结合计算机心电图和羊水指数预测新生儿出生时酸血症:一种改良的生物物理剖面。

J. Piazze, M. Anceschi, A. Berretta, S. Vitali, L. Maranghi, F. Amici, E. Cosmi
{"title":"结合计算机心电图和羊水指数预测新生儿出生时酸血症:一种改良的生物物理剖面。","authors":"J. Piazze, M. Anceschi, A. Berretta, S. Vitali, L. Maranghi, F. Amici, E. Cosmi","doi":"10.1080/714052769","DOIUrl":null,"url":null,"abstract":"Objective: To study the combination of computerized cardiotocography (cCTG) and the amniotic fluid index (AFI) in the prediction of neonatal acidemia at birth. Methods: A total of 89 singleton third-trimester high-risk pregnancies delivered by Cesarean section, with an AFI evaluated within 24 h from birth, and an antepartum cCTG performed within 6 h from delivery, were studied. The score was the sum of values for AFI (oligo/anhydramnios = 1, normal = 0) and cCTG (Dawes-Redman criteria, not met = 1, met = 0). The endpoint was to predict an abnormal neonatal outcome as defined by an umbilical artery pH of h 7.2. Results: Fifteen neonates had an umbilical artery pH of < 7.2. The combination of cCTG + AFI score was able to predict pH values ( h 7.20) with an OR = 2.83 ( p < 0.02). The diagnostic accuracy of the combination of cCTG + AFI was as follows: sensitivity 80%, specificity 58%, positive predictive value 28%, negative predictive value 83%. Comment: We suggest that the cCTG + AFI score may be of...","PeriodicalId":79464,"journal":{"name":"The Journal of maternal-fetal medicine","volume":"1 1","pages":"323-327"},"PeriodicalIF":0.0000,"publicationDate":"2001-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":"{\"title\":\"The combination of computerized cardiotocography and amniotic fluid index for the prediction of neonatal acidemia at birth: a modified biophysical profile.\",\"authors\":\"J. Piazze, M. Anceschi, A. Berretta, S. Vitali, L. Maranghi, F. Amici, E. Cosmi\",\"doi\":\"10.1080/714052769\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective: To study the combination of computerized cardiotocography (cCTG) and the amniotic fluid index (AFI) in the prediction of neonatal acidemia at birth. Methods: A total of 89 singleton third-trimester high-risk pregnancies delivered by Cesarean section, with an AFI evaluated within 24 h from birth, and an antepartum cCTG performed within 6 h from delivery, were studied. The score was the sum of values for AFI (oligo/anhydramnios = 1, normal = 0) and cCTG (Dawes-Redman criteria, not met = 1, met = 0). The endpoint was to predict an abnormal neonatal outcome as defined by an umbilical artery pH of h 7.2. Results: Fifteen neonates had an umbilical artery pH of < 7.2. The combination of cCTG + AFI score was able to predict pH values ( h 7.20) with an OR = 2.83 ( p < 0.02). The diagnostic accuracy of the combination of cCTG + AFI was as follows: sensitivity 80%, specificity 58%, positive predictive value 28%, negative predictive value 83%. Comment: We suggest that the cCTG + AFI score may be of...\",\"PeriodicalId\":79464,\"journal\":{\"name\":\"The Journal of maternal-fetal medicine\",\"volume\":\"1 1\",\"pages\":\"323-327\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2001-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"2\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"The Journal of maternal-fetal medicine\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1080/714052769\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"The Journal of maternal-fetal medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1080/714052769","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 2

摘要

目的:探讨计算机心脏造影(cCTG)与羊水指数(AFI)联合预测新生儿出生时酸血症的价值。方法:对89例剖宫产妊娠晚期高危单胎患者进行研究,这些患者在出生后24小时内进行AFI评估,并在分娩后6小时内进行产前cCTG检查。得分为AFI(羊水少/无= 1,正常= 0)和cCTG (Dawes-Redman标准,不符合= 1,符合= 0)的值之和。终点是通过脐动脉pH值h7.2来预测新生儿的异常结局。结果:15例新生儿脐动脉pH < 7.2。cCTG + AFI评分联合预测pH值(h 7.20), OR = 2.83 (p < 0.02)。cCTG + AFI联合诊断准确率:敏感性80%,特异性58%,阳性预测值28%,阴性预测值83%。评论:我们认为cCTG + AFI分数可能是…
本文章由计算机程序翻译,如有差异,请以英文原文为准。
The combination of computerized cardiotocography and amniotic fluid index for the prediction of neonatal acidemia at birth: a modified biophysical profile.
Objective: To study the combination of computerized cardiotocography (cCTG) and the amniotic fluid index (AFI) in the prediction of neonatal acidemia at birth. Methods: A total of 89 singleton third-trimester high-risk pregnancies delivered by Cesarean section, with an AFI evaluated within 24 h from birth, and an antepartum cCTG performed within 6 h from delivery, were studied. The score was the sum of values for AFI (oligo/anhydramnios = 1, normal = 0) and cCTG (Dawes-Redman criteria, not met = 1, met = 0). The endpoint was to predict an abnormal neonatal outcome as defined by an umbilical artery pH of h 7.2. Results: Fifteen neonates had an umbilical artery pH of < 7.2. The combination of cCTG + AFI score was able to predict pH values ( h 7.20) with an OR = 2.83 ( p < 0.02). The diagnostic accuracy of the combination of cCTG + AFI was as follows: sensitivity 80%, specificity 58%, positive predictive value 28%, negative predictive value 83%. Comment: We suggest that the cCTG + AFI score may be of...
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信