{"title":"廉价微波炉测量仪器:评价","authors":"Jr. Donald M. Witters, W. Herman","doi":"10.1080/16070658.1984.11689358","DOIUrl":null,"url":null,"abstract":"Instruments were tested. Depending upon the oven modulation and leak location these tests show that the Micromate will yield a warning reading at levels from 0.6 to 1.9 mW/cm/sup 2/, when the effects of modulation sensitivity and polarization ellipticity are considered. It is to be expected, then, that this instrument will yield warning readings even when the oven emission being measured lies well within applicable standards. Its apparent susceptibility to overload damage, however, could decrease its sensitivity with no clue to the user as to what had happened. Given the same considerations and parameters, the tests show that the Guard-Rod will yield a warning reading at levels from 3.7 to 47 mW/cm/sup 2/, depending on which measurement surface is used. This instrument will sometimes yield warnings for oven emissions lying within applicable standards, and sometimes will fail to yield such warnings for emissions 10 times greater than allowable levels. The same considerations show that the Interceptor/Microscan units will light up at levels from as low as 0.7 to 0.8 mW/cm/sup 2/ or may require levels of 7 to 9 mW/cm/sup 2/. At levels of 28 to 52 mW/cm/sup 2/ or higher, however, the warning LED does not light. If partmore » of the field is cross-polarized, this threshold of extinction may be greater. In any case, there is no apparent way for the user to discern the presence of some potentially hazardous fields, and no way to distinguish such fields from a very low-level field. In summary, instruments of the Interceptor/Microscan design may light up at very low power densities, but fail to light at very high power densities. Clearly, this characteristic is extremely undesirable and potentially dangerous to a user relying on such a device.All of these values assume that the instruments are used in accordance with manufacturer's instructions. Otherwise, results may be worse. Moreover, the range of values given take little or no account of unit-to-unit variation, which may be considerable.« less","PeriodicalId":76653,"journal":{"name":"The Journal of microwave power","volume":"19 1","pages":"119-126"},"PeriodicalIF":0.0000,"publicationDate":"1984-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"Inexpensive Microwave Oven Survey Instruments: An Evaluation\",\"authors\":\"Jr. Donald M. Witters, W. Herman\",\"doi\":\"10.1080/16070658.1984.11689358\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Instruments were tested. Depending upon the oven modulation and leak location these tests show that the Micromate will yield a warning reading at levels from 0.6 to 1.9 mW/cm/sup 2/, when the effects of modulation sensitivity and polarization ellipticity are considered. It is to be expected, then, that this instrument will yield warning readings even when the oven emission being measured lies well within applicable standards. Its apparent susceptibility to overload damage, however, could decrease its sensitivity with no clue to the user as to what had happened. Given the same considerations and parameters, the tests show that the Guard-Rod will yield a warning reading at levels from 3.7 to 47 mW/cm/sup 2/, depending on which measurement surface is used. This instrument will sometimes yield warnings for oven emissions lying within applicable standards, and sometimes will fail to yield such warnings for emissions 10 times greater than allowable levels. The same considerations show that the Interceptor/Microscan units will light up at levels from as low as 0.7 to 0.8 mW/cm/sup 2/ or may require levels of 7 to 9 mW/cm/sup 2/. At levels of 28 to 52 mW/cm/sup 2/ or higher, however, the warning LED does not light. If partmore » of the field is cross-polarized, this threshold of extinction may be greater. In any case, there is no apparent way for the user to discern the presence of some potentially hazardous fields, and no way to distinguish such fields from a very low-level field. In summary, instruments of the Interceptor/Microscan design may light up at very low power densities, but fail to light at very high power densities. Clearly, this characteristic is extremely undesirable and potentially dangerous to a user relying on such a device.All of these values assume that the instruments are used in accordance with manufacturer's instructions. Otherwise, results may be worse. Moreover, the range of values given take little or no account of unit-to-unit variation, which may be considerable.« less\",\"PeriodicalId\":76653,\"journal\":{\"name\":\"The Journal of microwave power\",\"volume\":\"19 1\",\"pages\":\"119-126\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1984-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"The Journal of microwave power\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1080/16070658.1984.11689358\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"The Journal of microwave power","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1080/16070658.1984.11689358","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

摘要

测试了仪器。根据烘箱调制和泄漏位置的不同,这些测试表明,当考虑调制灵敏度和极化椭圆率的影响时,Micromate将产生0.6至1.9 mW/cm/sup 2/的警告读数。因此,可以预期的是,即使被测量的烤箱排放完全在适用的标准范围内,该仪器也会产生警告读数。然而,它对过载损坏的明显敏感性可能会降低它的灵敏度,而用户却不知道发生了什么。考虑到相同的考虑因素和参数,测试表明,根据使用的测量表面,Guard-Rod将产生3.7至47 mW/cm/sup 2/的警告读数。该仪器有时会对在适用标准范围内的烤箱排放发出警告,有时会对超过允许水平10倍的排放发出警告。同样的考虑表明,Interceptor/Microscan装置将在低至0.7至0.8 mW/cm/sup 2/或可能需要7至9 mW/cm/sup 2/的电平下点亮。然而,在28至52 mW/cm/sup 2/或更高的电平时,警告LED不亮。如果场的一部分是交叉极化的,这个消光阈值可能更大。在任何情况下,用户都没有明显的方法来辨别某些潜在危险字段的存在,也没有办法将这些字段与非常低级别的字段区分开来。总之,Interceptor/Microscan设计的仪器可以在非常低的功率密度下发光,但不能在非常高的功率密度下发光。显然,这种特性对于依赖这种设备的用户来说是非常不可取的,并且具有潜在的危险。所有这些数值都假定仪器是按照制造商的说明使用的。否则,结果可能会更糟。此外,给出的数值范围很少或根本没有考虑到单位间的变化,这可能是相当大的。«少
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Inexpensive Microwave Oven Survey Instruments: An Evaluation
Instruments were tested. Depending upon the oven modulation and leak location these tests show that the Micromate will yield a warning reading at levels from 0.6 to 1.9 mW/cm/sup 2/, when the effects of modulation sensitivity and polarization ellipticity are considered. It is to be expected, then, that this instrument will yield warning readings even when the oven emission being measured lies well within applicable standards. Its apparent susceptibility to overload damage, however, could decrease its sensitivity with no clue to the user as to what had happened. Given the same considerations and parameters, the tests show that the Guard-Rod will yield a warning reading at levels from 3.7 to 47 mW/cm/sup 2/, depending on which measurement surface is used. This instrument will sometimes yield warnings for oven emissions lying within applicable standards, and sometimes will fail to yield such warnings for emissions 10 times greater than allowable levels. The same considerations show that the Interceptor/Microscan units will light up at levels from as low as 0.7 to 0.8 mW/cm/sup 2/ or may require levels of 7 to 9 mW/cm/sup 2/. At levels of 28 to 52 mW/cm/sup 2/ or higher, however, the warning LED does not light. If partmore » of the field is cross-polarized, this threshold of extinction may be greater. In any case, there is no apparent way for the user to discern the presence of some potentially hazardous fields, and no way to distinguish such fields from a very low-level field. In summary, instruments of the Interceptor/Microscan design may light up at very low power densities, but fail to light at very high power densities. Clearly, this characteristic is extremely undesirable and potentially dangerous to a user relying on such a device.All of these values assume that the instruments are used in accordance with manufacturer's instructions. Otherwise, results may be worse. Moreover, the range of values given take little or no account of unit-to-unit variation, which may be considerable.« less
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信