{"title":"利用星形跟踪装置分析6 mv和10 mv光子光束轮廓,保证直线加速器的质量","authors":"Basma Sabr Lazim, Basim Rejah, H. Alabedi","doi":"10.22038/IJMP.2019.39987.1543","DOIUrl":null,"url":null,"abstract":"Introduction: According to the American Society of Radiation Oncology, all patients receive radiation therapy during their illness, where radiation is delivered by the medical linear accelerator (Linac). The aim of this study was to evaluate the quality assurance (QA) of the Linac in analyzing the used dose profile in the treatment of cancer tumors. \nMaterial and Methods: This experimental study was performed using Linac (synergy device type) at Baghdad Radiotherapy and Nuclear Medicine laboratories, Baghdad, Iraq. The Star Track device was used for the routine quality assurance of the Linac, using photon beam for the reference Dmax and source to surface distance of 100 cm. The Star Track consists of 453 vented parallel plate ionization chambers. \nResults: The flatness and symmetry of beams for the reference field size did not exceed from ±2%, as they were within the allowed range. Moreover, the penumbra region showed a change in value that did not exceed from ±0.2 cm. using the Star Track method; maximum differences in beam symmetry and beam flatness were measured at 0.76%±2% and 1.17%±2%, respectively. Moreover, the maximum difference in the penumbra region was estimated at 0.12±0.2 cm. \nConclusion: The results indicated, the Star Track could successfully calculate the characteristics of dose profile during a time period of 2,500 ms, showing the superiority of this instrument over other verification devices.","PeriodicalId":14613,"journal":{"name":"Iranian Journal of Medical Physics","volume":"17 1","pages":"260-265"},"PeriodicalIF":0.0000,"publicationDate":"2020-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"Quality Assurance of LINAC by Analyzing the Profile of 6-MV and 10-MV Photon Beams Using Star Track Device\",\"authors\":\"Basma Sabr Lazim, Basim Rejah, H. Alabedi\",\"doi\":\"10.22038/IJMP.2019.39987.1543\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Introduction: According to the American Society of Radiation Oncology, all patients receive radiation therapy during their illness, where radiation is delivered by the medical linear accelerator (Linac). The aim of this study was to evaluate the quality assurance (QA) of the Linac in analyzing the used dose profile in the treatment of cancer tumors. \\nMaterial and Methods: This experimental study was performed using Linac (synergy device type) at Baghdad Radiotherapy and Nuclear Medicine laboratories, Baghdad, Iraq. The Star Track device was used for the routine quality assurance of the Linac, using photon beam for the reference Dmax and source to surface distance of 100 cm. The Star Track consists of 453 vented parallel plate ionization chambers. \\nResults: The flatness and symmetry of beams for the reference field size did not exceed from ±2%, as they were within the allowed range. Moreover, the penumbra region showed a change in value that did not exceed from ±0.2 cm. using the Star Track method; maximum differences in beam symmetry and beam flatness were measured at 0.76%±2% and 1.17%±2%, respectively. Moreover, the maximum difference in the penumbra region was estimated at 0.12±0.2 cm. \\nConclusion: The results indicated, the Star Track could successfully calculate the characteristics of dose profile during a time period of 2,500 ms, showing the superiority of this instrument over other verification devices.\",\"PeriodicalId\":14613,\"journal\":{\"name\":\"Iranian Journal of Medical Physics\",\"volume\":\"17 1\",\"pages\":\"260-265\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2020-07-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Iranian Journal of Medical Physics\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.22038/IJMP.2019.39987.1543\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Health Professions\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Iranian Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.22038/IJMP.2019.39987.1543","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Health Professions","Score":null,"Total":0}
引用次数: 1

摘要

简介:根据美国放射肿瘤学学会的说法,所有患者在患病期间都会接受放射治疗,通过医用线性加速器(Linac)进行放射治疗。本研究的目的是评估Linac在分析癌症肿瘤治疗中使用的剂量分布时的质量保证(QA)。材料和方法:本实验研究在伊拉克巴格达的巴格达放射治疗和核医学实验室使用Linac(协同装置型)进行。星轨装置用于直线加速器的常规质量保证,使用光子束作为参考Dmax,源到表面的距离为100厘米。星轨由453个通风平行板电离室组成。结果:参考场尺寸的光束的平面度和对称性不超过±2%,因为它们在允许范围内。此外,使用星轨法,半影区域的数值变化不超过±0.2厘米;光束对称性和光束平面度的最大差异分别为0.76%±2%和1.17%±2%。此外,半暗带区域的最大差异估计为0.12±0.2cm。结论:结果表明,Star Track可以成功地计算出2500ms时间段内的剂量分布特征,表明该仪器优于其他验证设备。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Quality Assurance of LINAC by Analyzing the Profile of 6-MV and 10-MV Photon Beams Using Star Track Device
Introduction: According to the American Society of Radiation Oncology, all patients receive radiation therapy during their illness, where radiation is delivered by the medical linear accelerator (Linac). The aim of this study was to evaluate the quality assurance (QA) of the Linac in analyzing the used dose profile in the treatment of cancer tumors. Material and Methods: This experimental study was performed using Linac (synergy device type) at Baghdad Radiotherapy and Nuclear Medicine laboratories, Baghdad, Iraq. The Star Track device was used for the routine quality assurance of the Linac, using photon beam for the reference Dmax and source to surface distance of 100 cm. The Star Track consists of 453 vented parallel plate ionization chambers. Results: The flatness and symmetry of beams for the reference field size did not exceed from ±2%, as they were within the allowed range. Moreover, the penumbra region showed a change in value that did not exceed from ±0.2 cm. using the Star Track method; maximum differences in beam symmetry and beam flatness were measured at 0.76%±2% and 1.17%±2%, respectively. Moreover, the maximum difference in the penumbra region was estimated at 0.12±0.2 cm. Conclusion: The results indicated, the Star Track could successfully calculate the characteristics of dose profile during a time period of 2,500 ms, showing the superiority of this instrument over other verification devices.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Iranian Journal of Medical Physics
Iranian Journal of Medical Physics Health Professions-Radiological and Ultrasound Technology
CiteScore
1.00
自引率
0.00%
发文量
0
审稿时长
8 weeks
期刊介绍: Iranian Journal of Medical Physics (IJMP) is the official scientific bimonthly publication of the Iranian Association of Medical Physicists. IJMP is an international and multidisciplinary journal, peer review, free of charge publication and open access. This journal devoted to publish Original Papers, Review Articles, Short Communications, Technical Notes, Editorial and Letters to the Editor in the field of “Medical Physics” involving both basic and clinical research. Submissions of manuscript from all countries are welcome and will be reviewed by at least two expert reviewers.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信