用计算机断层扫描金属伪影减少算法评估放射治疗计划

Q3 Health Professions
Kyungjin Kim, Jihun Choi, Joo-Wan Hong, Jaeouk Ahn
{"title":"用计算机断层扫描金属伪影减少算法评估放射治疗计划","authors":"Kyungjin Kim, Jihun Choi, Joo-Wan Hong, Jaeouk Ahn","doi":"10.22038/IJMP.2020.52418.1864","DOIUrl":null,"url":null,"abstract":"Introduction: The objective of this study was to evaluate the usefulness of Computed Tomography(CT) images acquired through repeated subtraction reconstruction algorithms to reduce metal artifacts in CT Treatment Planning System(TPS).Materials and methods: Origin images of Gammex phantom and Rando phantom and non orthopedic metal artifact reduction(O-MAR) images were obtained after high density implantation. O-MAR applied images were also obtained. For evaluation of images, regions of interest(ROI) were set at five tissue rods and three points directly affected by artifacts in Gammex phantom. CT number and noise were compared and analyzed. To investigate dose change according to CT number change of the surrounding artifact, a three virtual cylindrical target volume was formed on Rando phantom CT images. The average dose was then compared and analyzed.Results: CT number difference according to the application of O-MAR showed significant difference among lung and bone rod and 3 ROI directly affected. Noise difference according to O-MAR application was significantly different in rod except for bone rod. In the treatment plan using Rando phantom, non-O-MAR and O-MAR images showed -4.3 ~ 1.9% and -0.4 ~ 2.3% dose differences, respectively.Conclusion: Applying a metal artifact reduction reconstruction algorithm can reduce image distortion due to high density implantation, improve image quality, and correct CT number.","PeriodicalId":14613,"journal":{"name":"Iranian Journal of Medical Physics","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2020-12-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Evaluation of radiation treatment planning by computed tomography metal artifact reduction algorithm\",\"authors\":\"Kyungjin Kim, Jihun Choi, Joo-Wan Hong, Jaeouk Ahn\",\"doi\":\"10.22038/IJMP.2020.52418.1864\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Introduction: The objective of this study was to evaluate the usefulness of Computed Tomography(CT) images acquired through repeated subtraction reconstruction algorithms to reduce metal artifacts in CT Treatment Planning System(TPS).Materials and methods: Origin images of Gammex phantom and Rando phantom and non orthopedic metal artifact reduction(O-MAR) images were obtained after high density implantation. O-MAR applied images were also obtained. For evaluation of images, regions of interest(ROI) were set at five tissue rods and three points directly affected by artifacts in Gammex phantom. CT number and noise were compared and analyzed. To investigate dose change according to CT number change of the surrounding artifact, a three virtual cylindrical target volume was formed on Rando phantom CT images. The average dose was then compared and analyzed.Results: CT number difference according to the application of O-MAR showed significant difference among lung and bone rod and 3 ROI directly affected. Noise difference according to O-MAR application was significantly different in rod except for bone rod. In the treatment plan using Rando phantom, non-O-MAR and O-MAR images showed -4.3 ~ 1.9% and -0.4 ~ 2.3% dose differences, respectively.Conclusion: Applying a metal artifact reduction reconstruction algorithm can reduce image distortion due to high density implantation, improve image quality, and correct CT number.\",\"PeriodicalId\":14613,\"journal\":{\"name\":\"Iranian Journal of Medical Physics\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2020-12-09\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Iranian Journal of Medical Physics\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.22038/IJMP.2020.52418.1864\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Health Professions\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Iranian Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.22038/IJMP.2020.52418.1864","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Health Professions","Score":null,"Total":0}
引用次数: 0

摘要

引言:本研究的目的是评估通过重复减法重建算法获得的计算机断层扫描(CT)图像在CT治疗计划系统(TPS)中减少金属伪影的有用性。材料和方法:高密度后获得Gammex体模和Rando体模的原始图像以及非矫形金属伪影减少(O-MAR)图像植入。还获得了O-MAR应用图像。为了评估图像,将感兴趣区域(ROI)设置在Gammex体模中直接受伪影影响的五个组织棒和三个点上。比较分析CT数目和噪声。为了研究剂量随周围伪影CT数量变化的变化,在Rando体模CT图像上形成了一个三虚拟圆柱形目标体积。然后对平均剂量进行比较和分析。结果:根据O-MAR的应用,CT数字差异显示肺和骨棒之间存在显著差异,3个ROI直接受影响。根据O-MAR应用的噪声差异在除骨棒外的棒中有显著差异。在使用Rando体模的治疗计划中,非O-MAR和O-MAR图像显示的剂量差异分别为-4.3~1.9%和-0.4~2.3%。结论:应用金属伪影减少重建算法可以减少高密度植入引起的图像失真,提高图像质量,校正CT数量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Evaluation of radiation treatment planning by computed tomography metal artifact reduction algorithm
Introduction: The objective of this study was to evaluate the usefulness of Computed Tomography(CT) images acquired through repeated subtraction reconstruction algorithms to reduce metal artifacts in CT Treatment Planning System(TPS).Materials and methods: Origin images of Gammex phantom and Rando phantom and non orthopedic metal artifact reduction(O-MAR) images were obtained after high density implantation. O-MAR applied images were also obtained. For evaluation of images, regions of interest(ROI) were set at five tissue rods and three points directly affected by artifacts in Gammex phantom. CT number and noise were compared and analyzed. To investigate dose change according to CT number change of the surrounding artifact, a three virtual cylindrical target volume was formed on Rando phantom CT images. The average dose was then compared and analyzed.Results: CT number difference according to the application of O-MAR showed significant difference among lung and bone rod and 3 ROI directly affected. Noise difference according to O-MAR application was significantly different in rod except for bone rod. In the treatment plan using Rando phantom, non-O-MAR and O-MAR images showed -4.3 ~ 1.9% and -0.4 ~ 2.3% dose differences, respectively.Conclusion: Applying a metal artifact reduction reconstruction algorithm can reduce image distortion due to high density implantation, improve image quality, and correct CT number.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Iranian Journal of Medical Physics
Iranian Journal of Medical Physics Health Professions-Radiological and Ultrasound Technology
CiteScore
1.00
自引率
0.00%
发文量
0
审稿时长
8 weeks
期刊介绍: Iranian Journal of Medical Physics (IJMP) is the official scientific bimonthly publication of the Iranian Association of Medical Physicists. IJMP is an international and multidisciplinary journal, peer review, free of charge publication and open access. This journal devoted to publish Original Papers, Review Articles, Short Communications, Technical Notes, Editorial and Letters to the Editor in the field of “Medical Physics” involving both basic and clinical research. Submissions of manuscript from all countries are welcome and will be reviewed by at least two expert reviewers.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信