{"title":"一种自拔式外科吸引器的机械性能:一项随机对照试验","authors":"M. Stiehl","doi":"10.15438/RR.V4I1.54","DOIUrl":null,"url":null,"abstract":"Introduction: Obstruction of the surgical suction instrument is a common problem in orthopaedic surgery. Previous attempts have tried to address this problem. The ‘Super Sucker’ has a screen compartment that can be unclogged upon disassembly. The Yankauer sucker has small holes in its tip that strain larger bits of debris. The aim of this study is to clinically evaluate a new gas-actuated suction instrument in which a special screen at its tip is cleared, as needed, by a rapid burst of pressurized carbon dioxide gas. Methods: This IRB-approved, prospective, randomized study compared a gas-actuated suction instrument with the Super Sucker and Yankauer in 70 consecutive primary total joint arthroplasty cases. Outcome measures included: incidence of complete suction loss due to suction instrument obstruction; time lost while unplugging the suction instrument; number of additional suction instruments needed; and a subjective surgeon-assessed performance score (1 to 5, with 5 being most favorable) for type of suction instrument. Results: There were no cases in which the gas-actuated suction instrument could not be rapidly cleared of debris. The Super Sucker completely plugged in 71% of cases, requiring 67 minutes total to unplug (3.9 minutes per case, range 0 to 12 minutes). In four cases, replacement Super Suckers were required to finish the case. The Yankauer completely plugged in 47% of cases, requiring 52 minutes total to unplug (2.8 minutes per case, range 0 to 10 minutes). In three cases, replacement Yankauers were required to finish the case. The average performance score was 2.7 for the Super Sucker, 3.6 for the Yankauer, and 5 for the gas-actuated suction instrument on a scale of 1 to 5, with 5 being most favorable. Discussion: This study evaluated a suction instrument in which a screen tip prevents obstruction, and a burst of pressurized carbon dioxide gas clears debris from the tip. The new suction instrument was successful in 100% of cases, with considerably less time lost compared to the other suction instruments. The gas-actuated suction tool actively and rapidly cleared obstructive debris with minimal disruption to the surgical flow. Recent clinical experience has shown the gas-actuated suction tool to be particularly enabling in the settings of tourniquet-free total knee replacement, small incision total hip replacement, bipolar hemiarthroplasty, and revision total hip replacement.","PeriodicalId":20884,"journal":{"name":"Reconstructive Review","volume":"4 1","pages":"18-22"},"PeriodicalIF":0.0000,"publicationDate":"2014-03-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Mechanical Performance of a Self- Unplugging Surgical Suction Instrument: A Randomized Controlled Trial\",\"authors\":\"M. Stiehl\",\"doi\":\"10.15438/RR.V4I1.54\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Introduction: Obstruction of the surgical suction instrument is a common problem in orthopaedic surgery. Previous attempts have tried to address this problem. The ‘Super Sucker’ has a screen compartment that can be unclogged upon disassembly. The Yankauer sucker has small holes in its tip that strain larger bits of debris. The aim of this study is to clinically evaluate a new gas-actuated suction instrument in which a special screen at its tip is cleared, as needed, by a rapid burst of pressurized carbon dioxide gas. Methods: This IRB-approved, prospective, randomized study compared a gas-actuated suction instrument with the Super Sucker and Yankauer in 70 consecutive primary total joint arthroplasty cases. Outcome measures included: incidence of complete suction loss due to suction instrument obstruction; time lost while unplugging the suction instrument; number of additional suction instruments needed; and a subjective surgeon-assessed performance score (1 to 5, with 5 being most favorable) for type of suction instrument. Results: There were no cases in which the gas-actuated suction instrument could not be rapidly cleared of debris. The Super Sucker completely plugged in 71% of cases, requiring 67 minutes total to unplug (3.9 minutes per case, range 0 to 12 minutes). In four cases, replacement Super Suckers were required to finish the case. The Yankauer completely plugged in 47% of cases, requiring 52 minutes total to unplug (2.8 minutes per case, range 0 to 10 minutes). In three cases, replacement Yankauers were required to finish the case. The average performance score was 2.7 for the Super Sucker, 3.6 for the Yankauer, and 5 for the gas-actuated suction instrument on a scale of 1 to 5, with 5 being most favorable. Discussion: This study evaluated a suction instrument in which a screen tip prevents obstruction, and a burst of pressurized carbon dioxide gas clears debris from the tip. The new suction instrument was successful in 100% of cases, with considerably less time lost compared to the other suction instruments. The gas-actuated suction tool actively and rapidly cleared obstructive debris with minimal disruption to the surgical flow. Recent clinical experience has shown the gas-actuated suction tool to be particularly enabling in the settings of tourniquet-free total knee replacement, small incision total hip replacement, bipolar hemiarthroplasty, and revision total hip replacement.\",\"PeriodicalId\":20884,\"journal\":{\"name\":\"Reconstructive Review\",\"volume\":\"4 1\",\"pages\":\"18-22\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2014-03-31\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Reconstructive Review\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.15438/RR.V4I1.54\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Reconstructive Review","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.15438/RR.V4I1.54","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

手术吸引器阻塞是骨科手术中常见的问题。以前的尝试试图解决这个问题。“超级吸盘”有一个屏幕隔间,可以在拆卸时打开。洋考尔吸盘的顶端有小孔,可以吸出较大的碎片。本研究的目的是临床评估一种新型气体驱动吸引器,在需要时,通过加压二氧化碳气体的快速爆发清除其尖端的特殊屏幕。方法:这项经irb批准的前瞻性随机研究比较了70例连续一期全关节置换术患者的气体驱动吸引器与Super Sucker和Yankauer。结果指标包括:因吸器阻塞导致完全吸力丧失的发生率;拔吸器时间浪费;需要的额外吸痰器械数量;对抽吸器的类型进行主观的外科医生评估的性能评分(1 ~ 5分,5分为最有利)。结果:无一例气动式吸痰器不能快速清除脏物。在71%的情况下,Super Sucker完全插入,总共需要67分钟(每例3.9分钟,范围为0到12分钟)。在四个案例中,需要更换超级吸盘来完成手术。47%的情况下,Yankauer完全插入电源,总共需要52分钟(每例2.8分钟,范围为0到10分钟)。在三起案件中,需要替换杨考尔来完成案件。Super Sucker的平均得分为2.7分,Yankauer的平均得分为3.6分,气体驱动吸入器的平均得分为5分,满分为5分。讨论:本研究评估了一种吸力装置,其中筛网尖端可以防止阻塞,加压二氧化碳气体的爆发可以清除尖端的碎片。新的吸引器在100%的病例中成功,与其他吸引器相比,损失的时间大大减少。气体驱动的吸引工具能够主动、快速地清除阻塞的碎片,对手术流程的干扰最小。最近的临床经验表明,气体驱动吸引工具在无止血带全膝关节置换术、小切口全髋关节置换术、双极半关节置换术和翻修全髋关节置换术中特别有效。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Mechanical Performance of a Self- Unplugging Surgical Suction Instrument: A Randomized Controlled Trial
Introduction: Obstruction of the surgical suction instrument is a common problem in orthopaedic surgery. Previous attempts have tried to address this problem. The ‘Super Sucker’ has a screen compartment that can be unclogged upon disassembly. The Yankauer sucker has small holes in its tip that strain larger bits of debris. The aim of this study is to clinically evaluate a new gas-actuated suction instrument in which a special screen at its tip is cleared, as needed, by a rapid burst of pressurized carbon dioxide gas. Methods: This IRB-approved, prospective, randomized study compared a gas-actuated suction instrument with the Super Sucker and Yankauer in 70 consecutive primary total joint arthroplasty cases. Outcome measures included: incidence of complete suction loss due to suction instrument obstruction; time lost while unplugging the suction instrument; number of additional suction instruments needed; and a subjective surgeon-assessed performance score (1 to 5, with 5 being most favorable) for type of suction instrument. Results: There were no cases in which the gas-actuated suction instrument could not be rapidly cleared of debris. The Super Sucker completely plugged in 71% of cases, requiring 67 minutes total to unplug (3.9 minutes per case, range 0 to 12 minutes). In four cases, replacement Super Suckers were required to finish the case. The Yankauer completely plugged in 47% of cases, requiring 52 minutes total to unplug (2.8 minutes per case, range 0 to 10 minutes). In three cases, replacement Yankauers were required to finish the case. The average performance score was 2.7 for the Super Sucker, 3.6 for the Yankauer, and 5 for the gas-actuated suction instrument on a scale of 1 to 5, with 5 being most favorable. Discussion: This study evaluated a suction instrument in which a screen tip prevents obstruction, and a burst of pressurized carbon dioxide gas clears debris from the tip. The new suction instrument was successful in 100% of cases, with considerably less time lost compared to the other suction instruments. The gas-actuated suction tool actively and rapidly cleared obstructive debris with minimal disruption to the surgical flow. Recent clinical experience has shown the gas-actuated suction tool to be particularly enabling in the settings of tourniquet-free total knee replacement, small incision total hip replacement, bipolar hemiarthroplasty, and revision total hip replacement.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
2
审稿时长
24 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信