{"title":"基于LOCI™和eclia技术评估肿瘤标志物的方法比较","authors":"R. Dolscheid-Pommerich, Sarah Dolscheid, L. Eichhorn, B. Zur, S. Holdenrieder, B. Stoffel‐Wagner","doi":"10.1515/labmed-2016-0074","DOIUrl":null,"url":null,"abstract":"Abstract Background: Since the introduction of luminescent oxygen channeling immunoassays (LOCI™)-based assays in the daily laboratory routine of tumor marker measurements, only a small number of method comparisons with established immunoassays have been published. We performed a method comparison between LOCI™-based tumor marker assays for Dimension™ VISTA and electrochemiluminiscent immunoassays (ECLIA) for Cobas™ e411, for α-fetoprotein (AFP), carcinoembryonic antigen (CEA), CA 125, CA 15-3, CA 19-9, prostate-specific antigen (PSA) and free PSA (fPSA). Methods: Tumor markers were assessed in 1088 sera from routine diagnostics on the Dimension™ VISTA 1500 and Cobas™ e411 analyzers. Results: Strong correlations were achieved for PSA (r=0.999), AFP (r=0.994) and CEA (r=0.993). Results were quite comparable as only minor slopes of 1.05 (PSA), 1.02 (AFP) and 0.94 (CEA), respectively, were found. However, correlations for CA 125 (r=0.976), CA 19-9 (r=0.960), fPSA (r=0.950) and CA 15-3 (r=0.940) were only moderate, and considerable slopes were observed for these markers with higher values for CA 19-9 (slope 1.50) and lower ones for CA 15-3 (0.76), fPSA (0.75) and CA 125 (0.64), for Dimension™ VISTA 1500. Conclusions: We found excellent correlations and comparable values for AFP, CEA and PSA, but only moderate correlations for fPSA, CA 125, CA 15-3 and CA 19-9. The slopes for CA 19-9, CA 15-3, fPSA and CA 125 have to be considered when analysis methods for tumor markers are changed.","PeriodicalId":49926,"journal":{"name":"Laboratoriumsmedizin-Journal of Laboratory Medicine","volume":"14 1","pages":"11 - 3"},"PeriodicalIF":0.1000,"publicationDate":"2017-01-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"Method comparison of tumor markers assessed by LOCI™- and ECLIA-based technologies\",\"authors\":\"R. Dolscheid-Pommerich, Sarah Dolscheid, L. Eichhorn, B. Zur, S. Holdenrieder, B. Stoffel‐Wagner\",\"doi\":\"10.1515/labmed-2016-0074\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Abstract Background: Since the introduction of luminescent oxygen channeling immunoassays (LOCI™)-based assays in the daily laboratory routine of tumor marker measurements, only a small number of method comparisons with established immunoassays have been published. We performed a method comparison between LOCI™-based tumor marker assays for Dimension™ VISTA and electrochemiluminiscent immunoassays (ECLIA) for Cobas™ e411, for α-fetoprotein (AFP), carcinoembryonic antigen (CEA), CA 125, CA 15-3, CA 19-9, prostate-specific antigen (PSA) and free PSA (fPSA). Methods: Tumor markers were assessed in 1088 sera from routine diagnostics on the Dimension™ VISTA 1500 and Cobas™ e411 analyzers. Results: Strong correlations were achieved for PSA (r=0.999), AFP (r=0.994) and CEA (r=0.993). Results were quite comparable as only minor slopes of 1.05 (PSA), 1.02 (AFP) and 0.94 (CEA), respectively, were found. However, correlations for CA 125 (r=0.976), CA 19-9 (r=0.960), fPSA (r=0.950) and CA 15-3 (r=0.940) were only moderate, and considerable slopes were observed for these markers with higher values for CA 19-9 (slope 1.50) and lower ones for CA 15-3 (0.76), fPSA (0.75) and CA 125 (0.64), for Dimension™ VISTA 1500. Conclusions: We found excellent correlations and comparable values for AFP, CEA and PSA, but only moderate correlations for fPSA, CA 125, CA 15-3 and CA 19-9. The slopes for CA 19-9, CA 15-3, fPSA and CA 125 have to be considered when analysis methods for tumor markers are changed.\",\"PeriodicalId\":49926,\"journal\":{\"name\":\"Laboratoriumsmedizin-Journal of Laboratory Medicine\",\"volume\":\"14 1\",\"pages\":\"11 - 3\"},\"PeriodicalIF\":0.1000,\"publicationDate\":\"2017-01-31\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Laboratoriumsmedizin-Journal of Laboratory Medicine\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1515/labmed-2016-0074\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"OTORHINOLARYNGOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Laboratoriumsmedizin-Journal of Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1515/labmed-2016-0074","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"OTORHINOLARYNGOLOGY","Score":null,"Total":0}
引用次数: 1

摘要

背景:自从基于发光氧通道免疫测定法(LOCI™)的检测方法被引入肿瘤标志物的日常实验室常规测量中以来,只有少数方法与已建立的免疫测定法进行了比较。我们比较了基于LOCI™的肿瘤标志物分析法(Dimension™VISTA)和电化学发光免疫分析法(ECLIA) (Cobas™e411)、α-胎蛋白(AFP)、癌胚抗原(CEA)、ca125、ca15 -3、ca19 -9、前列腺特异性抗原(PSA)和游离PSA (fPSA))。方法:采用Dimension™VISTA 1500和Cobas™e411分析仪对1088例常规诊断血清中的肿瘤标志物进行评估。结果:PSA (r=0.999)、AFP (r=0.994)、CEA (r=0.993)具有较强的相关性。结果具有相当的可比性,只有较小的斜率分别为1.05 (PSA), 1.02 (AFP)和0.94 (CEA)。然而,CA 125 (r=0.976), CA 19-9 (r=0.960), fPSA (r=0.950)和CA 15-3 (r=0.940)的相关性仅为中等,并且这些标记的斜率相当大,对于Dimension™VISTA 1500而言,CA 19-9的值较高(斜率为1.50),CA 15-3 (0.76), fPSA(0.75)和CA 125(0.64)的值较低。结论:我们发现AFP、CEA和PSA具有良好的相关性和可比性,但fPSA、ca125、ca15 -3和ca19 -9只有中等相关性。当改变肿瘤标志物的分析方法时,必须考虑CA 19-9、CA 15-3、fPSA和CA 125的斜率。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Method comparison of tumor markers assessed by LOCI™- and ECLIA-based technologies
Abstract Background: Since the introduction of luminescent oxygen channeling immunoassays (LOCI™)-based assays in the daily laboratory routine of tumor marker measurements, only a small number of method comparisons with established immunoassays have been published. We performed a method comparison between LOCI™-based tumor marker assays for Dimension™ VISTA and electrochemiluminiscent immunoassays (ECLIA) for Cobas™ e411, for α-fetoprotein (AFP), carcinoembryonic antigen (CEA), CA 125, CA 15-3, CA 19-9, prostate-specific antigen (PSA) and free PSA (fPSA). Methods: Tumor markers were assessed in 1088 sera from routine diagnostics on the Dimension™ VISTA 1500 and Cobas™ e411 analyzers. Results: Strong correlations were achieved for PSA (r=0.999), AFP (r=0.994) and CEA (r=0.993). Results were quite comparable as only minor slopes of 1.05 (PSA), 1.02 (AFP) and 0.94 (CEA), respectively, were found. However, correlations for CA 125 (r=0.976), CA 19-9 (r=0.960), fPSA (r=0.950) and CA 15-3 (r=0.940) were only moderate, and considerable slopes were observed for these markers with higher values for CA 19-9 (slope 1.50) and lower ones for CA 15-3 (0.76), fPSA (0.75) and CA 125 (0.64), for Dimension™ VISTA 1500. Conclusions: We found excellent correlations and comparable values for AFP, CEA and PSA, but only moderate correlations for fPSA, CA 125, CA 15-3 and CA 19-9. The slopes for CA 19-9, CA 15-3, fPSA and CA 125 have to be considered when analysis methods for tumor markers are changed.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
0.80
自引率
0.00%
发文量
1
审稿时长
>12 weeks
期刊介绍: Information not localized
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信