{"title":"在线流动注射电感耦合等离子体原子发射光谱法,利用高比表面积tio_2进行预富集和后续测定","authors":"E. Vassileva","doi":"10.1051/ANALUSIS:2000160","DOIUrl":null,"url":null,"abstract":"A flow-injection analysis (FIA) system incorporating two microcolumns of TiO 2 was used for the developing of an on- line method for preconcentration and sequential determination of Cr 3+ and Cr6+ species by inductively coupled plasma atomic emission spectrometry (ICP-AES). The optimal conditions for both analytes have been obtained. With a sample volume of 10-50 ml and an elution volume of 500 µl, a signal was 20-100 times higher than the signal given by a continuous aspiration system. The method gave good reproducibility with precisions less than 7 % RSD for 5 replicates at the 10 ng/ml level for both Cr- species. Recoveries between 95 and 98 % were obtained for Cr 3+ and Cr6+. The preconcentration system and the methodology proposed were validated on several natural water samples, on reference water samples as well as on synthetic seawater. It was found that the on-line detection using ICP-AES allows sequential Cr speciation analysis at the low concentration levels (ng/l) usually encountered in these media.","PeriodicalId":8221,"journal":{"name":"Analusis","volume":"40 1","pages":"878-884"},"PeriodicalIF":0.0000,"publicationDate":"2000-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"10","resultStr":"{\"title\":\"Use of high surface area TiO$_2$ for preconcentration and following determination of Cr species by on-line flow injection inductively coupled plasma atomic emission spectrometry\",\"authors\":\"E. Vassileva\",\"doi\":\"10.1051/ANALUSIS:2000160\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"A flow-injection analysis (FIA) system incorporating two microcolumns of TiO 2 was used for the developing of an on- line method for preconcentration and sequential determination of Cr 3+ and Cr6+ species by inductively coupled plasma atomic emission spectrometry (ICP-AES). The optimal conditions for both analytes have been obtained. With a sample volume of 10-50 ml and an elution volume of 500 µl, a signal was 20-100 times higher than the signal given by a continuous aspiration system. The method gave good reproducibility with precisions less than 7 % RSD for 5 replicates at the 10 ng/ml level for both Cr- species. Recoveries between 95 and 98 % were obtained for Cr 3+ and Cr6+. The preconcentration system and the methodology proposed were validated on several natural water samples, on reference water samples as well as on synthetic seawater. It was found that the on-line detection using ICP-AES allows sequential Cr speciation analysis at the low concentration levels (ng/l) usually encountered in these media.\",\"PeriodicalId\":8221,\"journal\":{\"name\":\"Analusis\",\"volume\":\"40 1\",\"pages\":\"878-884\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2000-11-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"10\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Analusis\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1051/ANALUSIS:2000160\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Analusis","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1051/ANALUSIS:2000160","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 10

摘要

采用两微柱的流动注射分析(FIA)系统,建立了电感耦合等离子体原子发射光谱(ICP-AES)在线预富集和序列测定cr3 +和Cr6+的方法。得到了两种分析物的最佳条件。当样品体积为10-50 ml,洗脱体积为500µl时,信号比连续抽吸系统高20-100倍。在10 ng/ml浓度下,该方法重复性好,精密度小于7% RSD,重复5次。cr3 +和Cr6+的回收率在95 ~ 98%之间。在几种天然水样、参考水样和合成海水中对预富集系统和方法进行了验证。发现使用ICP-AES在线检测可以在这些介质中通常遇到的低浓度水平(ng/l)下进行顺序的Cr形态分析。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Use of high surface area TiO$_2$ for preconcentration and following determination of Cr species by on-line flow injection inductively coupled plasma atomic emission spectrometry
A flow-injection analysis (FIA) system incorporating two microcolumns of TiO 2 was used for the developing of an on- line method for preconcentration and sequential determination of Cr 3+ and Cr6+ species by inductively coupled plasma atomic emission spectrometry (ICP-AES). The optimal conditions for both analytes have been obtained. With a sample volume of 10-50 ml and an elution volume of 500 µl, a signal was 20-100 times higher than the signal given by a continuous aspiration system. The method gave good reproducibility with precisions less than 7 % RSD for 5 replicates at the 10 ng/ml level for both Cr- species. Recoveries between 95 and 98 % were obtained for Cr 3+ and Cr6+. The preconcentration system and the methodology proposed were validated on several natural water samples, on reference water samples as well as on synthetic seawater. It was found that the on-line detection using ICP-AES allows sequential Cr speciation analysis at the low concentration levels (ng/l) usually encountered in these media.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信