对血糖仪4型血糖系统用于近患者检测的适用性评估

M. Powell
{"title":"对血糖仪4型血糖系统用于近患者检测的适用性评估","authors":"M. Powell","doi":"10.1002/j.1528-252X.1994.tb00026.x","DOIUrl":null,"url":null,"abstract":"The performance of a new system for near‐patient monitoring of blood glucose levels based on the reference hexokinase/glucose‐6‐phosphate dehydrogenase method was evaluated in a routine out‐patient diabetes clinic. The system includes features designed to overcome operator dependency of results. Within‐batch precision was 1.1‐4.2% coefficient of variation, while between‐batch coefficients of variation of 2.6‐6.7% were achieved. The new system was assessed to be suitable for use by nurses and patients.","PeriodicalId":92116,"journal":{"name":"Practical diabetes international : the journal for diabetes care teams worldwide","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2021-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1002/j.1528-252X.1994.tb00026.x","citationCount":"1","resultStr":"{\"title\":\"An assessment of the suitability of the Glucometer 4 blood glucose system for near‐patient testing\",\"authors\":\"M. Powell\",\"doi\":\"10.1002/j.1528-252X.1994.tb00026.x\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The performance of a new system for near‐patient monitoring of blood glucose levels based on the reference hexokinase/glucose‐6‐phosphate dehydrogenase method was evaluated in a routine out‐patient diabetes clinic. The system includes features designed to overcome operator dependency of results. Within‐batch precision was 1.1‐4.2% coefficient of variation, while between‐batch coefficients of variation of 2.6‐6.7% were achieved. The new system was assessed to be suitable for use by nurses and patients.\",\"PeriodicalId\":92116,\"journal\":{\"name\":\"Practical diabetes international : the journal for diabetes care teams worldwide\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2021-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1002/j.1528-252X.1994.tb00026.x\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Practical diabetes international : the journal for diabetes care teams worldwide\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1002/j.1528-252X.1994.tb00026.x\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Practical diabetes international : the journal for diabetes care teams worldwide","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1002/j.1528-252X.1994.tb00026.x","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

摘要

在一个常规的糖尿病门诊评估了一种基于己糖激酶/葡萄糖- 6 -磷酸脱氢酶参考方法的近患者血糖水平监测新系统的性能。该系统包括一些功能,旨在克服操作员对结果的依赖。批内准确度为1.1 ~ 4.2%,批间准确度为2.6 ~ 6.7%。新系统被评估为适合护士和病人使用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
An assessment of the suitability of the Glucometer 4 blood glucose system for near‐patient testing
The performance of a new system for near‐patient monitoring of blood glucose levels based on the reference hexokinase/glucose‐6‐phosphate dehydrogenase method was evaluated in a routine out‐patient diabetes clinic. The system includes features designed to overcome operator dependency of results. Within‐batch precision was 1.1‐4.2% coefficient of variation, while between‐batch coefficients of variation of 2.6‐6.7% were achieved. The new system was assessed to be suitable for use by nurses and patients.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信