{"title":"利用星形跟踪装置分析6 mv和10 mv光子光束轮廓,保证直线加速器的质量","authors":"Basma Sabr Lazim, Basim Rejah, H. Alabedi","doi":"10.22038/IJMP.2019.39987.1543","DOIUrl":null,"url":null,"abstract":"Introduction: According to the American Society of Radiation Oncology, all patients receive radiation therapy during their illness, where radiation is delivered by the medical linear accelerator (Linac). The aim of this study was to evaluate the quality assurance (QA) of the Linac in analyzing the used dose profile in the treatment of cancer tumors. \nMaterial and Methods: This experimental study was performed using Linac (synergy device type) at Baghdad Radiotherapy and Nuclear Medicine laboratories, Baghdad, Iraq. The Star Track device was used for the routine quality assurance of the Linac, using photon beam for the reference Dmax and source to surface distance of 100 cm. The Star Track consists of 453 vented parallel plate ionization chambers. \nResults: The flatness and symmetry of beams for the reference field size did not exceed from ±2%, as they were within the allowed range. Moreover, the penumbra region showed a change in value that did not exceed from ±0.2 cm. using the Star Track method; maximum differences in beam symmetry and beam flatness were measured at 0.76%±2% and 1.17%±2%, respectively. Moreover, the maximum difference in the penumbra region was estimated at 0.12±0.2 cm. \nConclusion: The results indicated, the Star Track could successfully calculate the characteristics of dose profile during a time period of 2,500 ms, showing the superiority of this instrument over other verification devices.","PeriodicalId":14613,"journal":{"name":"Iranian Journal of Medical Physics","volume":"17 1","pages":"260-265"},"PeriodicalIF":0.0000,"publicationDate":"2020-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"Quality Assurance of LINAC by Analyzing the Profile of 6-MV and 10-MV Photon Beams Using Star Track Device\",\"authors\":\"Basma Sabr Lazim, Basim Rejah, H. Alabedi\",\"doi\":\"10.22038/IJMP.2019.39987.1543\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Introduction: According to the American Society of Radiation Oncology, all patients receive radiation therapy during their illness, where radiation is delivered by the medical linear accelerator (Linac). The aim of this study was to evaluate the quality assurance (QA) of the Linac in analyzing the used dose profile in the treatment of cancer tumors. \\nMaterial and Methods: This experimental study was performed using Linac (synergy device type) at Baghdad Radiotherapy and Nuclear Medicine laboratories, Baghdad, Iraq. The Star Track device was used for the routine quality assurance of the Linac, using photon beam for the reference Dmax and source to surface distance of 100 cm. The Star Track consists of 453 vented parallel plate ionization chambers. \\nResults: The flatness and symmetry of beams for the reference field size did not exceed from ±2%, as they were within the allowed range. Moreover, the penumbra region showed a change in value that did not exceed from ±0.2 cm. using the Star Track method; maximum differences in beam symmetry and beam flatness were measured at 0.76%±2% and 1.17%±2%, respectively. Moreover, the maximum difference in the penumbra region was estimated at 0.12±0.2 cm. \\nConclusion: The results indicated, the Star Track could successfully calculate the characteristics of dose profile during a time period of 2,500 ms, showing the superiority of this instrument over other verification devices.\",\"PeriodicalId\":14613,\"journal\":{\"name\":\"Iranian Journal of Medical Physics\",\"volume\":\"17 1\",\"pages\":\"260-265\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2020-07-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Iranian Journal of Medical Physics\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.22038/IJMP.2019.39987.1543\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Health Professions\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Iranian Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.22038/IJMP.2019.39987.1543","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Health Professions","Score":null,"Total":0}
Quality Assurance of LINAC by Analyzing the Profile of 6-MV and 10-MV Photon Beams Using Star Track Device
Introduction: According to the American Society of Radiation Oncology, all patients receive radiation therapy during their illness, where radiation is delivered by the medical linear accelerator (Linac). The aim of this study was to evaluate the quality assurance (QA) of the Linac in analyzing the used dose profile in the treatment of cancer tumors.
Material and Methods: This experimental study was performed using Linac (synergy device type) at Baghdad Radiotherapy and Nuclear Medicine laboratories, Baghdad, Iraq. The Star Track device was used for the routine quality assurance of the Linac, using photon beam for the reference Dmax and source to surface distance of 100 cm. The Star Track consists of 453 vented parallel plate ionization chambers.
Results: The flatness and symmetry of beams for the reference field size did not exceed from ±2%, as they were within the allowed range. Moreover, the penumbra region showed a change in value that did not exceed from ±0.2 cm. using the Star Track method; maximum differences in beam symmetry and beam flatness were measured at 0.76%±2% and 1.17%±2%, respectively. Moreover, the maximum difference in the penumbra region was estimated at 0.12±0.2 cm.
Conclusion: The results indicated, the Star Track could successfully calculate the characteristics of dose profile during a time period of 2,500 ms, showing the superiority of this instrument over other verification devices.
期刊介绍:
Iranian Journal of Medical Physics (IJMP) is the official scientific bimonthly publication of the Iranian Association of Medical Physicists. IJMP is an international and multidisciplinary journal, peer review, free of charge publication and open access. This journal devoted to publish Original Papers, Review Articles, Short Communications, Technical Notes, Editorial and Letters to the Editor in the field of “Medical Physics” involving both basic and clinical research. Submissions of manuscript from all countries are welcome and will be reviewed by at least two expert reviewers.