>The use of an intraoperative measuring guide could improve the accuracy of implant placement in sCAIS.</p>\",\"PeriodicalId\":50298,\"journal\":{\"name\":\"International Journal of Oral & Maxillofacial Implants\",\"volume\":\"38 2\",\"pages\":\"277-286\"},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2023-03-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International Journal of Oral & Maxillofacial Implants\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.11607/jomi.9898\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"DENTISTRY, ORAL SURGERY & MEDICINE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Oral & Maxillofacial Implants","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.11607/jomi.9898","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

摘要

目的:验证一种通过术中测量提高静态计算机辅助植入手术(sCAIS)准确性的新方法。材料和方法:本研究选取了47例缺牙或需要从前牙区拔牙的患者。将患者分为术中测量引导(MG)组和常规引导(CG)组。根据术前种植计划,设计并制作手术指南。MG组采用双臂氧化锆滑套进行钻头导向,并使用指示组件评估钻头的轴向。种植体采用树脂引导管引导。在CG组中,使用金属套和手柄引导钻头,种植体仅在金属套的引导下放置。在匹配术前和术后CBCT数据后,测量计划种植体位置与实际种植体位置在入口和尖端的角度和线性偏差。采用独立样本t检验比较MG组和CG组之间的偏差。结果:MG组入、尖部三维偏差分别为0.67±0.44 mm和0.93±0.40 mm。角度偏差为2.27±0.96度。MG组和CG组在进入点和根尖位置的三维偏差有统计学差异,MG组的偏差相对较小。结论:术中使用测量指南可提高sCAIS植入的准确性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Accuracy of Intraoperative Measuring Guide and Conventional Surgical Guide in Anterior Implant Surgery: A Retrospective Case-Control Study

Purpose: To verify a novel method that improves the accuracy of static computer-aided implant surgery (sCAIS) through intraoperative measurement.

Materials and methods: Forty-seven patients were selected for this study, each with a missing tooth or a tooth that required extraction from the anterior area. The patients were divided into the intraoperative measuring guide (MG) and conventional guide (CG) groups. Following the preoperative implant planning, the surgical guides were designed and fabricated. In the MG group, the drill was guided by double-armed zirconia sleeves, and the axial direction of the drill was assessed using the indicator components. The implant was guided using a resin guide tube. In the CG group, the drills were guided using a metal sleeve and handles, and the implants were placed with the guidance of the metal sleeve only. The angular and linear deviations at the entry and apex between the planned and actual implant positions were measured after matching the preoperative and postoperative CBCT data. The independent-samples t test was used to compare the deviation between the MG and CG groups.

Results: The 3D deviations for the MG group at the entry and apex were 0.67 ± 0.44 mm and 0.93 ± 0.40 mm, respectively. The angular deviation was 2.27 ± 0.96 degrees. Statistical differences were found in the 3D deviation at the entry point and apical position between the MG and CG groups, yielding relatively smaller deviations in the MG group.

Conclusion: The use of an intraoperative measuring guide could improve the accuracy of implant placement in sCAIS.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
3.30
自引率
5.00%
发文量
115
审稿时长
6 months
期刊介绍: Edited by Steven E. Eckert, DDS, MS ISSN (Print): 0882-2786 ISSN (Online): 1942-4434 This highly regarded, often-cited journal integrates clinical and scientific data to improve methods and results of oral and maxillofacial implant therapy. It presents pioneering research, technology, clinical applications, reviews of the literature, seminal studies, emerging technology, position papers, and consensus studies, as well as the many clinical and therapeutic innovations that ensue as a result of these efforts. The editorial board is composed of recognized opinion leaders in their respective areas of expertise and reflects the international reach of the journal. Under their leadership, JOMI maintains its strong scientific integrity while expanding its influence within the field of implant dentistry. JOMI’s popular regular feature "Thematic Abstract Review" presents a review of abstracts of recently published articles on a specific topical area of interest each issue.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信