eyeesys 2000测量标定球面表面高程的精度

Douglas G Horner PhD (OD, FAAO), Thomas O Salmon (OD, FAAO)
{"title":"eyeesys 2000测量标定球面表面高程的精度","authors":"Douglas G Horner PhD (OD, FAAO),&nbsp;Thomas O Salmon (OD, FAAO)","doi":"10.1016/S0892-8967(99)00005-X","DOIUrl":null,"url":null,"abstract":"<div><p>This article examines the accuracy of the EyeSys 2000 videokeratoscope in measuring aspheric surfaces for calculation of the wavefront aberrations of the anterior cornea. Six rotationally symmetric aspheric surfaces were measured. The surface elevation was computed from the normal measurement files. Both elevation error and relative error were computed from the average of the three maps. The root-mean-square errors for the various surfaces ranged from 1.48 to 6.55 microns, with less error on the oblate and spherical surfaces. The error found was very systematic, increasing monotonically toward the periphery. The article includes a strategy to compensate for the systematic error to meet the required 0.5-micron accuracy needed. An equation was developed that used only the apical radius and shape factor, which improved the accuracy to the required 0.5-micron level.</p></div>","PeriodicalId":80286,"journal":{"name":"International contact lens clinic (New York, N.Y.)","volume":"25 6","pages":"Pages 171-177"},"PeriodicalIF":0.0000,"publicationDate":"1998-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/S0892-8967(99)00005-X","citationCount":"6","resultStr":"{\"title\":\"Accuracy of the EyeSys 2000 in measuring surface elevation of calibrated aspheres\",\"authors\":\"Douglas G Horner PhD (OD, FAAO),&nbsp;Thomas O Salmon (OD, FAAO)\",\"doi\":\"10.1016/S0892-8967(99)00005-X\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><p>This article examines the accuracy of the EyeSys 2000 videokeratoscope in measuring aspheric surfaces for calculation of the wavefront aberrations of the anterior cornea. Six rotationally symmetric aspheric surfaces were measured. The surface elevation was computed from the normal measurement files. Both elevation error and relative error were computed from the average of the three maps. The root-mean-square errors for the various surfaces ranged from 1.48 to 6.55 microns, with less error on the oblate and spherical surfaces. The error found was very systematic, increasing monotonically toward the periphery. The article includes a strategy to compensate for the systematic error to meet the required 0.5-micron accuracy needed. An equation was developed that used only the apical radius and shape factor, which improved the accuracy to the required 0.5-micron level.</p></div>\",\"PeriodicalId\":80286,\"journal\":{\"name\":\"International contact lens clinic (New York, N.Y.)\",\"volume\":\"25 6\",\"pages\":\"Pages 171-177\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1998-11-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1016/S0892-8967(99)00005-X\",\"citationCount\":\"6\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International contact lens clinic (New York, N.Y.)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S089289679900005X\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International contact lens clinic (New York, N.Y.)","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S089289679900005X","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 6

摘要

本文考察了eyeesys 2000视频角化镜在测量非球面计算前角膜波前像差的准确性。测量了六个旋转对称非球面。地表高程由常规测量文件计算。高程误差和相对误差由三幅图的平均值计算。不同表面的均方根误差在1.48 ~ 6.55 μ m之间,扁面和球面误差较小。所发现的误差具有很强的系统性,向外围单调递增。本文包括补偿系统误差的策略,以满足所需的0.5微米精度。开发了一个仅使用根尖半径和形状因子的方程,将精度提高到所需的0.5微米水平。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Accuracy of the EyeSys 2000 in measuring surface elevation of calibrated aspheres

This article examines the accuracy of the EyeSys 2000 videokeratoscope in measuring aspheric surfaces for calculation of the wavefront aberrations of the anterior cornea. Six rotationally symmetric aspheric surfaces were measured. The surface elevation was computed from the normal measurement files. Both elevation error and relative error were computed from the average of the three maps. The root-mean-square errors for the various surfaces ranged from 1.48 to 6.55 microns, with less error on the oblate and spherical surfaces. The error found was very systematic, increasing monotonically toward the periphery. The article includes a strategy to compensate for the systematic error to meet the required 0.5-micron accuracy needed. An equation was developed that used only the apical radius and shape factor, which improved the accuracy to the required 0.5-micron level.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信