材料刚度对种植体稳定性的影响&有限元分析

Q4 Earth and Planetary Sciences
M. I. Ishak, R. Daud, S. M. Mohd Noor
{"title":"材料刚度对种植体稳定性的影响&有限元分析","authors":"M. I. Ishak, R. Daud, S. M. Mohd Noor","doi":"10.11113/aej.v13.18087","DOIUrl":null,"url":null,"abstract":"The perseverance of dental implant system in restoration of occlusion is highly dependent on biomechanical overloading factors such as implant macro geometries, parafunctional oral habits, and material. Different implant materials could affect the load transfer at the bone-implant interface differently which is related to stress shielding phenomenon. To date, the role of various implant materials on the surrounding tissues as well as implant stability is still debatable and unclear especially when the implant failure is of concern. Through this study, implant body with different materials or stiffnesses that are zirconia, Ti-6Al-4V, cpTi, TiZr, and PEEK were investigated via 3-D FEA. The bone tissues were modelled based on CT image datasets and subsequently be processed in SolidWorks software. All geometries were converted into finite element models and analysed in ANSYS software. The bone and implant models were assigned with anisotropic and isotropic properties, respectively. A dynamic occlusal loading of 300 N and pretension of 20 N were applied on the implant body and screw, respectively. The results showed that the less stiff implant increased the bone stress and decreased the implant body stress values compared to the stiffer implant. Moreover, the implant with lower stiffness exhibited lower bone strain and higher implant deformation than the implant with higher stiffness. Of all implant materials analysed, PEEK is observed to be the most satisfactory. However, further modifications on PEEK would be necessary to improve inherent bioactivity and osseointegration.","PeriodicalId":36749,"journal":{"name":"ASEAN Engineering Journal","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2023-02-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"THE EFFECT OF MATERIAL STIFFNESS ON DENTAL IMPLANT STABILITY – A FINITE ELEMENT ANALYSIS\",\"authors\":\"M. I. Ishak, R. Daud, S. M. Mohd Noor\",\"doi\":\"10.11113/aej.v13.18087\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The perseverance of dental implant system in restoration of occlusion is highly dependent on biomechanical overloading factors such as implant macro geometries, parafunctional oral habits, and material. Different implant materials could affect the load transfer at the bone-implant interface differently which is related to stress shielding phenomenon. To date, the role of various implant materials on the surrounding tissues as well as implant stability is still debatable and unclear especially when the implant failure is of concern. Through this study, implant body with different materials or stiffnesses that are zirconia, Ti-6Al-4V, cpTi, TiZr, and PEEK were investigated via 3-D FEA. The bone tissues were modelled based on CT image datasets and subsequently be processed in SolidWorks software. All geometries were converted into finite element models and analysed in ANSYS software. The bone and implant models were assigned with anisotropic and isotropic properties, respectively. A dynamic occlusal loading of 300 N and pretension of 20 N were applied on the implant body and screw, respectively. The results showed that the less stiff implant increased the bone stress and decreased the implant body stress values compared to the stiffer implant. Moreover, the implant with lower stiffness exhibited lower bone strain and higher implant deformation than the implant with higher stiffness. Of all implant materials analysed, PEEK is observed to be the most satisfactory. However, further modifications on PEEK would be necessary to improve inherent bioactivity and osseointegration.\",\"PeriodicalId\":36749,\"journal\":{\"name\":\"ASEAN Engineering Journal\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-02-28\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"ASEAN Engineering Journal\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.11113/aej.v13.18087\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Earth and Planetary Sciences\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"ASEAN Engineering Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.11113/aej.v13.18087","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Earth and Planetary Sciences","Score":null,"Total":0}
引用次数: 0

摘要

牙种植体系统在咬合修复中的持久性在很大程度上取决于生物力学过载因素,如种植体宏观几何形状、多功能口腔习惯和材料。不同的植入材料会对骨-植入物界面处的载荷传递产生不同的影响,这与应力屏蔽现象有关。到目前为止,各种植入材料对周围组织的作用以及植入物的稳定性仍然存在争议和不清楚,尤其是当植入物失效令人担忧时。通过本研究,通过三维有限元分析研究了不同材料或刚度的植入体,如氧化锆、Ti-6Al-4V、cpTi、TiZr和PEEK。骨组织基于CT图像数据集进行建模,随后在SolidWorks软件中进行处理。将所有几何形状转换为有限元模型,并在ANSYS软件中进行分析。骨和植入物模型分别具有各向异性和各向同性特性。在植入体和螺钉上分别施加300N的动态咬合载荷和20N的预张力。结果表明,与刚性较大的植入物相比,刚性较小的植入物增加了骨应力,并降低了植入物本体应力值。此外,与具有较高刚度的植入物相比,具有较低硬度的植入物表现出较低的骨应变和较高的植入物变形。在分析的所有植入材料中,PEEK是最令人满意的。然而,需要对PEEK进行进一步的修饰,以提高固有的生物活性和骨整合。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
THE EFFECT OF MATERIAL STIFFNESS ON DENTAL IMPLANT STABILITY – A FINITE ELEMENT ANALYSIS
The perseverance of dental implant system in restoration of occlusion is highly dependent on biomechanical overloading factors such as implant macro geometries, parafunctional oral habits, and material. Different implant materials could affect the load transfer at the bone-implant interface differently which is related to stress shielding phenomenon. To date, the role of various implant materials on the surrounding tissues as well as implant stability is still debatable and unclear especially when the implant failure is of concern. Through this study, implant body with different materials or stiffnesses that are zirconia, Ti-6Al-4V, cpTi, TiZr, and PEEK were investigated via 3-D FEA. The bone tissues were modelled based on CT image datasets and subsequently be processed in SolidWorks software. All geometries were converted into finite element models and analysed in ANSYS software. The bone and implant models were assigned with anisotropic and isotropic properties, respectively. A dynamic occlusal loading of 300 N and pretension of 20 N were applied on the implant body and screw, respectively. The results showed that the less stiff implant increased the bone stress and decreased the implant body stress values compared to the stiffer implant. Moreover, the implant with lower stiffness exhibited lower bone strain and higher implant deformation than the implant with higher stiffness. Of all implant materials analysed, PEEK is observed to be the most satisfactory. However, further modifications on PEEK would be necessary to improve inherent bioactivity and osseointegration.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
ASEAN Engineering Journal
ASEAN Engineering Journal Engineering-Engineering (all)
CiteScore
0.60
自引率
0.00%
发文量
75
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信