{"title":"椎弓根螺钉断裂一例研究","authors":"Giorgio E.O. Giacaglia ,&nbsp;Wendell de Queiroz Lamas","doi":"10.1016/j.csefa.2015.09.001","DOIUrl":null,"url":null,"abstract":"<div><p>In this work we present a technical description related to the rupture of a titanium alloy pedicle screw and connecting bar implanted in dorsal vertebras of a patient. Only metallurgical facts are described, with no attempt to identify any imperfections in the clinical aspects related to the rupture. The results described here are based on extensive analysis of the broken materials in a material sciences specialized laboratory. Excluding an incorrect prosthesis implantation in the surgical procedure and a possible low bone density, an information not available to the research team, with high probability the rupture of metallic pieces used in the prosthetic implant, was produced by the low fatigue resistance resulting by an improper machining process and excessive bending of the connecting bar prior to implant.</p></div>","PeriodicalId":91224,"journal":{"name":"Case studies in engineering failure analysis","volume":"4 ","pages":"Pages 64-75"},"PeriodicalIF":0.0000,"publicationDate":"2015-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/j.csefa.2015.09.001","citationCount":"6","resultStr":"{\"title\":\"Pedicle screw rupture: A case study\",\"authors\":\"Giorgio E.O. Giacaglia ,&nbsp;Wendell de Queiroz Lamas\",\"doi\":\"10.1016/j.csefa.2015.09.001\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><p>In this work we present a technical description related to the rupture of a titanium alloy pedicle screw and connecting bar implanted in dorsal vertebras of a patient. Only metallurgical facts are described, with no attempt to identify any imperfections in the clinical aspects related to the rupture. The results described here are based on extensive analysis of the broken materials in a material sciences specialized laboratory. Excluding an incorrect prosthesis implantation in the surgical procedure and a possible low bone density, an information not available to the research team, with high probability the rupture of metallic pieces used in the prosthetic implant, was produced by the low fatigue resistance resulting by an improper machining process and excessive bending of the connecting bar prior to implant.</p></div>\",\"PeriodicalId\":91224,\"journal\":{\"name\":\"Case studies in engineering failure analysis\",\"volume\":\"4 \",\"pages\":\"Pages 64-75\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2015-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1016/j.csefa.2015.09.001\",\"citationCount\":\"6\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Case studies in engineering failure analysis\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S221329021500019X\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Case studies in engineering failure analysis","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S221329021500019X","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 6

摘要

在这项工作中,我们提出了一个技术描述相关的钛合金椎弓根螺钉和连接杆植入病人的背椎体破裂。只描述了冶金方面的事实,没有试图确定与破裂有关的临床方面的任何缺陷。这里描述的结果是基于对材料科学专业实验室中破碎材料的广泛分析。排除手术过程中不正确的假体植入和可能的低骨密度,研究小组无法获得的信息表明,假体植入中使用的金属件破裂的大概率是由于加工工艺不当和植入前连接杆过度弯曲导致的低疲劳抗力造成的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Pedicle screw rupture: A case study

In this work we present a technical description related to the rupture of a titanium alloy pedicle screw and connecting bar implanted in dorsal vertebras of a patient. Only metallurgical facts are described, with no attempt to identify any imperfections in the clinical aspects related to the rupture. The results described here are based on extensive analysis of the broken materials in a material sciences specialized laboratory. Excluding an incorrect prosthesis implantation in the surgical procedure and a possible low bone density, an information not available to the research team, with high probability the rupture of metallic pieces used in the prosthetic implant, was produced by the low fatigue resistance resulting by an improper machining process and excessive bending of the connecting bar prior to implant.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信