临床测定红细胞沉降率的新型MINI-CUBE的验证。

IF 1.3 Q4 HEMATOLOGY
Journal of hematology Pub Date : 2023-10-01 Epub Date: 2023-10-25 DOI:10.14740/jh1165
Flaminia Tomassetti, Martina Pelagalli, Eleonora Nicolai, Serena Sarubbi, Cinzia Calabrese, Alfredo Giovannelli, Silvia Codella, Gemma Viola, Daniela Diamanti, Renato Massoud, Sergio Bernardini, Massimo Pieri
{"title":"临床测定红细胞沉降率的新型MINI-CUBE的验证。","authors":"Flaminia Tomassetti, Martina Pelagalli, Eleonora Nicolai, Serena Sarubbi, Cinzia Calabrese, Alfredo Giovannelli, Silvia Codella, Gemma Viola, Daniela Diamanti, Renato Massoud, Sergio Bernardini, Massimo Pieri","doi":"10.14740/jh1165","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Erythrocyte sedimentation rate (ESR) indirectly measures blood fibrinogen, and it is altered by all those pathological conditions that modify the aggregation of red blood cells. The international guidelines by the International Council for Standardization in Hematology (ICSH) define the Westergren method as the gold standard for ESR, although it is completely operator-dependent, time-consuming, and requires a patient's blood consumption. Therefore, the validation of new ESR analyzers is needed. The aim of this study is the validation of a new automated ESR analyzer, MINI-CUBE (DIESSE, Diagnostica Senese, Italy).</p><p><strong>Methods: </strong>The samples (n = 270) were collected at the University Hospital of the University of Rome Tor Vergata. A comparison between the automated instrument and the gold standard was performed. Statistical analyses were processed by MedCalc software.</p><p><strong>Results: </strong>The comparison analysis performed on the overall samples reported a good agreement, showing a Spearman rank correlation coefficient of 0.94 (P < 0.001), compared to the Westergren test. The Bland-Altman analysis showed a mean bias of 1.5 (maximum (max.):19.6; minimum (min.): -16.6). Inter-run (level 1 coefficient of variation (CV): 4.9%; level 2 CV: 0.8%), intra-run (level 1 CV: 21.1%; level 2 CV: 3.2%), and inter-instrument (level 1 CV: 27.1%; level 2 CV: 5.6%) precision were also assessed. The hematocrit value did not interfere with the analysis: Spearman rank correlation coefficient of 0.929 (P < 0.001); mean bias of 1.3 (max.:18.3; min.: -15.6).</p><p><strong>Conclusions: </strong>Overall results from MINI-CUBE asserted a good correlation rate with the gold standard, and it could be considered an accurate, and objective alternative for the Westergren test.</p>","PeriodicalId":15964,"journal":{"name":"Journal of hematology","volume":null,"pages":null},"PeriodicalIF":1.3000,"publicationDate":"2023-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10627362/pdf/","citationCount":"0","resultStr":"{\"title\":\"Validation of the New MINI-CUBE for Clinic Determination of Erythrocyte Sedimentation Rate.\",\"authors\":\"Flaminia Tomassetti, Martina Pelagalli, Eleonora Nicolai, Serena Sarubbi, Cinzia Calabrese, Alfredo Giovannelli, Silvia Codella, Gemma Viola, Daniela Diamanti, Renato Massoud, Sergio Bernardini, Massimo Pieri\",\"doi\":\"10.14740/jh1165\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Erythrocyte sedimentation rate (ESR) indirectly measures blood fibrinogen, and it is altered by all those pathological conditions that modify the aggregation of red blood cells. The international guidelines by the International Council for Standardization in Hematology (ICSH) define the Westergren method as the gold standard for ESR, although it is completely operator-dependent, time-consuming, and requires a patient's blood consumption. Therefore, the validation of new ESR analyzers is needed. The aim of this study is the validation of a new automated ESR analyzer, MINI-CUBE (DIESSE, Diagnostica Senese, Italy).</p><p><strong>Methods: </strong>The samples (n = 270) were collected at the University Hospital of the University of Rome Tor Vergata. A comparison between the automated instrument and the gold standard was performed. Statistical analyses were processed by MedCalc software.</p><p><strong>Results: </strong>The comparison analysis performed on the overall samples reported a good agreement, showing a Spearman rank correlation coefficient of 0.94 (P < 0.001), compared to the Westergren test. The Bland-Altman analysis showed a mean bias of 1.5 (maximum (max.):19.6; minimum (min.): -16.6). Inter-run (level 1 coefficient of variation (CV): 4.9%; level 2 CV: 0.8%), intra-run (level 1 CV: 21.1%; level 2 CV: 3.2%), and inter-instrument (level 1 CV: 27.1%; level 2 CV: 5.6%) precision were also assessed. The hematocrit value did not interfere with the analysis: Spearman rank correlation coefficient of 0.929 (P < 0.001); mean bias of 1.3 (max.:18.3; min.: -15.6).</p><p><strong>Conclusions: </strong>Overall results from MINI-CUBE asserted a good correlation rate with the gold standard, and it could be considered an accurate, and objective alternative for the Westergren test.</p>\",\"PeriodicalId\":15964,\"journal\":{\"name\":\"Journal of hematology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2023-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10627362/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of hematology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.14740/jh1165\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/10/25 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q4\",\"JCRName\":\"HEMATOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of hematology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.14740/jh1165","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/10/25 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"HEMATOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

背景:红细胞沉降率(ESR)间接测量血液纤维蛋白原,它被所有改变红细胞聚集的病理条件所改变。国际血液学标准化委员会(ICSH)的国际指南将Westergren方法定义为ESR的金标准,尽管它完全依赖于操作者,耗时且需要患者的血液消耗。因此,需要对新的ESR分析仪进行验证。本研究的目的是验证一种新的自动ESR分析仪MINI-CUBE(DISSE,Diagnostica Senese,Italy)。方法:在罗马大学Tor Vergata大学医院采集样本(n=270)。对自动化仪器和金标准进行了比较。统计分析采用MedCalc软件进行处理。结果:对总体样本进行的比较分析报告了良好的一致性,与Westergren检验相比,Spearman秩相关系数为0.94(P<0.001)。Bland-Altman分析显示平均偏差为1.5(最大值:19.6;最小值(最小值):-16.6)。内部运行(一级变异系数(CV):4.9%;2级CV:0.8%)、批内(1级CV:21.1%;2级CV:3.2%)和仪器间(1级CV:27.1%;2级CV:5.6%)的精密度也进行了评估。红细胞压积值不干扰分析:Spearman秩相关系数为0.929(P<0.001);平均偏差1.3(最大值:18.3;最小值:-15.6)。结论:MINI-CUBE的总体结果与金标准具有良好的相关性,可以被认为是Westergren检验的准确、客观的替代方案。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

Validation of the New MINI-CUBE for Clinic Determination of Erythrocyte Sedimentation Rate.

Validation of the New MINI-CUBE for Clinic Determination of Erythrocyte Sedimentation Rate.

Validation of the New MINI-CUBE for Clinic Determination of Erythrocyte Sedimentation Rate.

Validation of the New MINI-CUBE for Clinic Determination of Erythrocyte Sedimentation Rate.

Background: Erythrocyte sedimentation rate (ESR) indirectly measures blood fibrinogen, and it is altered by all those pathological conditions that modify the aggregation of red blood cells. The international guidelines by the International Council for Standardization in Hematology (ICSH) define the Westergren method as the gold standard for ESR, although it is completely operator-dependent, time-consuming, and requires a patient's blood consumption. Therefore, the validation of new ESR analyzers is needed. The aim of this study is the validation of a new automated ESR analyzer, MINI-CUBE (DIESSE, Diagnostica Senese, Italy).

Methods: The samples (n = 270) were collected at the University Hospital of the University of Rome Tor Vergata. A comparison between the automated instrument and the gold standard was performed. Statistical analyses were processed by MedCalc software.

Results: The comparison analysis performed on the overall samples reported a good agreement, showing a Spearman rank correlation coefficient of 0.94 (P < 0.001), compared to the Westergren test. The Bland-Altman analysis showed a mean bias of 1.5 (maximum (max.):19.6; minimum (min.): -16.6). Inter-run (level 1 coefficient of variation (CV): 4.9%; level 2 CV: 0.8%), intra-run (level 1 CV: 21.1%; level 2 CV: 3.2%), and inter-instrument (level 1 CV: 27.1%; level 2 CV: 5.6%) precision were also assessed. The hematocrit value did not interfere with the analysis: Spearman rank correlation coefficient of 0.929 (P < 0.001); mean bias of 1.3 (max.:18.3; min.: -15.6).

Conclusions: Overall results from MINI-CUBE asserted a good correlation rate with the gold standard, and it could be considered an accurate, and objective alternative for the Westergren test.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of hematology
Journal of hematology HEMATOLOGY-
自引率
0.00%
发文量
29
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信