{"title":"恢复时间、信号和噪声之间的相互作用:串联和并联调谐电路并不总是相同的","authors":"Joel B. Miller, B. Suits, A. N. Garroway, M. Hepp","doi":"10.1002/(SICI)1099-0534(2000)12:3<125::AID-CMR2>3.3.CO;2-G","DOIUrl":null,"url":null,"abstract":"The quality factor, Q, of the magnetic resonance probe and the manner in which the probe is coupled to the spectrometer play important roles in the quality of the obtained experimental data, affecting the signal-to-noise ratio (SNR), the probe recovery time, radiation-damping, and the coupling between multiple coils. For example, high SNR often requires a high-Q probe, while a short recovery time to observe of broad signals generally requires low Q. Here, we discuss effecting a compromise between SNR and recovery time by adjusting the coupling of the probe and preamplifier by mismatching their impedances. Unlike the more common impedance-matched case, the conditions for optimum performance for parallel- and series-tuned probes are found to differ significantly. © 2000 John Wiley & Sons, Inc. Concepts Magn Reson 12: 125–136, 2000","PeriodicalId":89665,"journal":{"name":"Concepts in magnetic resonance","volume":"277 6","pages":"125-136"},"PeriodicalIF":0.0000,"publicationDate":"2000-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"18","resultStr":"{\"title\":\"Interplay among recovery time, signal, and noise: series- and parallel-tuned circuits are not always the same\",\"authors\":\"Joel B. Miller, B. Suits, A. N. Garroway, M. Hepp\",\"doi\":\"10.1002/(SICI)1099-0534(2000)12:3<125::AID-CMR2>3.3.CO;2-G\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The quality factor, Q, of the magnetic resonance probe and the manner in which the probe is coupled to the spectrometer play important roles in the quality of the obtained experimental data, affecting the signal-to-noise ratio (SNR), the probe recovery time, radiation-damping, and the coupling between multiple coils. For example, high SNR often requires a high-Q probe, while a short recovery time to observe of broad signals generally requires low Q. Here, we discuss effecting a compromise between SNR and recovery time by adjusting the coupling of the probe and preamplifier by mismatching their impedances. Unlike the more common impedance-matched case, the conditions for optimum performance for parallel- and series-tuned probes are found to differ significantly. © 2000 John Wiley & Sons, Inc. Concepts Magn Reson 12: 125–136, 2000\",\"PeriodicalId\":89665,\"journal\":{\"name\":\"Concepts in magnetic resonance\",\"volume\":\"277 6\",\"pages\":\"125-136\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2000-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"18\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Concepts in magnetic resonance\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1002/(SICI)1099-0534(2000)12:3<125::AID-CMR2>3.3.CO;2-G\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Concepts in magnetic resonance","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1002/(SICI)1099-0534(2000)12:3<125::AID-CMR2>3.3.CO;2-G","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 18

摘要

磁共振探头的质量因子Q和探头与光谱仪的耦合方式对获得的实验数据质量有重要影响,影响到信噪比(SNR)、探头恢复时间、辐射阻尼和多线圈之间的耦合。例如,高信噪比通常需要高q的探头,而观察宽信号的短恢复时间通常需要低q。在这里,我们讨论通过调整探头和前置放大器的阻抗不匹配来调节信噪比和恢复时间之间的折衷。与更常见的阻抗匹配情况不同,并联和串联调谐探头的最佳性能条件有很大不同。©2000 John Wiley & Sons, Inc数学学报,12 (2):125-136,2000
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Interplay among recovery time, signal, and noise: series- and parallel-tuned circuits are not always the same
The quality factor, Q, of the magnetic resonance probe and the manner in which the probe is coupled to the spectrometer play important roles in the quality of the obtained experimental data, affecting the signal-to-noise ratio (SNR), the probe recovery time, radiation-damping, and the coupling between multiple coils. For example, high SNR often requires a high-Q probe, while a short recovery time to observe of broad signals generally requires low Q. Here, we discuss effecting a compromise between SNR and recovery time by adjusting the coupling of the probe and preamplifier by mismatching their impedances. Unlike the more common impedance-matched case, the conditions for optimum performance for parallel- and series-tuned probes are found to differ significantly. © 2000 John Wiley & Sons, Inc. Concepts Magn Reson 12: 125–136, 2000
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信