四极电感耦合等离子体质谱法精确测量铅同位素的噪声识别和采样频率测定

Analusis Pub Date : 2000-10-01 DOI:10.1051/ANALUSIS:2000146
F. Monna, J. Loizeau, B. Thomas, C. Guéguen, P. Favarger, R. Losno, J. Dominik
{"title":"四极电感耦合等离子体质谱法精确测量铅同位素的噪声识别和采样频率测定","authors":"F. Monna, J. Loizeau, B. Thomas, C. Guéguen, P. Favarger, R. Losno, J. Dominik","doi":"10.1051/ANALUSIS:2000146","DOIUrl":null,"url":null,"abstract":"Analytical precision of the isotope ratios measured by quadrupole-based ICP-MS is drastically controlled by the low- frequency noises which originate from nebulisation and vaporisation processes, and from sample introduction systems. The unde- sirable influence of these latter can be however reduced by choosing efficiently the operating parameters. In the present study , the choice of the stabilisation time necessary in peak jump mode, and the one of the number of sweeps are discussed in the light of noise power spectra obtained with various speeds of the peristaltic pump used as sample introduction system. Obviously, the settings proposed are probably efficient only on our own ICP-MS, but they can be determined without any difficulty on any other equipment, only by following the methodology detailed here. With such guidelines, experimental within-run RSD % were observed following closely those predicted by the counting statistics ( ~ 110 %). The isotopic measurements of natural samples (overbank sediments, airborne particulate matter, lichens and rainwater) exhibited fair accuracy and good reproducibility, making highly con - venient the use of the ICP-MS, at least in an environmental purpose.","PeriodicalId":8221,"journal":{"name":"Analusis","volume":"186 1","pages":"750-757"},"PeriodicalIF":0.0000,"publicationDate":"2000-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"24","resultStr":"{\"title\":\"Noise identification and sampling frequency determination for precise Pb isotopic measurements by quadrupole-based Inductively Coupled Plasma Mass Spectrometry\",\"authors\":\"F. Monna, J. Loizeau, B. Thomas, C. Guéguen, P. Favarger, R. Losno, J. Dominik\",\"doi\":\"10.1051/ANALUSIS:2000146\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Analytical precision of the isotope ratios measured by quadrupole-based ICP-MS is drastically controlled by the low- frequency noises which originate from nebulisation and vaporisation processes, and from sample introduction systems. The unde- sirable influence of these latter can be however reduced by choosing efficiently the operating parameters. In the present study , the choice of the stabilisation time necessary in peak jump mode, and the one of the number of sweeps are discussed in the light of noise power spectra obtained with various speeds of the peristaltic pump used as sample introduction system. Obviously, the settings proposed are probably efficient only on our own ICP-MS, but they can be determined without any difficulty on any other equipment, only by following the methodology detailed here. With such guidelines, experimental within-run RSD % were observed following closely those predicted by the counting statistics ( ~ 110 %). The isotopic measurements of natural samples (overbank sediments, airborne particulate matter, lichens and rainwater) exhibited fair accuracy and good reproducibility, making highly con - venient the use of the ICP-MS, at least in an environmental purpose.\",\"PeriodicalId\":8221,\"journal\":{\"name\":\"Analusis\",\"volume\":\"186 1\",\"pages\":\"750-757\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2000-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"24\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Analusis\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1051/ANALUSIS:2000146\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Analusis","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1051/ANALUSIS:2000146","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 24

摘要

四极ICP-MS测量的同位素比率的分析精度受到来自雾化和汽化过程以及样品导入系统的低频噪声的严重控制。但是,通过合理选择操作参数,可以减少这些因素的不良影响。本文根据作为进样系统的蠕动泵在不同转速下获得的噪声功率谱,讨论了跳峰模式所需稳定时间的选择和扫频次数的选择。显然,建议的设置可能只在我们自己的ICP-MS上有效,但它们可以在任何其他设备上毫无困难地确定,只需遵循这里详细介绍的方法。在此指导下,观察到的实验运行内RSD %与计数统计预测值(~ 110%)非常接近。自然样品(河岸沉积物、空气颗粒物、地衣和雨水)的同位素测量显示出相当的准确性和良好的再现性,使ICP-MS的使用非常方便,至少在环境目的方面。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Noise identification and sampling frequency determination for precise Pb isotopic measurements by quadrupole-based Inductively Coupled Plasma Mass Spectrometry
Analytical precision of the isotope ratios measured by quadrupole-based ICP-MS is drastically controlled by the low- frequency noises which originate from nebulisation and vaporisation processes, and from sample introduction systems. The unde- sirable influence of these latter can be however reduced by choosing efficiently the operating parameters. In the present study , the choice of the stabilisation time necessary in peak jump mode, and the one of the number of sweeps are discussed in the light of noise power spectra obtained with various speeds of the peristaltic pump used as sample introduction system. Obviously, the settings proposed are probably efficient only on our own ICP-MS, but they can be determined without any difficulty on any other equipment, only by following the methodology detailed here. With such guidelines, experimental within-run RSD % were observed following closely those predicted by the counting statistics ( ~ 110 %). The isotopic measurements of natural samples (overbank sediments, airborne particulate matter, lichens and rainwater) exhibited fair accuracy and good reproducibility, making highly con - venient the use of the ICP-MS, at least in an environmental purpose.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信