{"title":"透视时间作为心血管介入治疗中辐射剂量预警指标的可行性研究","authors":"Hai-Wen Ding, Yongchun Wang","doi":"10.3760/CMA.J.ISSN.0254-5098.2020.03.014","DOIUrl":null,"url":null,"abstract":"Objective \nTo collect the date of radiation dose in reference air kerma(AK) and dose-area product (DAP) values in order to evaluate the feasibility of fluoroscopy time as a monitoring and warning indicator of radiation exposure in cardiovascular interventions. \n \n \nMethods \nThe study conducted a retrospective analysis of 736 patients who underwent coronary angiography(CAG)and percutaneous coronary intervention(PCI)from November 2016 to January 2018 in Shanghai Changhai Hospital. Based on the imaging equipments(a Siemens Ceiling system and a Siemens Biplane system)and cardiovascular interventions(CAG and PCI), the fluoroscopy time, AK values and DAP values were collected. The correlation of the radiation dose and fluoroscopy time was analyzed using Spearman correlation statistics. \n \n \nResults \nThe mean values of fluoroscopy time, fluoroscopy AK, total AK, fluoroscopy DAP and total DAP were(8.9±7.8)min, (472±474), (703±595)mGy, (4 578±4 085)and(6 253±4 938)μGy·m2 for Ceiling system and(8.6±7.3)min, (510±509), (733±614)mGy, (4 255±3 781)and (5 681±4 432)μGy·m2 for Biplane system, respectively. The mean values of CAG and PCI fluoroscopy time were(2.4±0.9)and(15.7±4.9)min, respectively.The ratio of fluoroscopy radiation dose (AK and DAP) to total dose was 74% and 78% in PCI procedures. There was a strongly correlation between fluoroscopy time and total AK (r=0.822) or total DAP (r=0.844) in cardiovascular interventions (P<0.001). \n \n \nConclusions \nThe radiation dose of fluoroscopy acquisition is the main source of overall radiation dose in cardiovascular interventions. Radiation dose is expected to increase as fluoroscopy time increases.The fluoroscopy timer as a protective tool of radiation exposure has a good reference and warning value in the clinical application of cardiovascular interventions. \n \n \nKey words: \nCardiovascular intervention; Radiation dose; Fluoroscopy time","PeriodicalId":36403,"journal":{"name":"中华放射医学与防护杂志","volume":"2009 1","pages":"237-240"},"PeriodicalIF":0.0000,"publicationDate":"2020-03-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Feasibility study of using fluoroscopy time as warning indicator for radiation dose in cardiovascular interventions\",\"authors\":\"Hai-Wen Ding, Yongchun Wang\",\"doi\":\"10.3760/CMA.J.ISSN.0254-5098.2020.03.014\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective \\nTo collect the date of radiation dose in reference air kerma(AK) and dose-area product (DAP) values in order to evaluate the feasibility of fluoroscopy time as a monitoring and warning indicator of radiation exposure in cardiovascular interventions. \\n \\n \\nMethods \\nThe study conducted a retrospective analysis of 736 patients who underwent coronary angiography(CAG)and percutaneous coronary intervention(PCI)from November 2016 to January 2018 in Shanghai Changhai Hospital. Based on the imaging equipments(a Siemens Ceiling system and a Siemens Biplane system)and cardiovascular interventions(CAG and PCI), the fluoroscopy time, AK values and DAP values were collected. The correlation of the radiation dose and fluoroscopy time was analyzed using Spearman correlation statistics. \\n \\n \\nResults \\nThe mean values of fluoroscopy time, fluoroscopy AK, total AK, fluoroscopy DAP and total DAP were(8.9±7.8)min, (472±474), (703±595)mGy, (4 578±4 085)and(6 253±4 938)μGy·m2 for Ceiling system and(8.6±7.3)min, (510±509), (733±614)mGy, (4 255±3 781)and (5 681±4 432)μGy·m2 for Biplane system, respectively. The mean values of CAG and PCI fluoroscopy time were(2.4±0.9)and(15.7±4.9)min, respectively.The ratio of fluoroscopy radiation dose (AK and DAP) to total dose was 74% and 78% in PCI procedures. There was a strongly correlation between fluoroscopy time and total AK (r=0.822) or total DAP (r=0.844) in cardiovascular interventions (P<0.001). \\n \\n \\nConclusions \\nThe radiation dose of fluoroscopy acquisition is the main source of overall radiation dose in cardiovascular interventions. Radiation dose is expected to increase as fluoroscopy time increases.The fluoroscopy timer as a protective tool of radiation exposure has a good reference and warning value in the clinical application of cardiovascular interventions. \\n \\n \\nKey words: \\nCardiovascular intervention; Radiation dose; Fluoroscopy time\",\"PeriodicalId\":36403,\"journal\":{\"name\":\"中华放射医学与防护杂志\",\"volume\":\"2009 1\",\"pages\":\"237-240\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2020-03-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"中华放射医学与防护杂志\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.2020.03.014\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"中华放射医学与防护杂志","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.2020.03.014","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

目的收集参考空气克尔玛(AK)和剂量面积积(DAP)的辐射剂量数据,评价透视时间作为心血管干预中辐射暴露监测预警指标的可行性。方法回顾性分析2016年11月至2018年1月在上海长海医院行冠状动脉造影(CAG)和经皮冠状动脉介入治疗(PCI)的736例患者。基于成像设备(西门子Ceiling系统和西门子Biplane系统)和心血管干预(CAG和PCI),收集透视时间、AK值和DAP值。采用Spearman相关统计分析放射剂量与透视时间的相关性。结果Ceiling系统的透视时间、透视AK、总AK、透视DAP和总DAP的平均值分别为(8.9±7.8)min、(472±474)、(703±595)mGy、(4 578±4 085)和(6 253±4 938)μGy·m2, Biplane系统的平均值分别为(8.6±7.3)min、(510±509)、(733±614)mGy、(4 255±3 781)和(5 681±4 432)μGy·m2。CAG和PCI透视时间平均值分别为(2.4±0.9)min和(15.7±4.9)min。PCI手术中透视辐射剂量(AK和DAP)占总剂量的比例分别为74%和78%。在心血管干预中,透视时间与总AK (r=0.822)或总DAP (r=0.844)有很强的相关性(P<0.001)。结论透视获取的辐射剂量是心血管干预中总辐射剂量的主要来源。放射剂量预计随着透视时间的增加而增加。透视计时器作为辐射暴露的防护工具,在心血管干预的临床应用中具有很好的参考和警示价值。关键词:心血管干预;辐射剂量;透视时间
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Feasibility study of using fluoroscopy time as warning indicator for radiation dose in cardiovascular interventions
Objective To collect the date of radiation dose in reference air kerma(AK) and dose-area product (DAP) values in order to evaluate the feasibility of fluoroscopy time as a monitoring and warning indicator of radiation exposure in cardiovascular interventions. Methods The study conducted a retrospective analysis of 736 patients who underwent coronary angiography(CAG)and percutaneous coronary intervention(PCI)from November 2016 to January 2018 in Shanghai Changhai Hospital. Based on the imaging equipments(a Siemens Ceiling system and a Siemens Biplane system)and cardiovascular interventions(CAG and PCI), the fluoroscopy time, AK values and DAP values were collected. The correlation of the radiation dose and fluoroscopy time was analyzed using Spearman correlation statistics. Results The mean values of fluoroscopy time, fluoroscopy AK, total AK, fluoroscopy DAP and total DAP were(8.9±7.8)min, (472±474), (703±595)mGy, (4 578±4 085)and(6 253±4 938)μGy·m2 for Ceiling system and(8.6±7.3)min, (510±509), (733±614)mGy, (4 255±3 781)and (5 681±4 432)μGy·m2 for Biplane system, respectively. The mean values of CAG and PCI fluoroscopy time were(2.4±0.9)and(15.7±4.9)min, respectively.The ratio of fluoroscopy radiation dose (AK and DAP) to total dose was 74% and 78% in PCI procedures. There was a strongly correlation between fluoroscopy time and total AK (r=0.822) or total DAP (r=0.844) in cardiovascular interventions (P<0.001). Conclusions The radiation dose of fluoroscopy acquisition is the main source of overall radiation dose in cardiovascular interventions. Radiation dose is expected to increase as fluoroscopy time increases.The fluoroscopy timer as a protective tool of radiation exposure has a good reference and warning value in the clinical application of cardiovascular interventions. Key words: Cardiovascular intervention; Radiation dose; Fluoroscopy time
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
中华放射医学与防护杂志
中华放射医学与防护杂志 Medicine-Radiology, Nuclear Medicine and Imaging
CiteScore
0.60
自引率
0.00%
发文量
6377
期刊介绍:
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信