{"title":"一种定位超声显像导管的四自由度机器人。","authors":"P. M. Loschak, Alperen Değirmenci, Yaroslav Tenzer, R. Howe","doi":"10.1115/DETC2015-47693","DOIUrl":null,"url":null,"abstract":"In this paper we present the design, fabrication, and testing of a robot for automatically positioning ultrasound imaging catheters. Our system will point ultrasound (US) catheters to provide real-time imaging of anatomical structures and working instruments during minimally invasive surgeries. Manually navigating US catheters is difficult and requires extensive training in order to aim the US imager at desired targets. Therefore, a four DOF robotic system was developed to automatically navigate US imaging catheters for enhanced imaging. A rotational transmission enables three DOF for pitch, yaw, and roll of the imager. This transmission is translated by the fourth DOF. An accuracy analysis was conducted to calculate the maximum allowable joint motion error. Rotational joints must be accurate to within 1.5° and the translational joint must be accurate within 1.4 mm. Motion tests were then conducted to validate the accuracy of the robot. The average resulting errors in positioning of the rotational joints were measured to be 0.28°-0.38° with average measured backlash error 0.44°. Average translational positioning and backlash errors were measured to be significantly lower than the reported accuracy of the position sensor. The resulting joint motion errors were well within the required specifications for accurate robot motion. Such effective navigation of US imaging catheters will enable better visualization in various procedures ranging from cardiac arrhythmia treatment to tumor removal in urological cases.","PeriodicalId":74514,"journal":{"name":"Proceedings of the ... ASME Design Engineering Technical Conferences. ASME Design Engineering Technical Conferences","volume":"27 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2015-08-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"11","resultStr":"{\"title\":\"A 4-DOF Robot for Positioning Ultrasound Imaging Catheters.\",\"authors\":\"P. M. Loschak, Alperen Değirmenci, Yaroslav Tenzer, R. Howe\",\"doi\":\"10.1115/DETC2015-47693\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"In this paper we present the design, fabrication, and testing of a robot for automatically positioning ultrasound imaging catheters. Our system will point ultrasound (US) catheters to provide real-time imaging of anatomical structures and working instruments during minimally invasive surgeries. Manually navigating US catheters is difficult and requires extensive training in order to aim the US imager at desired targets. Therefore, a four DOF robotic system was developed to automatically navigate US imaging catheters for enhanced imaging. A rotational transmission enables three DOF for pitch, yaw, and roll of the imager. This transmission is translated by the fourth DOF. An accuracy analysis was conducted to calculate the maximum allowable joint motion error. Rotational joints must be accurate to within 1.5° and the translational joint must be accurate within 1.4 mm. Motion tests were then conducted to validate the accuracy of the robot. The average resulting errors in positioning of the rotational joints were measured to be 0.28°-0.38° with average measured backlash error 0.44°. Average translational positioning and backlash errors were measured to be significantly lower than the reported accuracy of the position sensor. The resulting joint motion errors were well within the required specifications for accurate robot motion. Such effective navigation of US imaging catheters will enable better visualization in various procedures ranging from cardiac arrhythmia treatment to tumor removal in urological cases.\",\"PeriodicalId\":74514,\"journal\":{\"name\":\"Proceedings of the ... ASME Design Engineering Technical Conferences. ASME Design Engineering Technical Conferences\",\"volume\":\"27 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2015-08-02\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"11\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Proceedings of the ... ASME Design Engineering Technical Conferences. ASME Design Engineering Technical Conferences\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1115/DETC2015-47693\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Proceedings of the ... ASME Design Engineering Technical Conferences. ASME Design Engineering Technical Conferences","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1115/DETC2015-47693","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 11

摘要

在本文中,我们提出了一个自动定位超声成像导管的机器人的设计,制造和测试。我们的系统将指向超声(US)导管,在微创手术中提供解剖结构和工作器械的实时成像。手动导航美国导管是困难的,需要广泛的培训,以瞄准美国成像仪期望的目标。因此,开发了一种四自由度机器人系统来自动导航US成像导管以增强成像。旋转传输使三自由度俯仰,偏航,和滚动的成像仪。该传输由第四自由度转换。进行了精度分析,计算了最大允许关节运动误差。旋转接头必须精确到1.5°以内,平移接头必须精确到1.4 mm以内。然后进行运动测试以验证机器人的准确性。测量的旋转关节定位平均误差为0.28°-0.38°,测量的平均间隙误差为0.44°。平均平移定位和间隙误差测量显着低于报告的位置传感器的精度。由此产生的关节运动误差完全在精确机器人运动所需的规格范围内。这种有效的导航超声显像导管将使从心律失常治疗到泌尿外科肿瘤切除的各种手术更好地可视化。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
A 4-DOF Robot for Positioning Ultrasound Imaging Catheters.
In this paper we present the design, fabrication, and testing of a robot for automatically positioning ultrasound imaging catheters. Our system will point ultrasound (US) catheters to provide real-time imaging of anatomical structures and working instruments during minimally invasive surgeries. Manually navigating US catheters is difficult and requires extensive training in order to aim the US imager at desired targets. Therefore, a four DOF robotic system was developed to automatically navigate US imaging catheters for enhanced imaging. A rotational transmission enables three DOF for pitch, yaw, and roll of the imager. This transmission is translated by the fourth DOF. An accuracy analysis was conducted to calculate the maximum allowable joint motion error. Rotational joints must be accurate to within 1.5° and the translational joint must be accurate within 1.4 mm. Motion tests were then conducted to validate the accuracy of the robot. The average resulting errors in positioning of the rotational joints were measured to be 0.28°-0.38° with average measured backlash error 0.44°. Average translational positioning and backlash errors were measured to be significantly lower than the reported accuracy of the position sensor. The resulting joint motion errors were well within the required specifications for accurate robot motion. Such effective navigation of US imaging catheters will enable better visualization in various procedures ranging from cardiac arrhythmia treatment to tumor removal in urological cases.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信