{"title":"磁共振成像(MRI)烧伤机制——简单环路可诱发热损伤。","authors":"T. Nakamura, K. Fukuda, K. Hayakawa, I. Aoki, K. Matsumoto, T. Sekine, H. Ueda, Y. Shimizu","doi":"10.1163/156855701316922298","DOIUrl":null,"url":null,"abstract":"To determine the mechanism of burn injury associated with magnetic resonance imaging (MRI), the induced current in the loops of a conductive lead was measured. Seven types of loops with effective areas within the range from 100 to 12 000 cm2 were made and then each loop was placed in the bore of a whole-body MR system at 0.5 T and MRI was carried out. During radio frequency (RF) irradiation, an induced voltage was observed in loops that were placed with their axes parallel to the linearly polarized transmitting RF field. The voltage had a sincfunction characteristic and was within the range 55-235 V at the RF pulse sequence for usual MR imaging. When the axis was vertical, negligible current was observed. A resistor inserted into the circuit of a 30 x 40 cm loop sparked and burned out. Simple loops of conductive material may result in the induction of a large and potentially hazardous voltage in the imaging system. Self-resonance of such a loop may add greatly to hazards by increasing the effective coupling to the RF transmitter. Also, impedance matching of the inserted resistor with the impedance of the loop increases the dissipated power at the resistor.","PeriodicalId":77139,"journal":{"name":"Frontiers of medical and biological engineering : the international journal of the Japan Society of Medical Electronics and Biological Engineering","volume":"2 1","pages":"117-29"},"PeriodicalIF":0.0000,"publicationDate":"2001-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"40","resultStr":"{\"title\":\"Mechanism of burn injury during magnetic resonance imaging (MRI)--simple loops can induce heat injury.\",\"authors\":\"T. Nakamura, K. Fukuda, K. Hayakawa, I. Aoki, K. Matsumoto, T. Sekine, H. Ueda, Y. Shimizu\",\"doi\":\"10.1163/156855701316922298\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"To determine the mechanism of burn injury associated with magnetic resonance imaging (MRI), the induced current in the loops of a conductive lead was measured. Seven types of loops with effective areas within the range from 100 to 12 000 cm2 were made and then each loop was placed in the bore of a whole-body MR system at 0.5 T and MRI was carried out. During radio frequency (RF) irradiation, an induced voltage was observed in loops that were placed with their axes parallel to the linearly polarized transmitting RF field. The voltage had a sincfunction characteristic and was within the range 55-235 V at the RF pulse sequence for usual MR imaging. When the axis was vertical, negligible current was observed. A resistor inserted into the circuit of a 30 x 40 cm loop sparked and burned out. Simple loops of conductive material may result in the induction of a large and potentially hazardous voltage in the imaging system. Self-resonance of such a loop may add greatly to hazards by increasing the effective coupling to the RF transmitter. Also, impedance matching of the inserted resistor with the impedance of the loop increases the dissipated power at the resistor.\",\"PeriodicalId\":77139,\"journal\":{\"name\":\"Frontiers of medical and biological engineering : the international journal of the Japan Society of Medical Electronics and Biological Engineering\",\"volume\":\"2 1\",\"pages\":\"117-29\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2001-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"40\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Frontiers of medical and biological engineering : the international journal of the Japan Society of Medical Electronics and Biological Engineering\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1163/156855701316922298\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Frontiers of medical and biological engineering : the international journal of the Japan Society of Medical Electronics and Biological Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1163/156855701316922298","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 40

摘要

为了确定与磁共振成像(MRI)相关的烧伤机制,测量了导电导线回路中的感应电流。制作了7种有效面积在100 ~ 12000 cm2范围内的环路,然后将每个环路放置在0.5 T的全身MR系统孔中并进行MRI。在射频(RF)辐照期间,在与线极化发射射频场平行的轴上放置的环路中观察到感应电压。电压具有自函数特性,在常规MR成像的RF脉冲序列下,电压在55-235 V范围内。当轴垂直时,观察到的电流可以忽略不计。插入30 × 40厘米回路的电阻器引发并烧毁。在成像系统中,简单的导电材料回路可能导致感应到一个大的和潜在危险的电压。这种回路的自共振会增加与射频发射机的有效耦合,从而大大增加危险。此外,插入电阻与回路阻抗的阻抗匹配增加了电阻处的耗散功率。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Mechanism of burn injury during magnetic resonance imaging (MRI)--simple loops can induce heat injury.
To determine the mechanism of burn injury associated with magnetic resonance imaging (MRI), the induced current in the loops of a conductive lead was measured. Seven types of loops with effective areas within the range from 100 to 12 000 cm2 were made and then each loop was placed in the bore of a whole-body MR system at 0.5 T and MRI was carried out. During radio frequency (RF) irradiation, an induced voltage was observed in loops that were placed with their axes parallel to the linearly polarized transmitting RF field. The voltage had a sincfunction characteristic and was within the range 55-235 V at the RF pulse sequence for usual MR imaging. When the axis was vertical, negligible current was observed. A resistor inserted into the circuit of a 30 x 40 cm loop sparked and burned out. Simple loops of conductive material may result in the induction of a large and potentially hazardous voltage in the imaging system. Self-resonance of such a loop may add greatly to hazards by increasing the effective coupling to the RF transmitter. Also, impedance matching of the inserted resistor with the impedance of the loop increases the dissipated power at the resistor.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信