CdTe和NaI (Tl)厚度作为小动物SPECT系统检测器的SIMIND Monte Carlo模拟研究

Q3 Health Professions
S. Abbaspour, B. Mahmoudian, S. Zakavi, Jalil Pirayesh Islamian
{"title":"CdTe和NaI (Tl)厚度作为小动物SPECT系统检测器的SIMIND Monte Carlo模拟研究","authors":"S. Abbaspour, B. Mahmoudian, S. Zakavi, Jalil Pirayesh Islamian","doi":"10.18502/fbt.v9i4.10424","DOIUrl":null,"url":null,"abstract":"Purpose: Micro Single Photon Emission Computed Tomography (Micro-SPECT) system has recently been introduced on nuclear medicine in the preclinical and research in which NaI (Tl) and Cadmium Telluride (CdTe) are used as the gamma-ray detectors with more generally use of NaI (Tl). The present study aimed to evaluate different thicknesses of the NaI (Tl) and (CdTe) detectors on functional parameters of a micro-SPECT system. \nMaterials and Methods: A Micro-SPECT system with CdTe semiconductor detector and a hexagonal parallel hole collimator with a hole diameter of 0.11 mm, high of 24.05 mm, and septal thickness of 0.016 mm was simulated by Simulating Medical Imaging Nuclear Detectors (SIMIND) Monte Carlo program. The system performance was assessed by comparing the functional parameters, including system efficiency, sensitivity, energy and spatial resolution with the NaI (Tl) detector. The simulated scans of a 99mTc point source, a digital micro-Jacszack phantom, and a voxelized MOBY mouse phantom with the system were prepared to evaluate image quality. \nResults: The functional parameters; sensitivity, efficiency, planar spatial resolution, and image contrast of CdTe detector were determined 1.4, 1.2, 1.7, and 1.8 times higher than those of NaI (Tl), respectively. Moreover, the calculated energy resolution of CdTe and NaI (Tl) detectors was obtained 6.2% and 10.2% at 141 KeV, respectively. In the filtered back projection (FBP) reconstructed images of the micro-Jacszack phantom, minimum detectable size of the cold rods with CdTe and NaI (Tl) detectors were obtained 0.79 mm and 0.95 mm, respectively. \nConclusion: The imaging system with a 5.5 mm thickness CdTe detector provided better image quality and showed considerable efficiency.","PeriodicalId":34203,"journal":{"name":"Frontiers in Biomedical Technologies","volume":"1 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2022-08-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"A SIMIND Monte Carlo Simulation Study on CdTe and NaI (Tl) Thickness as Detectors of a Small Animal SPECT System\",\"authors\":\"S. Abbaspour, B. Mahmoudian, S. Zakavi, Jalil Pirayesh Islamian\",\"doi\":\"10.18502/fbt.v9i4.10424\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Purpose: Micro Single Photon Emission Computed Tomography (Micro-SPECT) system has recently been introduced on nuclear medicine in the preclinical and research in which NaI (Tl) and Cadmium Telluride (CdTe) are used as the gamma-ray detectors with more generally use of NaI (Tl). The present study aimed to evaluate different thicknesses of the NaI (Tl) and (CdTe) detectors on functional parameters of a micro-SPECT system. \\nMaterials and Methods: A Micro-SPECT system with CdTe semiconductor detector and a hexagonal parallel hole collimator with a hole diameter of 0.11 mm, high of 24.05 mm, and septal thickness of 0.016 mm was simulated by Simulating Medical Imaging Nuclear Detectors (SIMIND) Monte Carlo program. The system performance was assessed by comparing the functional parameters, including system efficiency, sensitivity, energy and spatial resolution with the NaI (Tl) detector. The simulated scans of a 99mTc point source, a digital micro-Jacszack phantom, and a voxelized MOBY mouse phantom with the system were prepared to evaluate image quality. \\nResults: The functional parameters; sensitivity, efficiency, planar spatial resolution, and image contrast of CdTe detector were determined 1.4, 1.2, 1.7, and 1.8 times higher than those of NaI (Tl), respectively. Moreover, the calculated energy resolution of CdTe and NaI (Tl) detectors was obtained 6.2% and 10.2% at 141 KeV, respectively. In the filtered back projection (FBP) reconstructed images of the micro-Jacszack phantom, minimum detectable size of the cold rods with CdTe and NaI (Tl) detectors were obtained 0.79 mm and 0.95 mm, respectively. \\nConclusion: The imaging system with a 5.5 mm thickness CdTe detector provided better image quality and showed considerable efficiency.\",\"PeriodicalId\":34203,\"journal\":{\"name\":\"Frontiers in Biomedical Technologies\",\"volume\":\"1 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-08-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Frontiers in Biomedical Technologies\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.18502/fbt.v9i4.10424\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Health Professions\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Frontiers in Biomedical Technologies","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.18502/fbt.v9i4.10424","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Health Professions","Score":null,"Total":0}
引用次数: 1

摘要

目的:微单光子发射计算机断层扫描(Micro- spect)系统最近被引入核医学的临床前和研究中,其中NaI (Tl)和碲化镉(CdTe)作为伽马射线探测器,更广泛地使用NaI (Tl)。本研究旨在评估不同厚度的NaI (Tl)和(CdTe)探测器对微spect系统功能参数的影响。材料与方法:采用模拟医学成像核探测器(simmind)蒙特卡罗程序模拟了CdTe半导体探测器和孔径为0.11 mm、高24.05 mm、间隔厚度0.016 mm的六角形平行孔准直器的微spect系统。通过与NaI (Tl)探测器比较系统效率、灵敏度、能量和空间分辨率等功能参数来评估系统性能。利用该系统对99mTc点光源、数字微型杰克扎克幻影和体素化MOBY小鼠幻影进行了模拟扫描,以评估图像质量。结果:功能参数;CdTe探测器的灵敏度、效率、平面空间分辨率和图像对比度分别是NaI (Tl)的1.4倍、1.2倍、1.7倍和1.8倍。此外,CdTe和NaI (Tl)探测器在141 KeV下的计算能量分辨率分别为6.2%和10.2%。在微jackszack幻影的滤波后投影(FBP)重建图像中,CdTe和NaI (Tl)探测器可检测到的冷棒最小尺寸分别为0.79 mm和0.95 mm。结论:采用厚度为5.5 mm的CdTe探测器的成像系统具有较好的成像质量和较高的成像效率。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
A SIMIND Monte Carlo Simulation Study on CdTe and NaI (Tl) Thickness as Detectors of a Small Animal SPECT System
Purpose: Micro Single Photon Emission Computed Tomography (Micro-SPECT) system has recently been introduced on nuclear medicine in the preclinical and research in which NaI (Tl) and Cadmium Telluride (CdTe) are used as the gamma-ray detectors with more generally use of NaI (Tl). The present study aimed to evaluate different thicknesses of the NaI (Tl) and (CdTe) detectors on functional parameters of a micro-SPECT system. Materials and Methods: A Micro-SPECT system with CdTe semiconductor detector and a hexagonal parallel hole collimator with a hole diameter of 0.11 mm, high of 24.05 mm, and septal thickness of 0.016 mm was simulated by Simulating Medical Imaging Nuclear Detectors (SIMIND) Monte Carlo program. The system performance was assessed by comparing the functional parameters, including system efficiency, sensitivity, energy and spatial resolution with the NaI (Tl) detector. The simulated scans of a 99mTc point source, a digital micro-Jacszack phantom, and a voxelized MOBY mouse phantom with the system were prepared to evaluate image quality. Results: The functional parameters; sensitivity, efficiency, planar spatial resolution, and image contrast of CdTe detector were determined 1.4, 1.2, 1.7, and 1.8 times higher than those of NaI (Tl), respectively. Moreover, the calculated energy resolution of CdTe and NaI (Tl) detectors was obtained 6.2% and 10.2% at 141 KeV, respectively. In the filtered back projection (FBP) reconstructed images of the micro-Jacszack phantom, minimum detectable size of the cold rods with CdTe and NaI (Tl) detectors were obtained 0.79 mm and 0.95 mm, respectively. Conclusion: The imaging system with a 5.5 mm thickness CdTe detector provided better image quality and showed considerable efficiency.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Frontiers in Biomedical Technologies
Frontiers in Biomedical Technologies Health Professions-Medical Laboratory Technology
CiteScore
0.80
自引率
0.00%
发文量
34
审稿时长
12 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信