{"title":"含EDTA的未灌注血管:是否不适合用于糖化血红蛋白检测?","authors":"Berrak Guven, Ismail Benice, Murat Can","doi":"10.11613/BM.2023.010901","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Blood samples having inappropriate volume are a substantial part of preanalytical errors. Inadequate sample volume for glycated haemoglobin (HbA1c) test may be a common problem of patients with diabetes mellitus having vascular changes. In this study, we compared HbA1c concentrations of underfilled and appropriately filled blood collection tubes.</p><p><strong>Materials and methods: </strong>To compare HbA1c concentrations, blood samples were collected into 2 mL tubes containing K3-EDTA from 109 subjects. Two blood samples (underfilled and appropriately filled) were drawn from a patient by the same personnel and materials. HbA1c measurements were assayed on a Cobas 6000 analyser module c 501 (Roche Diagnostics, Mannheim, Germany). The HbA1c% results were compared by t-test and Wilcoxon's signed-rank statistical methods (SPSS Inc., Chicago, USA). Bias analysis was performed using Microsoft Excel 4.0.</p><p><strong>Results: </strong>Underfilled samples were classified three groups (group 1, N = 44; group 2, N = 36; and group 3, N = 29) according to the filling ratio of the samples; 0.5 mL and below (< 25%), 0.5-1.0 mL (25-50%), and 1.0-2.0 mL (> 50%), respectively. When we compared underfilled tubes with pairing filled tubes, there was a statistically significant difference only with tubes filled less than 25% (P = 0.030). Furthermore, we have done bias analysis between paired tubes according to the diagnostic cut-off value of 6.5%. The bias was more prominent in up to 50% underfilled blood tubes (1.1%), when HbA1c concentrations were below the diagnostic cut-off of 6.5%.</p><p><strong>Conclusions: </strong>We suggest that the blood tubes with EDTA for HbA1c measurement should be filled with at least 50% to avoid clinical variations.</p>","PeriodicalId":9021,"journal":{"name":"Biochemia Medica","volume":null,"pages":null},"PeriodicalIF":3.8000,"publicationDate":"2023-02-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9927728/pdf/","citationCount":"0","resultStr":"{\"title\":\"Undefilled blood tube containing EDTA: Is it an inappropriate sample for HbA1c assay?\",\"authors\":\"Berrak Guven, Ismail Benice, Murat Can\",\"doi\":\"10.11613/BM.2023.010901\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>Blood samples having inappropriate volume are a substantial part of preanalytical errors. Inadequate sample volume for glycated haemoglobin (HbA1c) test may be a common problem of patients with diabetes mellitus having vascular changes. In this study, we compared HbA1c concentrations of underfilled and appropriately filled blood collection tubes.</p><p><strong>Materials and methods: </strong>To compare HbA1c concentrations, blood samples were collected into 2 mL tubes containing K3-EDTA from 109 subjects. Two blood samples (underfilled and appropriately filled) were drawn from a patient by the same personnel and materials. HbA1c measurements were assayed on a Cobas 6000 analyser module c 501 (Roche Diagnostics, Mannheim, Germany). The HbA1c% results were compared by t-test and Wilcoxon's signed-rank statistical methods (SPSS Inc., Chicago, USA). Bias analysis was performed using Microsoft Excel 4.0.</p><p><strong>Results: </strong>Underfilled samples were classified three groups (group 1, N = 44; group 2, N = 36; and group 3, N = 29) according to the filling ratio of the samples; 0.5 mL and below (< 25%), 0.5-1.0 mL (25-50%), and 1.0-2.0 mL (> 50%), respectively. When we compared underfilled tubes with pairing filled tubes, there was a statistically significant difference only with tubes filled less than 25% (P = 0.030). Furthermore, we have done bias analysis between paired tubes according to the diagnostic cut-off value of 6.5%. The bias was more prominent in up to 50% underfilled blood tubes (1.1%), when HbA1c concentrations were below the diagnostic cut-off of 6.5%.</p><p><strong>Conclusions: </strong>We suggest that the blood tubes with EDTA for HbA1c measurement should be filled with at least 50% to avoid clinical variations.</p>\",\"PeriodicalId\":9021,\"journal\":{\"name\":\"Biochemia Medica\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":3.8000,\"publicationDate\":\"2023-02-15\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9927728/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Biochemia Medica\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.11613/BM.2023.010901\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biochemia Medica","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.11613/BM.2023.010901","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
Undefilled blood tube containing EDTA: Is it an inappropriate sample for HbA1c assay?
Introduction: Blood samples having inappropriate volume are a substantial part of preanalytical errors. Inadequate sample volume for glycated haemoglobin (HbA1c) test may be a common problem of patients with diabetes mellitus having vascular changes. In this study, we compared HbA1c concentrations of underfilled and appropriately filled blood collection tubes.
Materials and methods: To compare HbA1c concentrations, blood samples were collected into 2 mL tubes containing K3-EDTA from 109 subjects. Two blood samples (underfilled and appropriately filled) were drawn from a patient by the same personnel and materials. HbA1c measurements were assayed on a Cobas 6000 analyser module c 501 (Roche Diagnostics, Mannheim, Germany). The HbA1c% results were compared by t-test and Wilcoxon's signed-rank statistical methods (SPSS Inc., Chicago, USA). Bias analysis was performed using Microsoft Excel 4.0.
Results: Underfilled samples were classified three groups (group 1, N = 44; group 2, N = 36; and group 3, N = 29) according to the filling ratio of the samples; 0.5 mL and below (< 25%), 0.5-1.0 mL (25-50%), and 1.0-2.0 mL (> 50%), respectively. When we compared underfilled tubes with pairing filled tubes, there was a statistically significant difference only with tubes filled less than 25% (P = 0.030). Furthermore, we have done bias analysis between paired tubes according to the diagnostic cut-off value of 6.5%. The bias was more prominent in up to 50% underfilled blood tubes (1.1%), when HbA1c concentrations were below the diagnostic cut-off of 6.5%.
Conclusions: We suggest that the blood tubes with EDTA for HbA1c measurement should be filled with at least 50% to avoid clinical variations.
期刊介绍:
Biochemia Medica is the official peer-reviewed journal of the Croatian Society of Medical Biochemistry and Laboratory Medicine. Journal provides a wide coverage of research in all aspects of clinical chemistry and laboratory medicine. Following categories fit into the scope of the Journal: general clinical chemistry, haematology and haemostasis, molecular diagnostics and endocrinology. Development, validation and verification of analytical techniques and methods applicable to clinical chemistry and laboratory medicine are welcome as well as studies dealing with laboratory organization, automation and quality control. Journal publishes on a regular basis educative preanalytical case reports (Preanalytical mysteries), articles dealing with applied biostatistics (Lessons in biostatistics) and research integrity (Research integrity corner).