{"title":"一些常规x射线照相设备预测与测量x射线输出的比较","authors":"Calvin Didier Njiki, Yolande Huguette Ebele Yigbedeck, Joseph Eric Martial Ndjaka Manyol, Thierry Ndzana Ndah","doi":"10.4236/IJMPCERO.2019.84018","DOIUrl":null,"url":null,"abstract":"This study aims to investigate how accurate are TASMICS and TASMIP \nmodels in predicting the X-ray output of some Conventional Radiology X-ray \nunits with high frequency generators. The X-ray output in microgray per milliampere \nseconds (μGy/mAs) at 100 cm from the X-ray tube was determined \nfor selected high voltages and taking into account the total filtration. The \nX-ray output was then measured directly with the multi-purpose detectors \n(MPD), Raysafe X2. The maximum relative error between measured and predicted \nvalues was found to be equal to 20%. The maximum relative error between \nmeasured and predicted values obtained demonstrates the difficulty of \naccurately predicting the X-ray tube output using TASMICS and TASMIP \nmodels since they are based on fixed anode angles and different composition \nof the tungsten anode.","PeriodicalId":14028,"journal":{"name":"International Journal of Medical Physics, Clinical Engineering and Radiation Oncology","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2019-10-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"Comparison between Predicted and Measured X-Ray Output in Some Conventional Radiography Units\",\"authors\":\"Calvin Didier Njiki, Yolande Huguette Ebele Yigbedeck, Joseph Eric Martial Ndjaka Manyol, Thierry Ndzana Ndah\",\"doi\":\"10.4236/IJMPCERO.2019.84018\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"This study aims to investigate how accurate are TASMICS and TASMIP \\nmodels in predicting the X-ray output of some Conventional Radiology X-ray \\nunits with high frequency generators. The X-ray output in microgray per milliampere \\nseconds (μGy/mAs) at 100 cm from the X-ray tube was determined \\nfor selected high voltages and taking into account the total filtration. The \\nX-ray output was then measured directly with the multi-purpose detectors \\n(MPD), Raysafe X2. The maximum relative error between measured and predicted \\nvalues was found to be equal to 20%. The maximum relative error between \\nmeasured and predicted values obtained demonstrates the difficulty of \\naccurately predicting the X-ray tube output using TASMICS and TASMIP \\nmodels since they are based on fixed anode angles and different composition \\nof the tungsten anode.\",\"PeriodicalId\":14028,\"journal\":{\"name\":\"International Journal of Medical Physics, Clinical Engineering and Radiation Oncology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2019-10-09\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International Journal of Medical Physics, Clinical Engineering and Radiation Oncology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4236/IJMPCERO.2019.84018\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Medical Physics, Clinical Engineering and Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4236/IJMPCERO.2019.84018","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Comparison between Predicted and Measured X-Ray Output in Some Conventional Radiography Units
This study aims to investigate how accurate are TASMICS and TASMIP
models in predicting the X-ray output of some Conventional Radiology X-ray
units with high frequency generators. The X-ray output in microgray per milliampere
seconds (μGy/mAs) at 100 cm from the X-ray tube was determined
for selected high voltages and taking into account the total filtration. The
X-ray output was then measured directly with the multi-purpose detectors
(MPD), Raysafe X2. The maximum relative error between measured and predicted
values was found to be equal to 20%. The maximum relative error between
measured and predicted values obtained demonstrates the difficulty of
accurately predicting the X-ray tube output using TASMICS and TASMIP
models since they are based on fixed anode angles and different composition
of the tungsten anode.