孟加拉国原子能委员会核医学物理研究所PET-CT设施的辐射屏蔽分析和设计,孟加拉国

M. F. Uddin, R. Khatun, S. Akter, H. M. Jamil, A. N. Monika, Mansur Rahaman, Rasel Das, R. Sharmin, M. Rahman, M. M. Ahasan
{"title":"孟加拉国原子能委员会核医学物理研究所PET-CT设施的辐射屏蔽分析和设计,孟加拉国","authors":"M. F. Uddin, R. Khatun, S. Akter, H. M. Jamil, A. N. Monika, Mansur Rahaman, Rasel Das, R. Sharmin, M. Rahman, M. M. Ahasan","doi":"10.4236/IJMPCERO.2019.81001","DOIUrl":null,"url":null,"abstract":"High \nresolution (4 mm) tof PET-CT (positron emission tomography-computed tomography) from Philips of model Ingenuity TF is newly installed at Institute \nof Nuclear Medical Physics (INMP). 128 slice CT component incorporated with PET \nprovides comparatively lower dose than the 511 keV annihilation photons \nassociated with positron decay from PET scan. So, for designing shielding in \nour PET-CT facility, only 511 keV annihilation photons energy has been \nconsidered. The main objective of this paper is to show what measures have been \ntaken to protect patients, occupational workers as well as environment from \nPET-CT radiation hazard through a cost effective design that satisfy the \nnational regulatory demand. In this paper, AAPM (American Associations of \nPhysicists in Medicine) Task group 108 analysis for PET and PET-CT shielding \nrequirements is followed for our PET-CT facility shielding design. From theoretical calculation as shielding \nrequirement, 1.1 cm Pb thickness or, 13 cm concrete thicknesses are \nfound. Practically, all walls and ceiling are of 30.48 cm (1 foot) thick made \nof concrete with density 2.35 gcm-3 for more safety. As x-ray from \nCT is not \ntaken into account for shielding analysis, Bangladesh Atomic Energy Commission \n(BAEC) conducted an extensive radiation survey at controlled, supervised and \npublic area for CT. The report that is found meets the national regulatory \nrequirements.","PeriodicalId":14028,"journal":{"name":"International Journal of Medical Physics, Clinical Engineering and Radiation Oncology","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2019-01-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"Radiation Shielding Analysis and Design of tof PET-CT Facility at Institute of Nuclear Medical Physics under Bangladesh Atomic Energy Commission, Bangladesh\",\"authors\":\"M. F. Uddin, R. Khatun, S. Akter, H. M. Jamil, A. N. Monika, Mansur Rahaman, Rasel Das, R. Sharmin, M. Rahman, M. M. Ahasan\",\"doi\":\"10.4236/IJMPCERO.2019.81001\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"High \\nresolution (4 mm) tof PET-CT (positron emission tomography-computed tomography) from Philips of model Ingenuity TF is newly installed at Institute \\nof Nuclear Medical Physics (INMP). 128 slice CT component incorporated with PET \\nprovides comparatively lower dose than the 511 keV annihilation photons \\nassociated with positron decay from PET scan. So, for designing shielding in \\nour PET-CT facility, only 511 keV annihilation photons energy has been \\nconsidered. The main objective of this paper is to show what measures have been \\ntaken to protect patients, occupational workers as well as environment from \\nPET-CT radiation hazard through a cost effective design that satisfy the \\nnational regulatory demand. In this paper, AAPM (American Associations of \\nPhysicists in Medicine) Task group 108 analysis for PET and PET-CT shielding \\nrequirements is followed for our PET-CT facility shielding design. From theoretical calculation as shielding \\nrequirement, 1.1 cm Pb thickness or, 13 cm concrete thicknesses are \\nfound. Practically, all walls and ceiling are of 30.48 cm (1 foot) thick made \\nof concrete with density 2.35 gcm-3 for more safety. As x-ray from \\nCT is not \\ntaken into account for shielding analysis, Bangladesh Atomic Energy Commission \\n(BAEC) conducted an extensive radiation survey at controlled, supervised and \\npublic area for CT. The report that is found meets the national regulatory \\nrequirements.\",\"PeriodicalId\":14028,\"journal\":{\"name\":\"International Journal of Medical Physics, Clinical Engineering and Radiation Oncology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2019-01-21\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International Journal of Medical Physics, Clinical Engineering and Radiation Oncology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4236/IJMPCERO.2019.81001\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Medical Physics, Clinical Engineering and Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4236/IJMPCERO.2019.81001","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

摘要

高分辨率(4毫米)的PET-CT(正电子发射断层扫描-计算机断层扫描)从飞利浦的独创性TF模型是新安装在核医学物理研究所(INMP)。与PET扫描中与正电子衰变相关的511 keV湮灭光子相比,结合PET的128层CT组件提供了相对较低的剂量。因此,在PET-CT设备的屏蔽设计中,仅考虑511 keV湮灭光子能量。本文的主要目的是通过符合国家监管要求的具有成本效益的设计,展示采取了哪些措施来保护患者、职业工人和环境免受PET-CT辐射危害。本文根据AAPM(美国医学物理学家协会)任务小组108对PET和PET- ct屏蔽要求的分析,对我们的PET- ct设备屏蔽设计进行了分析。根据理论计算得出屏蔽要求为1.1 cm的铅厚度或13 cm的混凝土厚度。实际上,所有的墙壁和天花板都是30.48厘米(1英尺)厚,由密度为2.35 gcm-3的混凝土制成,以提高安全性。由于CT的x射线没有被考虑到屏蔽分析,孟加拉国原子能委员会(BAEC)在控制、监督和公共区域对CT进行了广泛的辐射调查。发现的报告符合国家法规要求。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Radiation Shielding Analysis and Design of tof PET-CT Facility at Institute of Nuclear Medical Physics under Bangladesh Atomic Energy Commission, Bangladesh
High resolution (4 mm) tof PET-CT (positron emission tomography-computed tomography) from Philips of model Ingenuity TF is newly installed at Institute of Nuclear Medical Physics (INMP). 128 slice CT component incorporated with PET provides comparatively lower dose than the 511 keV annihilation photons associated with positron decay from PET scan. So, for designing shielding in our PET-CT facility, only 511 keV annihilation photons energy has been considered. The main objective of this paper is to show what measures have been taken to protect patients, occupational workers as well as environment from PET-CT radiation hazard through a cost effective design that satisfy the national regulatory demand. In this paper, AAPM (American Associations of Physicists in Medicine) Task group 108 analysis for PET and PET-CT shielding requirements is followed for our PET-CT facility shielding design. From theoretical calculation as shielding requirement, 1.1 cm Pb thickness or, 13 cm concrete thicknesses are found. Practically, all walls and ceiling are of 30.48 cm (1 foot) thick made of concrete with density 2.35 gcm-3 for more safety. As x-ray from CT is not taken into account for shielding analysis, Bangladesh Atomic Energy Commission (BAEC) conducted an extensive radiation survey at controlled, supervised and public area for CT. The report that is found meets the national regulatory requirements.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信