{"title":"减少气管插管患者气管粘膜损伤的装置系统研究","authors":"Dai Won Suh, Seung Bong Lee, Sung Min Kim","doi":"10.1115/1.4054334","DOIUrl":null,"url":null,"abstract":"\n The purpose of this study was to develop an automatic bronchial aspiration system to minimize tracheal mucosal damage in ventilator patients. Operation performances of the system's suction pressure, bronchial intubation depth, suction cycle, and tube cuff pressure were tested. To check clinical results, subjects underwent endoscopy after applying the previous manual method for 24 hours. After that, they underwent endoscopy after applying the proposed suction system for 24 hours. For quantitative evaluation of test results, tracheal mucosal injury was divided into five grades: Grade 0 = normal, Grade 1 = erythema or edema, Grade 2 = erosion, Grade 3 = hemorrhage, and Grade 4 = ulcer or necrosis. In the performance test, an error of up to 12 mmHg occurred within the normal operation error range for suction pressure control. The insertion depth control had a maximum error of 7.0 mm within the normal operation error range. On the other hand, there was no error in the time control or the tube cuff pressure control. In the clinical trial, after using the proposed system for five subjects to find changes in tracheal mucosal injury by endoscopy, reduced injury or no change in injury was found. The system proposed in this study is confirmed to be able to remove sputum while minimizing tracheal mucosal injury that can occur when using previous manual suction device.","PeriodicalId":49305,"journal":{"name":"Journal of Medical Devices-Transactions of the Asme","volume":null,"pages":null},"PeriodicalIF":0.8000,"publicationDate":"2022-04-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Study On Device System to Reduce Tracheal Mucosal Injury in Intubation Patients\",\"authors\":\"Dai Won Suh, Seung Bong Lee, Sung Min Kim\",\"doi\":\"10.1115/1.4054334\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"\\n The purpose of this study was to develop an automatic bronchial aspiration system to minimize tracheal mucosal damage in ventilator patients. Operation performances of the system's suction pressure, bronchial intubation depth, suction cycle, and tube cuff pressure were tested. To check clinical results, subjects underwent endoscopy after applying the previous manual method for 24 hours. After that, they underwent endoscopy after applying the proposed suction system for 24 hours. For quantitative evaluation of test results, tracheal mucosal injury was divided into five grades: Grade 0 = normal, Grade 1 = erythema or edema, Grade 2 = erosion, Grade 3 = hemorrhage, and Grade 4 = ulcer or necrosis. In the performance test, an error of up to 12 mmHg occurred within the normal operation error range for suction pressure control. The insertion depth control had a maximum error of 7.0 mm within the normal operation error range. On the other hand, there was no error in the time control or the tube cuff pressure control. In the clinical trial, after using the proposed system for five subjects to find changes in tracheal mucosal injury by endoscopy, reduced injury or no change in injury was found. The system proposed in this study is confirmed to be able to remove sputum while minimizing tracheal mucosal injury that can occur when using previous manual suction device.\",\"PeriodicalId\":49305,\"journal\":{\"name\":\"Journal of Medical Devices-Transactions of the Asme\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.8000,\"publicationDate\":\"2022-04-14\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Medical Devices-Transactions of the Asme\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1115/1.4054334\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Devices-Transactions of the Asme","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1115/1.4054334","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

摘要

本研究的目的是开发一种自动支气管抽吸系统,以最大限度地减少呼吸机患者的气管粘膜损伤。系统的运行性能';s抽吸压力、支气管插管深度、抽吸周期和袖带压力。为了检查临床结果,受试者在应用之前的手动方法24小时后接受内窥镜检查。之后,他们在应用拟议的抽吸系统24小时后接受了内窥镜检查。为了定量评估测试结果,将气管粘膜损伤分为五个等级:0级=正常,1级=红斑或水肿,2级=侵蚀,3级=出血,4级=溃疡或坏死。在性能测试中,在吸入压力控制的正常操作误差范围内出现了高达12毫米汞柱的误差。插入深度控制在正常操作误差范围内具有7.0mm的最大误差。另一方面,时间控制或袖带压力控制没有错误。在临床试验中,在对五名受试者使用所提出的系统通过内窥镜检查发现气管粘膜损伤的变化后,发现损伤减少或损伤没有变化。本研究中提出的系统已被证实能够清除痰液,同时最大限度地减少使用以前的手动抽吸装置时可能发生的气管粘膜损伤。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Study On Device System to Reduce Tracheal Mucosal Injury in Intubation Patients
The purpose of this study was to develop an automatic bronchial aspiration system to minimize tracheal mucosal damage in ventilator patients. Operation performances of the system's suction pressure, bronchial intubation depth, suction cycle, and tube cuff pressure were tested. To check clinical results, subjects underwent endoscopy after applying the previous manual method for 24 hours. After that, they underwent endoscopy after applying the proposed suction system for 24 hours. For quantitative evaluation of test results, tracheal mucosal injury was divided into five grades: Grade 0 = normal, Grade 1 = erythema or edema, Grade 2 = erosion, Grade 3 = hemorrhage, and Grade 4 = ulcer or necrosis. In the performance test, an error of up to 12 mmHg occurred within the normal operation error range for suction pressure control. The insertion depth control had a maximum error of 7.0 mm within the normal operation error range. On the other hand, there was no error in the time control or the tube cuff pressure control. In the clinical trial, after using the proposed system for five subjects to find changes in tracheal mucosal injury by endoscopy, reduced injury or no change in injury was found. The system proposed in this study is confirmed to be able to remove sputum while minimizing tracheal mucosal injury that can occur when using previous manual suction device.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
1.80
自引率
11.10%
发文量
56
审稿时长
6-12 weeks
期刊介绍: The Journal of Medical Devices presents papers on medical devices that improve diagnostic, interventional and therapeutic treatments focusing on applied research and the development of new medical devices or instrumentation. It provides special coverage of novel devices that allow new surgical strategies, new methods of drug delivery, or possible reductions in the complexity, cost, or adverse results of health care. The Design Innovation category features papers focusing on novel devices, including papers with limited clinical or engineering results. The Medical Device News section provides coverage of advances, trends, and events.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信