{"title":"电子手腕血压监测装置的准确性","authors":"B. Bamgbose, B. Bavitz, F. Qian","doi":"10.61172/ndj.v23i2.24","DOIUrl":null,"url":null,"abstract":"Background: Assessment of blood pressure is the most common diagnostic procedure performed in the outpatient clinic. The purpose of this study was to compare the accuracy of the wrist-cuff electronic oscillometric device with the universally-accepted gold standard measurement made with Korotkoff sound technique mercury sphygmomanometer.Materials and Methods: This was a randomized, crossover study with a total of 60 subjects. All subjects were at least 30 years of age. Three blood pressure readings were recorded for each subject, two with a wrist-cuff device and one with a mercury sphygmomanometer by the secondary investigator. The order of the 3 readings were determined by chance.Results: Patients (32 women, 28 men) aged 30-94 (mean age = 55, SD =16) years were included in the study. Based on the paired-sample t-test, the data provided strong evidence that there was no statistically significant difference in systolic or diastolic blood pressure readings between the wrist-cuff electronic oscillometric device and mercury sphygmomanometer.Conclusions: The evaluation of blood pressure remains a basic diagnostic step in every clinical practice as it positively imparts the clinical well-being of the patient. It is, therefore, important to maintain a good degree of accuracy and reliability in the blood pressure monitoring device. \n ","PeriodicalId":79241,"journal":{"name":"Nigerian Dental Journal","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2020-09-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"The Accuracy of Electronic Wrist Blood Pressure Monitoring Devices\",\"authors\":\"B. Bamgbose, B. Bavitz, F. Qian\",\"doi\":\"10.61172/ndj.v23i2.24\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Background: Assessment of blood pressure is the most common diagnostic procedure performed in the outpatient clinic. The purpose of this study was to compare the accuracy of the wrist-cuff electronic oscillometric device with the universally-accepted gold standard measurement made with Korotkoff sound technique mercury sphygmomanometer.Materials and Methods: This was a randomized, crossover study with a total of 60 subjects. All subjects were at least 30 years of age. Three blood pressure readings were recorded for each subject, two with a wrist-cuff device and one with a mercury sphygmomanometer by the secondary investigator. The order of the 3 readings were determined by chance.Results: Patients (32 women, 28 men) aged 30-94 (mean age = 55, SD =16) years were included in the study. Based on the paired-sample t-test, the data provided strong evidence that there was no statistically significant difference in systolic or diastolic blood pressure readings between the wrist-cuff electronic oscillometric device and mercury sphygmomanometer.Conclusions: The evaluation of blood pressure remains a basic diagnostic step in every clinical practice as it positively imparts the clinical well-being of the patient. It is, therefore, important to maintain a good degree of accuracy and reliability in the blood pressure monitoring device. \\n \",\"PeriodicalId\":79241,\"journal\":{\"name\":\"Nigerian Dental Journal\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2020-09-23\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Nigerian Dental Journal\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.61172/ndj.v23i2.24\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nigerian Dental Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.61172/ndj.v23i2.24","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
The Accuracy of Electronic Wrist Blood Pressure Monitoring Devices
Background: Assessment of blood pressure is the most common diagnostic procedure performed in the outpatient clinic. The purpose of this study was to compare the accuracy of the wrist-cuff electronic oscillometric device with the universally-accepted gold standard measurement made with Korotkoff sound technique mercury sphygmomanometer.Materials and Methods: This was a randomized, crossover study with a total of 60 subjects. All subjects were at least 30 years of age. Three blood pressure readings were recorded for each subject, two with a wrist-cuff device and one with a mercury sphygmomanometer by the secondary investigator. The order of the 3 readings were determined by chance.Results: Patients (32 women, 28 men) aged 30-94 (mean age = 55, SD =16) years were included in the study. Based on the paired-sample t-test, the data provided strong evidence that there was no statistically significant difference in systolic or diastolic blood pressure readings between the wrist-cuff electronic oscillometric device and mercury sphygmomanometer.Conclusions: The evaluation of blood pressure remains a basic diagnostic step in every clinical practice as it positively imparts the clinical well-being of the patient. It is, therefore, important to maintain a good degree of accuracy and reliability in the blood pressure monitoring device.