用伏安法分析水介质中的钴(II)和镍(II)

Devi Aliefiyardi Aulia Widowati, Fanny Kurniawan, Shynta Ramadhan
{"title":"用伏安法分析水介质中的钴(II)和镍(II)","authors":"Devi Aliefiyardi Aulia Widowati, Fanny Kurniawan, Shynta Ramadhan","doi":"10.56425/cma.v2i2.52","DOIUrl":null,"url":null,"abstract":"Currently, the electrochemical method as one of the fast, accurate, and reliable techniques for detecting heavy metals such as Co and Ni has been widely developed. Cyclic voltammetry methodology was successfully employed for determination of trace cobalt (Co) and nickel (Ni). Co and Ni levels are studied because they can cause water pollution and can be toxic, and non-biodegradable which influences the health of living things. To obtain results with high accuracy, this study tested several electrochemical parameters. Based on the results processed, the CoNi level in the solution can be measured well in the voltage range, scan rate, and pH of the solution, respectively, at -1.00 V to 0.75 V, 250 mV/s, and pH 3 (Ni) and pH 7 (Co). This work have been validated and showed good results with correlation coefficients of Co and Ni. The detection limit and quantity limit for Co and Ni were 510.2 ppm and 723.5 ppm, respectively, with 274.99 ppm and 916.66 ppm. Co and Ni metal RSDs also showed good values (<5%). In this study, Co and Ni levels in lab waste obtained by the cyclic voltammetry method were 510.2 ppm and 1666.47 ppm, and the results obtained from atomic absorption spectroscopy (AAS) measurements of 510.7 ppm and 1676 ppm. The developed electrochemical systems based are giving new inputs to the existing devices or leading to the development of novel heavy metal detection tools with interest for applications in fields such as environmental, safety, security controls or other industries.","PeriodicalId":9724,"journal":{"name":"chemistry and materials research","volume":"46 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2023-06-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Analysis of Cobalt(II) and Nickel(II) in Water Medium using Voltammetry Techniques\",\"authors\":\"Devi Aliefiyardi Aulia Widowati, Fanny Kurniawan, Shynta Ramadhan\",\"doi\":\"10.56425/cma.v2i2.52\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Currently, the electrochemical method as one of the fast, accurate, and reliable techniques for detecting heavy metals such as Co and Ni has been widely developed. Cyclic voltammetry methodology was successfully employed for determination of trace cobalt (Co) and nickel (Ni). Co and Ni levels are studied because they can cause water pollution and can be toxic, and non-biodegradable which influences the health of living things. To obtain results with high accuracy, this study tested several electrochemical parameters. Based on the results processed, the CoNi level in the solution can be measured well in the voltage range, scan rate, and pH of the solution, respectively, at -1.00 V to 0.75 V, 250 mV/s, and pH 3 (Ni) and pH 7 (Co). This work have been validated and showed good results with correlation coefficients of Co and Ni. The detection limit and quantity limit for Co and Ni were 510.2 ppm and 723.5 ppm, respectively, with 274.99 ppm and 916.66 ppm. Co and Ni metal RSDs also showed good values (<5%). In this study, Co and Ni levels in lab waste obtained by the cyclic voltammetry method were 510.2 ppm and 1666.47 ppm, and the results obtained from atomic absorption spectroscopy (AAS) measurements of 510.7 ppm and 1676 ppm. The developed electrochemical systems based are giving new inputs to the existing devices or leading to the development of novel heavy metal detection tools with interest for applications in fields such as environmental, safety, security controls or other industries.\",\"PeriodicalId\":9724,\"journal\":{\"name\":\"chemistry and materials research\",\"volume\":\"46 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-06-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"chemistry and materials research\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.56425/cma.v2i2.52\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"chemistry and materials research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.56425/cma.v2i2.52","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目前,电化学方法作为一种快速、准确、可靠的检测Co、Ni等重金属的技术得到了广泛的发展。采用循环伏安法成功地测定了痕量钴(Co)和镍(Ni)。研究Co和Ni的水平,因为它们会造成水污染,有毒,不可生物降解,影响生物的健康。为了获得高精度的结果,本研究测试了几个电化学参数。结果表明,在-1.00 V ~ 0.75 V、250 mV/s、pH 3 (Ni)和pH 7 (Co)的电压范围、扫描速率和溶液pH下,可以很好地测量溶液中的CoNi水平。用Co和Ni的相关系数对这一工作进行了验证,得到了很好的结果。Co和Ni的检出限分别为510.2 ppm和723.5 ppm,检出限分别为274.99 ppm和916.66 ppm。Co和Ni金属的rsd值也很好(<5%)。本研究中,循环伏安法测定的实验室废弃物中Co和Ni含量分别为510.2 ppm和1666.47 ppm,原子吸收光谱法测定的结果分别为510.7 ppm和1676 ppm。开发的电化学系统为现有设备提供了新的输入,或导致新型重金属检测工具的开发,这些工具对环境,安全,安全控制或其他行业等领域的应用感兴趣。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Analysis of Cobalt(II) and Nickel(II) in Water Medium using Voltammetry Techniques
Currently, the electrochemical method as one of the fast, accurate, and reliable techniques for detecting heavy metals such as Co and Ni has been widely developed. Cyclic voltammetry methodology was successfully employed for determination of trace cobalt (Co) and nickel (Ni). Co and Ni levels are studied because they can cause water pollution and can be toxic, and non-biodegradable which influences the health of living things. To obtain results with high accuracy, this study tested several electrochemical parameters. Based on the results processed, the CoNi level in the solution can be measured well in the voltage range, scan rate, and pH of the solution, respectively, at -1.00 V to 0.75 V, 250 mV/s, and pH 3 (Ni) and pH 7 (Co). This work have been validated and showed good results with correlation coefficients of Co and Ni. The detection limit and quantity limit for Co and Ni were 510.2 ppm and 723.5 ppm, respectively, with 274.99 ppm and 916.66 ppm. Co and Ni metal RSDs also showed good values (<5%). In this study, Co and Ni levels in lab waste obtained by the cyclic voltammetry method were 510.2 ppm and 1666.47 ppm, and the results obtained from atomic absorption spectroscopy (AAS) measurements of 510.7 ppm and 1676 ppm. The developed electrochemical systems based are giving new inputs to the existing devices or leading to the development of novel heavy metal detection tools with interest for applications in fields such as environmental, safety, security controls or other industries.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信