{"title":"设计了一种简单的马弗炉,用于灰分分析中的温度优化","authors":"F. R. Abadi","doi":"10.18860/NEU.V10I2.4559","DOIUrl":null,"url":null,"abstract":"<p class=\"abstrak\">Analysis of the ash content in biomass conducted through the incineration process in the muffle furnace at a given temperature of 575<span style=\"text-decoration: underline;\">+</span>25<sup>o</sup>C. The result depends on the accuracy level on the temperature sensor in the muffle furnace. It was attempted to achieve the optimum temperature and to detect the muffle furnace temperature by installing the thermocouple sensor type K directly  the porcelain sample plate. It yielded that by direct incineration process and gradual process the setup point reached in 60-90 minutes and 150-180 minutes before attaining the steady temperature, respectively. The temperature distributed uniformly for all treatments at <span style=\"text-decoration: underline;\">+</span> 6.61 <sup>o</sup>C. The temperature measurement accuracy was <span style=\"text-decoration: underline;\">+</span> 1.18%, and the overall temperature gradient was <span style=\"text-decoration: underline;\">+</span> 28.69 <sup>o</sup>C. Moreover, there was no difference in the temperature fluctuation among the treatments at the temperature of  <span style=\"text-decoration: underline;\">+</span> 20.95 <sup>o</sup>C. The optimum temperature of the chamber was  588 <sup>o</sup>C for analyzing ash content of biomass.</p>","PeriodicalId":17685,"journal":{"name":"Jurnal Neutrino","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2018-07-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"DESIGN OF A SIMPLE MUFFLE FURNACE FOR TEMPERATURE OPTIMIZATION IN ASH CONTENT ANALYSIS\",\"authors\":\"F. R. Abadi\",\"doi\":\"10.18860/NEU.V10I2.4559\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p class=\\\"abstrak\\\">Analysis of the ash content in biomass conducted through the incineration process in the muffle furnace at a given temperature of 575<span style=\\\"text-decoration: underline;\\\">+</span>25<sup>o</sup>C. The result depends on the accuracy level on the temperature sensor in the muffle furnace. It was attempted to achieve the optimum temperature and to detect the muffle furnace temperature by installing the thermocouple sensor type K directly  the porcelain sample plate. It yielded that by direct incineration process and gradual process the setup point reached in 60-90 minutes and 150-180 minutes before attaining the steady temperature, respectively. The temperature distributed uniformly for all treatments at <span style=\\\"text-decoration: underline;\\\">+</span> 6.61 <sup>o</sup>C. The temperature measurement accuracy was <span style=\\\"text-decoration: underline;\\\">+</span> 1.18%, and the overall temperature gradient was <span style=\\\"text-decoration: underline;\\\">+</span> 28.69 <sup>o</sup>C. Moreover, there was no difference in the temperature fluctuation among the treatments at the temperature of  <span style=\\\"text-decoration: underline;\\\">+</span> 20.95 <sup>o</sup>C. The optimum temperature of the chamber was  588 <sup>o</sup>C for analyzing ash content of biomass.</p>\",\"PeriodicalId\":17685,\"journal\":{\"name\":\"Jurnal Neutrino\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2018-07-18\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Jurnal Neutrino\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.18860/NEU.V10I2.4559\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Jurnal Neutrino","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.18860/NEU.V10I2.4559","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

在给定温度575+25℃条件下,对马弗炉内焚烧过程中生物质灰分含量进行分析。结果取决于马弗炉温度传感器的精度水平。将K型热电偶传感器直接安装在瓷样板上,试图达到最佳温度并检测马弗炉温度。结果表明,直接焚烧法和渐进式焚烧法分别在60 ~ 90分钟和150 ~ 180分钟达到温度稳定点。在+ 6.61℃时,各处理温度分布均匀。测温精度为+ 1.18%,总温度梯度为+ 28.69 oC。在+ 20.95 oC温度下,各处理间的温度波动无显著差异。测定生物质灰分的最佳温度为588℃。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
DESIGN OF A SIMPLE MUFFLE FURNACE FOR TEMPERATURE OPTIMIZATION IN ASH CONTENT ANALYSIS

Analysis of the ash content in biomass conducted through the incineration process in the muffle furnace at a given temperature of 575+25oC. The result depends on the accuracy level on the temperature sensor in the muffle furnace. It was attempted to achieve the optimum temperature and to detect the muffle furnace temperature by installing the thermocouple sensor type K directly  the porcelain sample plate. It yielded that by direct incineration process and gradual process the setup point reached in 60-90 minutes and 150-180 minutes before attaining the steady temperature, respectively. The temperature distributed uniformly for all treatments at + 6.61 oC. The temperature measurement accuracy was + 1.18%, and the overall temperature gradient was + 28.69 oC. Moreover, there was no difference in the temperature fluctuation among the treatments at the temperature of  + 20.95 oC. The optimum temperature of the chamber was  588 oC for analyzing ash content of biomass.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信