{"title":"图像引导对准系统对四触感疏水单焦环形人工晶状体的疗效","authors":"D. Shin, Jeong Min Lee, Tae-im Kim, K. Y. Seo, K. Koh","doi":"10.1097/ICL.0000000000000901","DOIUrl":null,"url":null,"abstract":"Objective: To compare the difference in performance between the image-guided alignment system and the manual-marking method in the four-haptic hydrophobic monofocal toric intraocular lens (IOL). Methods: Medical records of patients who underwent cataract surgery with a four-haptic hydrophobic monofocal toric IOL implantation between May 2020 and April 2021 and with 3-month visit data available were investigated. Toric IOL misalignment, residual astigmatism, and mean prediction errors between the two groups were compared. Results: This study included 49 eyes of 44 patients (women: 68%; mean age: 67.2±7.0 [range: 47–82] years). Twenty-nine eyes of 26 patients were treated with toric IOL implantation using the image-guided system and 20 eyes of 18 patients were treated using the manual-marking method. No statistical differences were observed regarding the baseline characteristics of the two groups. Three months after the surgery, the misalignment of the toric IOL was significantly lower in the image-guided group (2.18°±0.65°, range: 1.26°–3.95°) than in the manual; marking group (4.72°±0.74°, range: 3.44°–6.21°; P<0.001). Conclusion: In comparison to the manual-marking method, the image-guided system reduced the misalignment of a four-haptic hydrophobic monofocal toric IOL.","PeriodicalId":12216,"journal":{"name":"Eye & Contact Lens: Science & Clinical Practice","volume":"30 1","pages":"396 - 402"},"PeriodicalIF":0.0000,"publicationDate":"2022-05-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Efficacy of the Image-Guided Alignment System for a Four-Haptic Hydrophobic Monofocal Toric Intraocular Lens\",\"authors\":\"D. Shin, Jeong Min Lee, Tae-im Kim, K. Y. Seo, K. Koh\",\"doi\":\"10.1097/ICL.0000000000000901\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective: To compare the difference in performance between the image-guided alignment system and the manual-marking method in the four-haptic hydrophobic monofocal toric intraocular lens (IOL). Methods: Medical records of patients who underwent cataract surgery with a four-haptic hydrophobic monofocal toric IOL implantation between May 2020 and April 2021 and with 3-month visit data available were investigated. Toric IOL misalignment, residual astigmatism, and mean prediction errors between the two groups were compared. Results: This study included 49 eyes of 44 patients (women: 68%; mean age: 67.2±7.0 [range: 47–82] years). Twenty-nine eyes of 26 patients were treated with toric IOL implantation using the image-guided system and 20 eyes of 18 patients were treated using the manual-marking method. No statistical differences were observed regarding the baseline characteristics of the two groups. Three months after the surgery, the misalignment of the toric IOL was significantly lower in the image-guided group (2.18°±0.65°, range: 1.26°–3.95°) than in the manual; marking group (4.72°±0.74°, range: 3.44°–6.21°; P<0.001). Conclusion: In comparison to the manual-marking method, the image-guided system reduced the misalignment of a four-haptic hydrophobic monofocal toric IOL.\",\"PeriodicalId\":12216,\"journal\":{\"name\":\"Eye & Contact Lens: Science & Clinical Practice\",\"volume\":\"30 1\",\"pages\":\"396 - 402\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-05-17\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Eye & Contact Lens: Science & Clinical Practice\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1097/ICL.0000000000000901\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Eye & Contact Lens: Science & Clinical Practice","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1097/ICL.0000000000000901","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的:比较图像引导对准系统与手动标记方法在四触感疏水单焦环形人工晶状体(IOL)中的性能差异。方法:对2020年5月至2021年4月行四触觉疏水单焦点环形人工晶状体植入术的患者病历及3个月访视资料进行调查。比较两组人工晶状体不对准、剩余散光和平均预测误差。结果:本研究纳入44例患者的49只眼(女性占68%;平均年龄:67.2±7.0岁(范围:47-82岁)。26例29眼采用图像引导人工晶状体植入术,18例20眼采用人工标记法植入术。两组的基线特征无统计学差异。术后3个月,图像引导组环形人工晶状体的不对准度(2.18°±0.65°,范围:1.26°-3.95°)明显低于手动组;标记组(4.72°±0.74°,范围:3.44°-6.21°;P < 0.001)。结论:与手工标记方法相比,图像引导系统减少了四触感疏水单焦环形人工晶状体的错位。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Efficacy of the Image-Guided Alignment System for a Four-Haptic Hydrophobic Monofocal Toric Intraocular Lens
Objective: To compare the difference in performance between the image-guided alignment system and the manual-marking method in the four-haptic hydrophobic monofocal toric intraocular lens (IOL). Methods: Medical records of patients who underwent cataract surgery with a four-haptic hydrophobic monofocal toric IOL implantation between May 2020 and April 2021 and with 3-month visit data available were investigated. Toric IOL misalignment, residual astigmatism, and mean prediction errors between the two groups were compared. Results: This study included 49 eyes of 44 patients (women: 68%; mean age: 67.2±7.0 [range: 47–82] years). Twenty-nine eyes of 26 patients were treated with toric IOL implantation using the image-guided system and 20 eyes of 18 patients were treated using the manual-marking method. No statistical differences were observed regarding the baseline characteristics of the two groups. Three months after the surgery, the misalignment of the toric IOL was significantly lower in the image-guided group (2.18°±0.65°, range: 1.26°–3.95°) than in the manual; marking group (4.72°±0.74°, range: 3.44°–6.21°; P<0.001). Conclusion: In comparison to the manual-marking method, the image-guided system reduced the misalignment of a four-haptic hydrophobic monofocal toric IOL.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信